Comparison of hybrid and pure iterative reconstruction techniques with conventional filtered back projection: dose reduction potential in the abdomen.

Sarabjeet Singh1, Mannudeep K Kalra, Synho Do, Jean Baptiste Thibault, Homer Pien, Owen J O'Connor, Owen O J Connor, Michael A Blake.   

Abstract

PURPOSE: Assess the effect of filtered back projection (FBP) and hybrid (adaptive statistical iterative reconstruction [ASIR]) and pure (model-based iterative reconstruction [MBIR]) iterative reconstructions on abdominal computed tomography (CT) acquired with 75% radiation dose reduction.
MATERIALS AND METHODS: In an institutional review board-approved prospective study, 10 patients (mean [standard deviation] age, 60 (8) years; 4 men and 6 women) gave informed consent for acquisition of additional abdominal images on 64-slice multidetector-row CT (GE 750HD, GE Healthcare). Scanning was repeated over a 10-cm scan length at 200 and 50 milliampere second (mA s), with remaining parameters held constant at 120 kilovolt (peak), 0.984:1 pitch, and standard reconstruction kernel. Projection data were deidentified, exported, and reconstructed to obtain 4 data sets (200-mA s FBP, 50-mA s FBP, 50-mA s ASIR, 50-mA s MBIR), which were evaluated by 2 abdominal radiologists for lesions and subjective image quality. Objective noise and noise spectral density were measured for each image series.
RESULTS: Among the 10 patients, the maximum weight recorded was 123 kg, with maximum transverse diameter measured as 43.7 cm. Lesion conspicuity at 50-mA s MBIR was better than on 50-mA s FBP and ASIR images (P < 0.01). Image noise was rated as suboptimal on low-dose FBP and ASIR but deemed acceptable in MBIR images. Objective noise with 50-mA s MBIR was 2 to 3 folds lower compared to 50-mA s ASIR, 50-mA s FBP, and 200-mA s FBP (P < 0.0001). Noise spectral density analyses demonstrated that ASIR retains the noise spectrum signature of FBP, whereas MBIR has much lower noise with a more regularized noise spectrum pattern.
CONCLUSION: Model-based iterative reconstruction renders acceptable image quality and diagnostic confidence in 50- mA s abdominal CT images, whereas FBP and ASIR images are associated with suboptimal image quality at this radiation dose level.

Entities:  

Mesh:

Year:  2012        PMID: 22592622     DOI: 10.1097/RCT.0b013e31824e639e

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  41 in total

1.  Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom.

Authors:  Young Jin Ryu; Young Hun Choi; Jung-Eun Cheon; Seongmin Ha; Woo Sun Kim; In-One Kim
Journal:  Pediatr Radiol       Date:  2015-11-06

2.  Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp.

Authors:  Jin Hyeok Kim; Ki Seok Choo; Tae Yong Moon; Jun Woo Lee; Ung Bae Jeon; Tae Un Kim; Jae Yeon Hwang; Myeong-Ja Yun; Dong Wook Jeong; Soo Jin Lim
Journal:  Eur Radiol       Date:  2015-10-20       Impact factor: 5.315

3.  Image quality and dose optimisation for infant CT using a paediatric phantom.

Authors:  Jack W Lambert; Andrew S Phelps; Jesse L Courtier; Robert G Gould; John D MacKenzie
Journal:  Eur Radiol       Date:  2015-08-26       Impact factor: 5.315

4.  Quantifying the effect of slice thickness, intravenous contrast and tube current on muscle segmentation: Implications for body composition analysis.

Authors:  Georg Fuchs; Yves R Chretien; Julia Mario; Synho Do; Matthias Eikermann; Bob Liu; Kai Yang; Florian J Fintelmann
Journal:  Eur Radiol       Date:  2018-01-09       Impact factor: 5.315

5.  Assessment of sub-milli-sievert abdominal computed tomography with iterative reconstruction techniques of different vendors.

Authors:  Atul Padole; Nisha Sainani; Diego Lira; Ranish Deedar Ali Khawaja; Sarvenaz Pourjabbar; Roberto Lo Gullo; Alexi Otrakji; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2016-06-28

6.  Prioritizing examination-centered over patient-centered dose reduction: a hazard of institutional "benchmarking".

Authors:  Jonathan D Eisenberg; Michael E Gilmore; Mannudeep K Kalra; Chung Yin Kong; Pari V Pandharipande
Journal:  AJR Am J Roentgenol       Date:  2014-05       Impact factor: 3.959

7.  Standardization and optimization of CT protocols to achieve low dose.

Authors:  Sigal Trattner; Gregory D N Pearson; Cynthia Chin; Dianna D Cody; Rajiv Gupta; Christopher P Hess; Mannudeep K Kalra; James M Kofler; Mayil S Krishnam; Andrew J Einstein
Journal:  J Am Coll Radiol       Date:  2014-03       Impact factor: 5.532

8.  Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT.

Authors:  Ethan A Smith; Jonathan R Dillman; Mitchell M Goodsitt; Emmanuel G Christodoulou; Nahid Keshavarzi; Peter J Strouse
Journal:  Radiology       Date:  2013-10-29       Impact factor: 11.105

9.  Evaluation of Abdominal Computed Tomography Image Quality Using a New Version of Vendor-Specific Model-Based Iterative Reconstruction.

Authors:  Corey T Jensen; Morgan E Telesmanich; Nicolaus A Wagner-Bartak; Xinming Liu; John Rong; Janio Szklaruk; Aliya Qayyum; Wei Wei; Adam G Chandler; Eric P Tamm
Journal:  J Comput Assist Tomogr       Date:  2017-01       Impact factor: 1.826

10.  CT reconstruction algorithms affect histogram and texture analysis: evidence for liver parenchyma, focal solid liver lesions, and renal cysts.

Authors:  Su Joa Ahn; Jung Hoon Kim; Sang Min Lee; Sang Joon Park; Joon Koo Han
Journal:  Eur Radiol       Date:  2018-11-19       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.