An electromagnetic "Tracker-in-Table" configuration for X-ray fluoroscopy and cone-beam CT-guided surgery.

J Yoo1, S Schafer, A Uneri, Y Otake, A J Khanna, J H Siewerdsen.   

Abstract

PURPOSE: A novel electromagnetic tracking configuration was characterized and implemented for image-guided surgery incorporating C-arm fluoroscopy and/or cone-beam CT (CBCT). The tracker employed a field generator (FG) with an open rectangular aperture and a frame enclosure with two essentially hollow sides, yielding a design that presents little or no X-ray attenuation across the C-arm orbit. The "Window" FG (WFG) was characterized in comparison with a conventional "Aurora" FG (AFG), and a configuration in which the WFG was incorporated directly into the operating table was investigated in preclinical phantom studies.
METHOD: The geometric accuracy and field of view (FOV) of the WFG and AFG were evaluated in terms of target registration error (TRE) using an acrylic phantom on an (electromagnetic compatible) experimental bench. The WFG design was incorporated in a prototype operating table featuring a carbon fiber top beneath, which the FG could be translated for positioning under the patient. The X-ray compatibility was evaluated using a prototype mobile C-arm for fluoroscopy and CBCT in an anthropomorphic chest phantom. The susceptibility to EM field distortion associated with surgical tools (e.g., spine screws) and the C-arm itself was investigated in terms of TRE, and calibration methods were tested to provide robust image-world registration with minimal perturbation from the rotational C-arm.
RESULTS: The WFG demonstrated mean TRE of 1.28 ± 0.79 mm compared to 1.13 ± 0.72 mm for the AFG, with no statistically significant difference between the two (p = 0.32 and n = 250). The WFG exhibited a deeper field of view by ~10 cm providing an equivalent degree of geometric accuracy to a depth of z ~55 cm, compared to z ~45 cm for the AFG. Although the presence of a small number of spine screws did not degrade tracker accuracy, the mobile C-arm perturbed the electromagnetic field sufficiently to degrade TRE; however, a calibration method was identified to mitigate the effect. Specifically, the average calibration between posterior-anterior and lateral orientations of the C-arm was found to yield fairly robust registration for any C-arm pose with only a slight reduction in geometric accuracy (1.43 ± 0.31 mm in comparison with 1.28 ± 0.79 mm, p = 0.05). The WFG demonstrated reasonable X-ray compatibility, although the initial design of the window frame included suboptimal material and shape of the side bars that caused a level of streak artifacts in CBCT reconstructions. The streak artifacts were of sufficient magnitude to degrade soft-tissue visibility in CBCT but were negligible in the context of high-contrast imaging tasks (e.g., bone visualization).
CONCLUSION: The open frame of the WFG offers a potentially valuable configuration for electromagnetic trackers in image-guided surgery applications that are based on X-ray fluoroscopy and/or CBCT. The geometric accuracy and FOV are comparable to the conventional AFG and offers increased depth (z-direction) FOV. Incorporation directly within the operating table offers a streamlined implementation in which the tracker is in place but "invisible," potentially simplifying tableside logistics, avoidance of the sterile field, and compatibility with X-ray imaging.

Entities:  

Mesh:

Year:  2012        PMID: 22585463      PMCID: PMC9092243          DOI: 10.1007/s11548-012-0744-z

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   3.421


  27 in total

1.  Derivation of linear attenuation coefficients from CT numbers for low-energy photons.

Authors:  Y Watanabe
Journal:  Phys Med Biol       Date:  1999-09       Impact factor: 3.609

2.  Interference during the use of an electromagnetic tracking system under OR conditions.

Authors:  François Poulin; L-P Amiot
Journal:  J Biomech       Date:  2002-06       Impact factor: 2.712

3.  Evaluation of a miniature electromagnetic position tracker.

Authors:  Johann Hummel; Michael Figl; Christian Kollmann; Helmar Bergmann; Wolfgang Birkfellner
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

4.  Accuracy assessment protocols for electromagnetic tracking systems.

Authors:  D D Frantz; A D Wiles; S E Leis; S R Kirsch
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

5.  Three-dimensional tomosynthesis and cone-beam computed tomography: an experimental study for fast, low-dose intraoperative imaging technology for guidance of sinus and skull base surgery.

Authors:  Gideon Bachar; Emma Barker; Sajendra Nithiananthan; Harley Chan; Michael J Daly; Jonathan C Irish; Jeffrey H Siewerdsen
Journal:  Laryngoscope       Date:  2009-03       Impact factor: 3.325

6.  Electromagnetic tracking in the clinical environment.

Authors:  Ziv Yaniv; Emmanuel Wilson; David Lindisch; Kevin Cleary
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

7.  Cone beam CT-based three-dimensional planning in high-dose-rate brachytherapy for cervical cancer.

Authors:  Hani Al-Halabi; Lorraine Portelance; Marie Duclos; Brigitte Reniers; Boris Bahoric; Luis Souhami
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-30       Impact factor: 7.038

8.  Mobile C-arm cone-beam CT for guidance of spine surgery: image quality, radiation dose, and integration with interventional guidance.

Authors:  S Schafer; S Nithiananthan; D J Mirota; A Uneri; J W Stayman; W Zbijewski; C Schmidgunst; G Kleinszig; A J Khanna; J H Siewerdsena
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

9.  Evaluation of a new electromagnetic tracking system using a standardized assessment protocol.

Authors:  J Hummel; M Figl; W Birkfellner; M R Bax; R Shahidi; C R Maurer; H Bergmann
Journal:  Phys Med Biol       Date:  2006-04-26       Impact factor: 3.609

10.  Intraoperative cone-beam CT for image-guided tibial plateau fracture reduction.

Authors:  A Khoury; J H Siewerdsen; C M Whyne; M J Daly; H J Kreder; D J Moseley; D A Jaffray
Journal:  Comput Aided Surg       Date:  2007-07
View more
  4 in total

1.  Preclinical evaluation of a prototype freehand drill video guidance system for orthopedic surgery.

Authors:  Niral Sheth; Prasad Vagdargi; Alejandro Sisniega; Ali Uneri; Gregory Osgood; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2022-08-26

2.  Planning, guidance, and quality assurance of pelvic screw placement using deformable image registration.

Authors:  J Goerres; A Uneri; M Jacobson; B Ramsay; T De Silva; M Ketcha; R Han; A Manbachi; S Vogt; G Kleinszig; J-P Wolinsky; G Osgood; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2017-11-13       Impact factor: 3.609

3.  Drill-mounted video guidance for orthopaedic trauma surgery.

Authors:  Prasad Vagdargi; Niral Sheth; Alejandro Sisniega; Ali Uneri; Tharindu De Silva; Greg M Osgood; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2021-02-12

4.  Robust methods for automatic image-to-world registration in cone-beam CT interventional guidance.

Authors:  H Dang; Y Otake; S Schafer; J W Stayman; G Kleinszig; J H Siewerdsen
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.