Literature DB >> 16675856

Evaluation of a new electromagnetic tracking system using a standardized assessment protocol.

J Hummel1, M Figl, W Birkfellner, M R Bax, R Shahidi, C R Maurer, H Bergmann.   

Abstract

This note uses a published protocol to evaluate a newly released 6 degrees of freedom electromagnetic tracking system (Aurora, Northern Digital Inc.). A practice for performance monitoring over time is also proposed. The protocol uses a machined base plate to measure relative error in position and orientation as well as the influence of metallic objects in the operating volume. Positional jitter (E(RMS)) was found to be 0.17 mm +/- 0.19 mm. A relative positional error of 0.25 mm +/- 0.22 mm at 50 mm offsets and 0.97 mm +/- 1.01 mm at 300 mm offsets was found. The mean of the relative rotation error was found to be 0.20 degrees +/- 0.14 degrees with respect to the axial and 0.91 degrees +/- 0.68 degrees for the longitudinal rotation. The most significant distortion caused by metallic objects is caused by 400-series stainless steel. A 9.4 mm maximum error occurred when the rod was closest to the emitter, 10 mm away. The improvement compared to older generations of the Aurora with respect to accuracy is substantial.

Entities:  

Mesh:

Year:  2006        PMID: 16675856     DOI: 10.1088/0031-9155/51/10/N01

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  19 in total

1.  Marker-free registration for the accurate integration of CT images and the subject's anatomy during navigation surgery of the maxillary sinus.

Authors:  S-H Kang; M-K Kim; J-H Kim; H-K Park; W Park
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

2.  Development of a navigation system for minimally invasive esophagectomy.

Authors:  H G Kenngott; J Neuhaus; B P Müller-Stich; I Wolf; M Vetter; H-P Meinzer; J Köninger; M W Büchler; C N Gutt
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

3.  A low-cost tracked C-arm (TC-arm) upgrade system for versatile quantitative intraoperative imaging.

Authors:  Shahram Amiri; David R Wilson; Bassam A Masri; Carolyn Anglin
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-10       Impact factor: 2.924

4.  A software solution to dynamically reduce metallic distortions of electromagnetic tracking systems for image-guided surgery.

Authors:  Mengfei Li; Christian Hansen; Georg Rose
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-03       Impact factor: 2.924

5.  Anser EMT: the first open-source electromagnetic tracking platform for image-guided interventions.

Authors:  Herman Alexander Jaeger; Alfred Michael Franz; Kilian O'Donoghue; Alexander Seitel; Fabian Trauzettel; Lena Maier-Hein; Pádraig Cantillon-Murphy
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-29       Impact factor: 2.924

6.  Synthetic tracked aperture ultrasound imaging: design, simulation, and experimental evaluation.

Authors:  Haichong K Zhang; Alexis Cheng; Nick Bottenus; Xiaoyu Guo; Gregg E Trahey; Emad M Boctor
Journal:  J Med Imaging (Bellingham)       Date:  2016-04-08

7.  A simulator for advanced analysis of a 5-DOF EM tracking systems in use for image-guided surgery.

Authors:  Mengfei Li; Christian Hansen; Georg Rose
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-08-29       Impact factor: 2.924

8.  An open electromagnetic tracking framework applied to targeted liver tumour ablation.

Authors:  Stephen Hinds; Herman Alexander Jaeger; Richard Burke; Brodie O'Sullivan; Joseph Keane; Fabian Trauzettel; Bruno Marques; Stéphane Cotin; Brian Bird; Håkon Olav Leira; Erlend Fagertun Hofstad; Ole Vegard Solberg; Thomas Langø; Pádraig Cantillon-Murphy
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-04-27       Impact factor: 2.924

9.  Current Evidence in Image-Guided Liver Surgery.

Authors:  Amber L Simpson; T Peter Kingham
Journal:  J Gastrointest Surg       Date:  2016-03-08       Impact factor: 3.452

10.  Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy.

Authors:  Jochen Krücker; Sheng Xu; Neil Glossop; Anand Viswanathan; Jörn Borgert; Heinrich Schulz; Bradford J Wood
Journal:  J Vasc Interv Radiol       Date:  2007-09       Impact factor: 3.464

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.