Pitfalls of electromagnetic tracking in clinical routine using multiple or adjacent sensors.

Ingmar Wegner1, Dogu Teber, Boris Hadaschik, Sascha Pahernik, Markus Hohenfellner, Hans-Peter Meinzer, Johannes Huber.   

Abstract

BACKGROUND: While electromagnetic tracking (EMT) holds great promise, there are substantiated concerns about interference within the clinical environment. The purpose of this study was to address accuracy and isolate pitfalls for using multiple or adjacent EMT sensors in clinical routine.
METHODS: A phantom simulating an EMT-guided puncture of the kidney was used to analyse the effects of multiple sensors in the direct vicinity, common clinical environments and the influence of endo-urological instruments.
RESULTS: No relevant interference due to the investigated instruments was discovered. However, there was a great decrease in accuracy in the vicinity of a C-arm's image intensifier, especially affecting the precision of two adjacent sensors.
CONCLUSION: Endo-urological instruments can be safely used for multi-sensor EMT-guided procedures. When avoiding the immediate proximity of the surgical table and the image intensifier, there is a comfortable accuracy corridor. Thus, EMT-assistance is promising for a wide range of procedures if basic conditions are met.
Copyright © 2012 John Wiley & Sons, Ltd.

Keywords:  accuracy analysis; electromagnetic tracking; navigation

Mesh:

Year:  2012        PMID: 22473654     DOI: 10.1002/rcs.1431

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  3 in total

1.  C-arm angle measurement with accelerometer for brachytherapy: an accuracy study.

Authors:  Thomas Wolff; Andras Lasso; Markus Eblenkamp; Erich Wintermantel; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-07-03       Impact factor: 2.924

2.  Electromagnetically tracked personalized templates for surgical navigation.

Authors:  Andrew W L Dickinson; Michelle L Zec; David R Pichora; Brian J Rasquinha; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-22       Impact factor: 2.924

3.  Electromagnetic tracking in surgical and interventional environments: usability study.

Authors:  Elodie Lugez; Hossein Sadjadi; David R Pichora; Randy E Ellis; Selim G Akl; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-09-06       Impact factor: 2.924

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.