| Literature DB >> 22643044 |
Derek R Cooney1, Norma L Cooney, Harry Wallus, Susan Wojcik.
Abstract
BACKGROUND: This study was designed to evaluate emergency physician success and satisfaction using a video-assisted semi-rigid fiberoptic stylet, the Clarus Video System (CVS), during a simulated difficult airway scenario.Entities:
Year: 2012 PMID: 22643044 PMCID: PMC3407739 DOI: 10.1186/1865-1380-5-24
Source DB: PubMed Journal: Int J Emerg Med ISSN: 1865-1372
Figure 1The Clarus Video System (Clarus Medical, Minneapolis, MN).
Figure 2Endotracheal tube in position over the CVS.
Figure 3Views as seen on the CVS screen: A. View of the larynx of the mannequin with ETT loaded on the CVS. B. View of the mannequin’s trachea after positioning the end of the stylet through the vocal cords. C. View of the ETT passing through the cords as it is advanced off of the CVS. D. Confirmation of ETT placement by viewing the larynx with the CVS.
Participant demographics
| Training Year: | 1st year = 6 (30%) | 2nd year = 4 (20%) | 3rd year = 3 (15%) | Attending = 7 (35%) |
| Handedness: | right = 17(85%) | left = 3(15%) | ambidextrous = 0(0%) | |
| Eye dominance: | right = 11(55%) | left = 2(10%) | unknown = 7(35%) | |
| Age Range: | 25 yo–43 yo (mean = 33 yo) | | | |
| Sex: | male = 16 (80%) / female = 4 (20%) | | | |
| | none | simulated patient | actual patient | |
| Claurs Video Scope: | 20 (100%) | 0 (0%) | 0 (0%) | |
| other optical stylet: | 14 (70%) | 5 (25%) | 2 (10%) | |
| flexible fiberscope: | 14 (70%) | 3 (15%) | 3 (15%) | |
| Glidescope: | 3 (15%) | 7 (35%) | 16 (80%) | |
| other video-laryngoscope: | 16 (80%) | 3 (15%) | 2 (10%) | |
| Airtraq: | 17 (85%) | 2 (10%) | 0 (0%) | |
| lighted stylet: | 16 (80%) | 3 (15%) | 1 (5%) | |
| gum elastic bougie: | 4 (20%) | 9 (45%) | 13 (65%) | |
Individual results for EP intubation attempts by technique
| | DL attempt #1 | DL attempt #2 | DL attempt #3 | CVS attempt #1 | CVS attempt #2 | CVS attempt #3 |
| 1 | 118.5 | 74.6* | n/a | 102.3* | n/a | n/a |
| 2 | 70.9 | 59.4 | 43.8* | 27.7* | n/a | n/a |
| 3 | 39.8* | n/a | n/a | 134.8* | n/a | n/a |
| 4 | 43.1* | n/a | n/a | 15* | n/a | n/a |
| 5 | 15.6* | n/a | n/a | 11.7* | n/a | n/a |
| 6 | 82.4 | 39.9* | n/a | 16.7* | n/a | n/a |
| 7 | 13* | n/a | n/a | 8.2* | n/a | n/a |
| 8 | 15.9* | n/a | n/a | 21* | n/a | n/a |
| 9 | 100.6 | 58 | 25.4 | 86.8* | n/a | n/a |
| 10 | 92.4 | 79.7 | 46* | 26.6* | n/a | n/a |
| 11 | 34.1 | 62.3* | n/a | 50.5* | n/a | n/a |
| 12 | 106* | n/a | n/a | 20.9* | n/a | n/a |
| 13 | 22.3* | n/a | n/a | 26.4* | n/a | n/a |
| 14 | 35.7* | n/a | n/a | 16.5* | n/a | n/a |
| 15 | 20.6* | n/a | n/a | 60.8* | n/a | n/a |
| 16 | 17.5* | n/a | n/a | 14* | n/a | n/a |
| 17 | 58.3* | n/a | n/a | 26.3* | n/a | n/a |
| 18 | 24.7* | n/a | n/a | 25.1* | n/a | n/a |
| 19 | 29.3 | 36.4* | n/a | 22.5* | n/a | n/a |
| 20 | 24.8* | n/a | n/a | 60.3* | n/a | n/a |
Mean number of intubation attempts and times for intubation attempts (per attempt and cumulative)
| | Direct laryngoscopy (DL) | Clarus Video System(CVS) | |
| Mean number of attempts per provider: | 1.5 (SD = ±0.76) | 1 (SD = ±0) | |
| Mean time for each attempt: | 43.41 (SD = ±26.82) | 38.71 (SD = ±34.14) | |
| Mean time for all attempts: | 74.55 (SD = ±68.40) | 38.71 (SD = ±34.14) | |