Quantification accuracy and partial volume effect in dependence of the attenuation correction of a state-of-the-art small animal PET scanner.

Julia G Mannheim1, Martin S Judenhofer, Andreas Schmid, Julia Tillmanns, Detlef Stiller, Vesna Sossi, Bernd J Pichler.   

Abstract

Quantification accuracy and partial volume effect (PVE) of the Siemens Inveon PET scanner were evaluated. The influence of transmission source activities (40 and 160 MBq) on the quantification accuracy and the PVE were determined. Dynamic range, object size and PVE for different sphere sizes, contrast ratios and positions in the field of view (FOV) were evaluated. The acquired data were reconstructed using different algorithms and correction methods. The activity level of the transmission source and the total emission activity in the FOV strongly influenced the attenuation maps. Reconstruction algorithms, correction methods, object size and location within the FOV had a strong influence on the PVE in all configurations. All evaluated parameters potentially influence the quantification accuracy. Hence, all protocols should be kept constant during a study to allow a comparison between different scans.

Mesh:

Year:  2012        PMID: 22647967     DOI: 10.1088/0031-9155/57/12/3981

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  15 in total

1.  Scatter characterization and correction for simultaneous multiple small-animal PET imaging.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2014-04       Impact factor: 3.488

2.  The Synergistic Effect of Selumetinib/Docetaxel Combination Therapy Monitored by [(18)F]FDG/[(18)F]FLT PET and Diffusion-Weighted Magnetic Resonance Imaging in a Colorectal Tumor Xenograft Model.

Authors:  Valerie S Honndorf; Holger Schmidt; Stefan Wiehr; Hans F Wehrl; Leticia Quintanilla-Martinez; Anke Stahlschmidt; Hervé Barjat; Sally-Ann Emmas; Bernd J Pichler
Journal:  Mol Imaging Biol       Date:  2016-04       Impact factor: 3.488

Review 3.  Quantitative Rodent Brain Receptor Imaging.

Authors:  Kristina Herfert; Julia G Mannheim; Laura Kuebler; Sabina Marciano; Mario Amend; Christoph Parl; Hanna Napieczynska; Florian M Maier; Salvador Castaneda Vega; Bernd J Pichler
Journal:  Mol Imaging Biol       Date:  2020-04       Impact factor: 3.488

4.  Feasibility of sequential PET/MRI using a state-of-the-art small animal PET and a 1 T benchtop MRI.

Authors:  Andreas Schmid; Jennifer Schmitz; Julia G Mannheim; Florian C Maier; Kerstin Fuchs; Hans F Wehrl; Bernd J Pichler
Journal:  Mol Imaging Biol       Date:  2013-04       Impact factor: 3.488

Review 5.  Standardization of Small Animal Imaging-Current Status and Future Prospects.

Authors:  Julia G Mannheim; Firat Kara; Janine Doorduin; Kerstin Fuchs; Gerald Reischl; Sayuan Liang; Marleen Verhoye; Felix Gremse; Laura Mezzanotte; Marc C Huisman
Journal:  Mol Imaging Biol       Date:  2018-10       Impact factor: 3.488

6.  Influence of Co-57 and CT Transmission Measurements on the Quantification Accuracy and Partial Volume Effect of a Small Animal PET Scanner.

Authors:  Julia G Mannheim; Andreas M Schmid; Bernd J Pichler
Journal:  Mol Imaging Biol       Date:  2017-12       Impact factor: 3.488

7.  Utility of 3'-[(18)F]fluoro-3'-deoxythymidine as a PET tracer to monitor response to gene therapy in a xenograft model of head and neck carcinoma.

Authors:  Neale S Mason; Brian J Lopresti; James Ruszkiewicz; Xinxin Dong; Sonali Joyce; George Leef; Malabika Sen; Abdus S Wahed; Chester A Mathis; Jennifer R Grandis; Sufi M Thomas
Journal:  Am J Nucl Med Mol Imaging       Date:  2013-01-05

8.  Positron emission tomography based in-vivo imaging of early phase stem cell retention after intramyocardial delivery in the mouse model.

Authors:  Cajetan Lang; Sebastian Lehner; Andrei Todica; Guido Boening; Wolfgang-Michael Franz; Peter Bartenstein; Marcus Hacker; Robert David
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-07-17       Impact factor: 9.236

9.  Evaluation of a commercial multi-dimensional echocardiography technique for ventricular volumetry in small animals.

Authors:  Jana Grune; Annelie Blumrich; Sarah Brix; Sarah Jeuthe; Cathleen Drescher; Tilman Grune; Anna Foryst-Ludwig; Daniel Messroghli; Wolfgang M Kuebler; Christiane Ott; Ulrich Kintscher
Journal:  Cardiovasc Ultrasound       Date:  2018-07-03       Impact factor: 2.062

10.  Preclinical evaluation of the anti-tumor effects of the natural isoflavone genistein in two xenograft mouse models monitored by [18F]FDG, [18F]FLT, and [64Cu]NODAGA-cetuximab small animal PET.

Authors:  Valerie S Honndorf; Stefan Wiehr; Anna-Maria Rolle; Julia Schmitt; Luisa Kreft; Letitia Quintanilla-Martinez; Ursula Kohlhofer; Gerald Reischl; Andreas Maurer; Karsten Boldt; Michael Schwarz; Holger Schmidt; Bernd J Pichler
Journal:  Oncotarget       Date:  2016-05-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.