Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT.

David A Hirschl1, Vineet R Jain, Hugo Spindola-Franco, Jay N Gross, Linda B Haramati.   

Abstract

BACKGROUND: Pacemakers and implantable cardiac defibrillators (ICDs) are widely used for the management of cardiac arrhythmias and congestive heart failure (CHF). Acute implantation complication rates range from 3% to 7%. The aim of this study is to describe the incidence of lead perforation on computed tomography (CT), and correlate these findings with electrophysiologic data.
METHODS: Images of 100 consecutive patients with permanent pacemakers (n = 72) or ICDs (n = 28) who underwent multidetector CTs of the chest were identified. Cases were reviewed by 2 cardiothoracic radiologists, and a third if there was disagreement. Each CT was reviewed for device and fixation type, tip position, and presence of pericardial effusion. Results were correlated with lead impedance and pacing threshold, when available [79% (79/100)]. A cardiac electrophysiologist interpreted device data.
RESULTS: All 100 patients had right ventricular leads (58 passive, 42 active) and 61 had right atrial leads (12 passive, 49 active). 15% (15/100) of patients had a lead perforation. Perforation rates were 15% (9/61) for atrial and 6% (6/100) for ventricular leads (P < 0.05, chi square). Four of 28 (14%) right ventricular ICD leads and 2 of 72 (3%)right ventricular pacemaker leads were perforated (P < 0.05, chi square). 12% (6/49) of active right atrial leads, and 25% (3/12) of passive right atrial leads perforated (P = NS, chi square). 7% (3/42) of active right ventricular leads, and 5% (3/58) of passive ventricular leads perforated (P = NS, chi square). Electrophysiologic parameters did not differ significantly between perforated and nonperforated leads.
CONCLUSION: Asymptomatic perforation is a common phenomenon and rarely resulting in electrophysiologic consequences. Atrial leads perforated more frequently than ventricular leads, and ventricular ICD leads perforated more frequently than ventricular pacemaker leads.

Entities:  

Mesh:

Year:  2007        PMID: 17241311     DOI: 10.1111/j.1540-8159.2007.00575.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  50 in total

1.  What could be the cause of late syncope after pacemaker implantation?

Authors:  Mohammad Hossein Mandegar; Farideh Roshanali; Helia Farokhzad; Mahmood Eftekharzadeh
Journal:  BMJ Case Rep       Date:  2012-02-25

2.  Uncomplicated prior right ventricular perforation due to a pacemaker lead - an incidental finding during coronary artery bypass surgery.

Authors:  Can Yerebakan; Bernd Westphal; Selcan Bayraktar; Anthony Alozie; Andreas Liebold; Gustav Steinhoff
Journal:  Wien Med Wochenschr       Date:  2011-12

3.  An Itchy Lead: First Reported Case of Ventricular Pacemaker Lead Self-Extraction.

Authors:  Brent Klinkhammer; Mevan Wijetunga; Yassar Almanaseer
Journal:  J Atr Fibrillation       Date:  2018-08-31

Review 4.  [Strategies for the avoidance and treatment of complications during pacemaker implantation].

Authors:  C Kolb
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-12

5.  Pacemaker lead perforation causing hemopericardium eight years after implantation.

Authors:  Jackson J Liang; Ammar M Killu; Michael J Osborn; Allan S Jaffe
Journal:  Indian Heart J       Date:  2013-04-09

Review 6.  Radiology of cardiac devices and their complications.

Authors:  J Dipoce; A Bernheim; H Spindola-Franco
Journal:  Br J Radiol       Date:  2014-11-20       Impact factor: 3.039

Review 7.  Management of Accidental and Iatrogenic Foreign Body Injuries to Heart- Case Series.

Authors:  Rupesh Kumar; Sandeep Singh Rana; Sanjay Kumar; Deepanwita Das; Monalisa Datta
Journal:  J Clin Diagn Res       Date:  2017-03-01

8.  Assessment of Cardiac Lead Perforation: Comparison Among Chest Radiography, Transthoracic Echocardiography and Electrocardiography-gated Contrast-enhanced Cardiac CT.

Authors:  Xiang Zhang; Chushan Zheng; Peiwei Wang; Dongye Wang; Boshui Huang; Guozhao Li; Huijun Hu; Zehong Yang; Xiaohui Duan; Shaoxin Zheng; Pinming Liu; Jingfeng Wang; Jun Shen
Journal:  Eur Radiol       Date:  2018-07-17       Impact factor: 5.315

9.  Late perforation of Durata ICD lead--a case report.

Authors:  Bharati Das; Sachin Yalagudri; S Maiya; Sunita Abhraham; J Kannan; Ravi Kishore
Journal:  Indian Heart J       Date:  2014 Jul-Aug

Review 10.  State-of-the-art narrative review: multimodality imaging in electrophysiology and cardiac device therapies.

Authors:  Balint Laczay; Divyang Patel; Richard Grimm; Bo Xu
Journal:  Cardiovasc Diagn Ther       Date:  2021-06