David M Huland, Christopher M Brown, Scott S Howard, Dimitre G Ouzounov, Ina Pavlova, Ke Wang, David R Rivera, Watt W Webb, Chris Xu.
Abstract
We characterize long (up to 285 mm) gradient index (GRIN) lens endoscope systems for multiphoton imaging. We fabricate a portable, rigid endoscope system suitable for imaging unstained tissues, potentially deep within the body, using a GRIN lens system of 1 mm diameter and 8 cm length. The portable device is capable of imaging a ~200 µm diameter field of view at 4 frames/s. The lateral and axial resolution in water is 0.85 µm and 7.4 µm respectively. In vivo images of unstained tissues in live, anesthetized rats using the portable device are presented. These results show great promise for GRIN endoscopy to be used clinically.Entities:
Keywords: (110.2760) Gradient-index lenses; (170.2150) Endoscopic imaging; (180.4315) Nonlinear microscopy
Year: 2012 PMID: 22567597 PMCID: PMC3342183 DOI: 10.1364/BOE.3.001077
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732
Summary of Optical Characterization Resultsa
| GRIN System | Part Number | Diameter (mm) | Length (mm) | Relay Lens Pitch | FWHM (µm) | Diameter of FOV (µm) | |
|---|---|---|---|---|---|---|---|
| Lateral | Axial | ||||||
| 1A | GT-ERLS-100-075-11-50-NC | 1.00 | 35.9 | 0.75 | 0.94 | 10.8 | 199 |
| 1B | GT-ERLS-100-175-11-50-NC | 1.00 | 79.9 | 1.75 | 0.97 | 12.0 | 195 |
| 2A | GT-ERLS-200-075-11-50-NC | 2.00 | 81.4 | 0.75 | 0.99 | 12.6 | 370 |
| 2B | GT-ERLS-200-125-11-50-NC | 2.00 | 132.6 | 1.25 | 1.05 | 15.6 | 365 |
| 2C | GT-ERLS-200-275-11-50-NC | 2.00 | 285.0 | 2.75 | 1.17 | 25.0 | 359 |
aFive different doublet GRIN systems were designed by and purchased from GRINTECH GmbH. The lateral and axial two-photon resolutions were determined in air using subresolution fluorescent beads and a thin rhodamine film respectively.
Fig. 1Experimental setup used for the optical characterization of the long gradient index endoscope systems and close-up of the doublet GRIN system design (shown here using a 0.75 relay lens pitch).
Fig. 2Two-photon lateral and axial resolution of GRIN system 2C (285 mm length). (a) Lateral intensity line profile across a subresolution fluorescent bead with a Gaussian fit in black. (b) Axial intensity profile across a thin film rhodamine slide with a Lorentzian fit in black.
Fig. 3Normalized one photon transmission intensity across the field of view for (a) GRIN system 1B and (b) GRIN system 2C.
Fig. 4Off-axis performance. Axial FWHM of GRIN system 2C in µm plotted (a) Across the FOV and (b) Across a line (dashed blue in (a)) through the center of the FOV. Scale bar is 50µm.
Fig. 5Portable GRIN endoscope. (a) Optical drawing and (b) Solidworks drawing of the GRIN based endoscope system. Total system length of the portable device is 10.6” (including GRIN system).
Fig. 6Unaveraged in vivo images of unstained rat tissue. The pseudo-color images show red SHG signal (<405 nm) and green intrinsic fluorescent emission (405-700 nm). (a) Image of the superficial kidney renal cortex shows dark renal interstitium (RI), dark cellular nuclei (N) and bright intrinsic fluorescent cytoplasm (CY) that form the epithelial cells in the renal tubules (RT), SHG signal from the tough fibrous layer that forms the renal capsule (RC), and the dark blood filled lumen (L) inside the renal tubules. (b) Image of the inner colon wall shows bright intrinsic fluorescent signal from entrocytes (E) surrounding dark circular crypts (C). (c) Image of the rat liver showing ~20 µm diameter hapatocytes (coarse dashed line) with dark nuclei (N, solid line) chained together to form hepatic chords (HC), a dark bile duct (BD, fine dashed line) and bright intrinsically fluorescent bile salts (BS), as well as SHG emission from the septa (S) a fibrous tissue bands that separates hapatocyte nodules. (d) Image of the rat liver without labels shown for clarity. In these images, scale bars are 20 µm.