Optimal reproducibility of gated sestamibi and thallium myocardial perfusion study left ventricular ejection fractions obtained on a solid-state CZT cardiac camera requires operator input.

Martin H Cherk1, Jason Ky, Kenneth S K Yap, Patrina Campbell, Catherine McGrath, Michael Bailey, Victor Kalff.   

Abstract

AIM: To evaluate the reproducibility of serial re-acquisitions of gated Tl-201 and Tc-99m sestamibi left ventricular ejection fraction (LVEF) measurements obtained on a new generation solid-state cardiac camera system during myocardial perfusion imaging and the importance of manual operator optimization of left ventricular wall tracking.
METHODS: Resting blinded automated (auto) and manual operator optimized (opt) LVEF measurements were measured using ECT toolbox (ECT) and Cedars-Sinai QGS software in two separate cohorts of 55 Tc-99m sestamibi (MIBI) and 50 thallium (Tl-201) myocardial perfusion studies (MPS) acquired in both supine and prone positions on a cadmium zinc telluride (CZT) solid-state camera system. Resting supine and prone automated LVEF measurements were similarly obtained in a further separate cohort of 52 gated cardiac blood pool scans (GCBPS) for validation of methodology and comparison. Appropriate use of Bland-Altman, chi-squared and Levene's equality of variance tests was used to analyse the resultant data comparisons.
RESULTS: For all radiotracer and software combinations, manual checking and optimization of valve planes (+/- centre radius with ECT software) resulted in significant improvement in MPS LVEF reproducibility that approached that of planar GCBPS. No difference was demonstrated between optimized MIBI/Tl-201 QGS and planar GCBPS LVEF reproducibility (P = .17 and P = .48, respectively). ECT required significantly more manual optimization compared to QGS software in both supine and prone positions independent of radiotracer used (P < .02).
CONCLUSIONS: Reproducibility of gated sestamibi and Tl-201 LVEF measurements obtained during myocardial perfusion imaging with ECT toolbox or QGS software packages using a new generation solid-state cardiac camera with improved image quality approaches that of planar GCBPS however requires visual quality control and operator optimization of left ventricular wall tracking for best results. Using this superior cardiac technology, Tl-201 reproducibility also appears at least equivalent to sestamibi for measuring LVEF.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22547397     DOI: 10.1007/s12350-012-9561-6

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  9 in total

1.  Comparison of post-stress ejection fraction and relative left ventricular volumes by automatic analysis of gated myocardial perfusion single-photon emission computed tomography acquired in the supine and prone positions.

Authors:  D Berman; G Germano; H Lewin; X Kang; P B Kavanagh; P Tapnio; M Harris; J Friedman
Journal:  J Nucl Cardiol       Date:  1998 Jan-Feb       Impact factor: 5.952

2.  Thallium myocardial perfusion scintigraphy: influence of perfusion, scatter, and photon energy on the detection of lesions.

Authors:  P H Vos; E K Pauwels; A M Vossepoel
Journal:  Cardiovasc Intervent Radiol       Date:  1986       Impact factor: 2.740

3.  Reproducibility of ejection fraction and ventricular volume by gated radionuclide angiography after myocardial infarction.

Authors:  R Slutsky; J Karliner; A Battler; M Pfisterer; S Swanson; W Ashburn
Journal:  Radiology       Date:  1979-07       Impact factor: 11.105

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Early image acquisition using a solid-state cardiac camera for fast myocardial perfusion imaging.

Authors:  J Wells Askew; Todd D Miller; Royce L Ruter; Lennon G Jordan; David O Hodge; Raymond J Gibbons; Michael K O'Connor
Journal:  J Nucl Cardiol       Date:  2011-07-12       Impact factor: 5.952

6.  Clinical validation of fully automated computation of ejection fraction from gated equilibrium blood-pool scintigrams.

Authors:  J H Reiber; S P Lie; M L Simoons; C Hoek; J J Gerbrands; W Wijns; W H Bakker; P P Kooij
Journal:  J Nucl Med       Date:  1983-12       Impact factor: 10.057

7.  Prone versus supine patient positioning during gated 99mTc-sestamibi SPECT: effect on left ventricular volumes, ejection fraction, and heart rate.

Authors:  Wolfgang M Schaefer; Claudia S A Lipke; Harald P Kühl; Karl-Christian Koch; Hans-Juergen Kaiser; Patrick Reinartz; Bernd Nowak; Udalrich Buell
Journal:  J Nucl Med       Date:  2004-12       Impact factor: 10.057

8.  Reproducibility of ejection-fraction determinations by equilibrium radionuclide angiography in response to supine bicycle exercise: concise communication.

Authors:  M E Pfisterer; A Battler; S M Swanson; R Slutsky; V Froelicher; W L Ashburn
Journal:  J Nucl Med       Date:  1979-06       Impact factor: 10.057

9.  Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging.

Authors:  Daniel S Berman; Xingping Kang; Balaji Tamarappoo; Arik Wolak; Sean W Hayes; Ryo Nakazato; Louise E J Thomson; Faith Kite; Ishac Cohen; Piotr J Slomka; Andrew J Einstein; John D Friedman
Journal:  JACC Cardiovasc Imaging       Date:  2009-03
  9 in total
  4 in total

Review 1.  Recent advances in cardiac SPECT instrumentation and system design.

Authors:  Mark F Smith
Journal:  Curr Cardiol Rep       Date:  2013-08       Impact factor: 2.931

2.  Comparison of contemporaneous left ventricular ejection fraction (LVEF) obtained from planar gated cardiac blood pool scans (GCBPS) and Tl-201 gated myocardial perfusion scans (MPS) using a novel solid state dedicated cardiac camera.

Authors:  Kenneth S K Yap; Martin Cherk; Bruce Van Every; Michael Bailey; Michael J Kelly; Victor Kalff
Journal:  J Nucl Cardiol       Date:  2013-02-27       Impact factor: 5.952

3.  Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance.

Authors:  Alban Bailliez; Tanguy Blaire; Frédéric Mouquet; R Legghe; B Etienne; Damien Legallois; Denis Agostini; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2014-05-09       Impact factor: 5.952

4.  Gated cardiac SPECT: can it be used for serial assessment of left ventricular function in oncology patients?

Authors:  Penelope J Thorley; Meg Bould; Fahmid U Chowdhury
Journal:  Nucl Med Commun       Date:  2013-12       Impact factor: 1.690