Literature DB >> 22541180

Sutureless perceval aortic valve replacement: results of two European centers.

Thierry A Folliguet1, François Laborde, Konstantinos Zannis, Gabriel Ghorayeb, Axel Haverich, Malakh Shrestha.   

Abstract

BACKGROUND: The Perceval S bioprosthesis (Sorin Biomedica Cardio Srl, Sallugia, Italy) is a self-expanding valve designed to preserve aortic sinuses and sinotubular junction. We report the midterm results of a prospective, multicenter clinical study evaluating the safety and efficacy of this stented bioprosthesis in patients undergoing aortic valve replacement with or without concomitant procedures.
METHODS: From January 2007 to September 2011, a total of 208 high-risk patients (mean European system for cardiac operative risk evaluation: 8.7±5.3 years) received a Perceval bioprosthesis in 2 European centers. Median follow up was 10±20 months and 100% complete, and the total accumulated follow-up was 156 patient-years. Ten patients have reached a 4-year follow-up. Valve function was assessed in all patients.
RESULTS: Valve implantation resulted in significant improvement of patients' symptoms. Mean preoperative and postoperative gradients were 48.6±18.6 mm Hg and 10.4±4.3 mm Hg, respectively, and preoperative and postoperative mean effective orifice areas were 0.7±0.2 and 1.4±0.4 cm2. Survival at 12 months was 87.1%, success of implantation was 95%, and freedom from reoperation was 96%. In hospital mortality was 2.4%. During follow-up, 9 patients (4%) required reoperation for paravalvular regurgitation; 7 early and 2 late reoperations. Mean cross-clamp time (CCT) and extracorporeal circulation time (ECT) were, respectively, 33±14 minutes and 54±24 minutes, including 45 patients who underwent surgery through ministernotomy. Concomitant coronary bypass was done in 48 patients with mean CCT 43±13 and ECT 68±25 minutes.
CONCLUSIONS: Perceval sutureless is a safe bioprosthesis that can easily be implanted, including by a minimally invasive technique. It provides excellent hemodynamic with significant clinical improvement. Overall, these data confirm the safety and utility of the Perceval bioprosthesis aortic valve replacement for high-risk patients.
Copyright © 2012 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22541180     DOI: 10.1016/j.athoracsur.2012.01.071

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  45 in total

1.  Minimally Invasive Sutureless Aortic Valve Replacement With the Perceval S Bioprosthesis Through Ministernotomy: A Single-Center Experience.

Authors:  Krishan Ramsaransing; Vikash Hindori; Athiná Kougioumtzoglou; Abdullah Kaya; Eva Verbeek
Journal:  Cureus       Date:  2020-10-28

2.  Do we need sutureless or self-anchoring aortic valve prostheses?

Authors:  Malakh Shrestha
Journal:  Ann Cardiothorac Surg       Date:  2015-03

3.  Left ventricular mass regression after sutureless implantation of the Perceval S aortic valve bioprosthesis: preliminary results.

Authors:  Giuseppe Santarpino; Steffen Pfeiffer; Francesco Pollari; Giovanni Concistrè; Ferdinand Vogt; Theodor Fischlein
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-10-08

Review 4.  Aortic valve replacement through J-shaped partial upper sternotomy.

Authors:  Shahzad G Raja; Umberto Benedetto; Mohamed Amrani
Journal:  J Thorac Dis       Date:  2013-11       Impact factor: 2.895

Review 5.  Sutureless aortic valve replacement: a systematic review and meta-analysis.

Authors:  Kevin Phan; Yi-Chin Tsai; Nithya Niranjan; Denis Bouchard; Thierry P Carrel; Otto E Dapunt; Harald C Eichstaedt; Theodor Fischlein; Borut Gersak; Mattia Glauber; Axel Haverich; Martin Misfeld; Peter J Oberwalder; Giuseppe Santarpino; Malakh Lal Shrestha; Marco Solinas; Marco Vola; Tristan D Yan; Marco Di Eusanio
Journal:  Ann Cardiothorac Surg       Date:  2015-03

6.  Sutureless aortic valve implantation: first experience in Asia.

Authors:  Guohao Chang; duc Thang Vu; Kristine Leok Kheng Teoh; Lian Kah Ti; Theodoros Kofidis
Journal:  Singapore Med J       Date:  2014-11       Impact factor: 1.858

Review 7.  Minimally invasive aortic valve surgery: state of the art and future directions.

Authors:  Mattia Glauber; Matteo Ferrarini; Antonio Miceli
Journal:  Ann Cardiothorac Surg       Date:  2015-01

8.  Early results of the Sorin® Perceval S sutureless valve: systematic review and meta-analysis.

Authors:  Karan Sian; Sheila Li; Daneish Selvakumar; Ross Mejia
Journal:  J Thorac Dis       Date:  2017-03       Impact factor: 2.895

Review 9.  [Surgical treatment of aortic valve stenosis].

Authors:  O J Liakopoulos; J Merkle; T Wahlers; Y-H Choi
Journal:  Herz       Date:  2017-09       Impact factor: 1.443

10.  Sutureless versus Conventional Aortic Valve Replacement: Outcomes in 70 High-Risk Patients Undergoing Concomitant Cardiac Procedures.

Authors:  Muhammet Onur Hanedan; Mehmet Ali Yuruk; Ali Ihsan Parlar; Ugur Ziyrek; Ali Kemal Arslan; Ufuk Sayar; Ilker Mataraci
Journal:  Tex Heart Inst J       Date:  2018-02-01