Literature DB >> 22482615

High intensity focused ultrasound with large aperture transducers: a MRI based focal point correction for tissue heterogeneity.

Charles Mougenot1, Matti Tillander, Julius Koskela, Max O Köhler, Chrit Moonen, Mario Ries.   

Abstract

PURPOSE: The risk of undesired tissue damage to thoracic cage, heart, and lung during MR guided HIFU ablations of breast cancer can be greatly reduced if a phased array transducer design with a lateral beam direction is used in combination with a large aperture. The disadvantage is an increased sensitivity to focus aberrations due to tissue heterogeneity. Here, the authors propose to restore the focal coherence by using a matched aperture phase correction, which is based on a noninvasively obtained tissue model.
METHODS: The method combines high resolution MRI with ultrasound wave measurements of different tissue types to determine a phase correction, which compensates focal point aberrations caused by tissue heterogeneity. 3D segmentation of tissue is used to quantify the relative proportion of each tissue type along a line running from the center of each element of the phased array to the target focal point.
RESULTS: For tissue types with a celerity difference of 3%, the proposed method allows to quantify the phase aberration with an accuracy of 6° ± 20° and a correlation factor R(2) = 0.95. Using the refocusing method for a complex heterogeneous phantom resulted in 95% of the maximal pressure, whereas only 70% of the maximal pressure is obtained in absence of any phase correction.
CONCLUSIONS: Since the proposed refocusing algorithm is compatible with a standard interventional preplanning and requires only a minimal amount of processing, it presents a promising approach to compensate for aberration in heterogeneous tissues such as the human breast.

Entities:  

Mesh:

Year:  2012        PMID: 22482615     DOI: 10.1118/1.3693051

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  A simulation technique for 3D MR-guided acoustic radiation force imaging.

Authors:  Allison Payne; Josh de Bever; Alexis Farrer; Brittany Coats; Dennis L Parker; Douglas A Christensen
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

2.  Unsupervised clustering method to convert high-resolution magnetic resonance volumes to three-dimensional acoustic models for full-wave ultrasound simulations.

Authors:  Kevin Looby; Carl D Herickhoff; Christopher Sandino; Tao Zhang; Shreyas Vasanawala; Jeremy J Dahl
Journal:  J Med Imaging (Bellingham)       Date:  2019-07-22

3.  Breath-hold MR-HIFU hyperthermia: phantom and in vivo feasibility.

Authors:  Chenchen Bing; Bingbing Cheng; Robert M Staruch; Joris Nofiele; Michelle Wodzak Staruch; Debra Szczepanski; Alan Farrow-Gillespie; Adeline Yang; Theodore W Laetsch; Rajiv Chopra
Journal:  Int J Hyperthermia       Date:  2019       Impact factor: 3.914

4.  Non-surgical treatment of early breast cancer: techniques on the way.

Authors:  Alaa Hamza; Shymaa Elrefaey
Journal:  Gland Surg       Date:  2014-08

5.  Focal point determination in magnetic resonance-guided focused ultrasound using tracking coils.

Authors:  Bryant T Svedin; Michael J Beck; J Rock Hadley; Robb Merrill; Joshua T de Bever; Bradley D Bolster; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2016-07-15       Impact factor: 4.668

6.  Real-time implementation of a dual-mode ultrasound array system: in vivo results.

Authors:  Andrew J Casper; Dalong Liu; John R Ballard; Emad S Ebbini
Journal:  IEEE Trans Biomed Eng       Date:  2013-05-21       Impact factor: 4.538

Review 7.  High-intensity focused ultrasound: advances in technology and experimental trials support enhanced utility of focused ultrasound surgery in oncology.

Authors:  G Malietzis; L Monzon; J Hand; H Wasan; E Leen; M Abel; A Muhammad; P Price; P Abel
Journal:  Br J Radiol       Date:  2013-02-12       Impact factor: 3.039

8.  Phase aberration simulation study of MRgFUS breast treatments.

Authors:  Alexis I Farrer; Scott Almquist; Christopher R Dillon; Leigh A Neumayer; Dennis L Parker; Douglas A Christensen; Allison Payne
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

9.  Evaluation of a three-dimensional MR acoustic radiation force imaging pulse sequence using a novel unbalanced bipolar motion encoding gradient.

Authors:  Joshua T de Bever; Henrik Odéen; Nick Todd; Alexis I Farrer; Dennis L Parker
Journal:  Magn Reson Med       Date:  2015-10-07       Impact factor: 4.668

10.  A Breast-Specific MR Guided Focused Ultrasound Platform and Treatment Protocol: First-in-Human Technical Evaluation.

Authors:  Allison Payne; Robb Merrill; Emilee Minalga; J Rock Hadley; Henrik Odeen; Lorne W Hofstetter; Sara Johnson; Christine Tunon de Lara; Sophie Auriol; Stephanie Recco; Erik Dumont; Dennis L Parker; Jean Palussiere
Journal:  IEEE Trans Biomed Eng       Date:  2021-02-19       Impact factor: 4.538

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.