Xuan Liu1, Marcin Balicki, Russell H Taylor, Jin U Kang.
Abstract
We present a new automatic spectral calibration (ASC) method for spectral Domain optical coherence tomography (SD-OCT). Our ASC method calibrates the spectral mapping of the spectrometer in SD-OCT, and does not require external calibrating light source or a commercial spectral analyzer. The ASC method simultaneously calibrates the physical pixel spacing of the A-scan in static and dynamic environments. Experimental results show that the proposed ASC method can provide satisfactory calibration for SD-OCT to achieve high axial resolution and high ranging accuracy, without increasing hardware complexity.Entities:
Year: 2011 PMID: 22611481 PMCID: PMC3354775 DOI: 10.1117/12.875210
Source DB: PubMed Journal: Proc SPIE Int Soc Opt Eng ISSN: 0277-786X