Literature DB >> 2250062

Failures of pin removal after in situ pinning of slipped capital femoral epiphyses: a comparison of different pin types.

E J Vresilovic1, K P Spindler, W W Robertson, R S Davidson, D S Drummond.   

Abstract

We reviewed 75 pin retrievals after in situ pinning of 35 children with slipped capital femoral epiphysis (SCFE). Four types of pins were used. Ten pins in 8 hips could not be removed because of pin breakage or stripping. Our failure rate was significantly related to pin type and size (p less than 0.039). As compared with the expected values using the chi-square method of analysis, the noncannulated large steel pins and cannulated steel pin groups had fewer failures (p less than 0.001) as compared with the cannulated titanium and noncannulated small steel pin groups. We now avoid using cannulated titanium or noncannulated small pins in treatment of SCFE.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2250062     DOI: 10.1097/01241398-199011000-00011

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  4 in total

1.  Complications and associated risk factors at screw removal in slipped capital femoral epiphysis treated by cannulated stainless steel screws.

Authors:  Juan Pretell-Mazzini; Victor Rodriguez-Vega; Jorge Muñoz-Ledesma; Eva María Andrés-Esteban; Rafael Marti-Ciruelos; José Luis González-López; Angel Curto-De la Mano
Journal:  J Child Orthop       Date:  2012-08-01       Impact factor: 1.548

2.  In vivo evaluation of Zr-based bulk metallic glass alloy intramedullary nails in rat femora.

Authors:  Kazuhiro Imai; Sachiko Hiromoto
Journal:  J Mater Sci Mater Med       Date:  2013-11-27       Impact factor: 3.896

3.  Locking Compression Plates are more difficult to remove than conventional non-locking plates.

Authors:  G D Musters; P Boele van Hensbroek; K J Ponsen; J S K Luitse; J C Goslings
Journal:  Eur J Trauma Emerg Surg       Date:  2013-01-11       Impact factor: 3.693

4.  In Vivo Evaluation of Bulk Metallic Glasses for Osteosynthesis Devices.

Authors:  Kazuhiro Imai; Sachiko Hiromoto
Journal:  Materials (Basel)       Date:  2016-08-09       Impact factor: 3.623

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.