Literature DB >> 22472797

Virtual monochromatic spectral imaging for the evaluation of hypovascular hepatic metastases: the optimal monochromatic level with fast kilovoltage switching dual-energy computed tomography.

Yoshitake Yamada1, Masahiro Jinzaki, Yutaka Tanami, Takayuki Abe, Sachio Kuribayashi.   

Abstract

OBJECTIVES: The purpose of this study was to select the optimal monochromatic level for virtual monochromatic spectral (VMS) imaging to minimize the image noise of the liver parenchyma and to acquire a high contrast-to-noise ratio (CNR) of hypovascular hepatic metastases in the portal-dominant phase.
MATERIALS AND METHODS: This study was conducted with the approval of our institutional review board, and written informed consent was obtained from all the participating patients. Ninety patients with hepatic metastases were scanned by fast kilovoltage switching dual-energy computed tomography in the portal-dominant phase. One hundred one sets of VMS images in the range of 40 to 140 keV at 1-keV intervals were reconstructed. The image noise of the liver parenchyma in each patient and the CNR of each metastasis (n = 303) were measured on all the 101 VMS image sets. Data were analyzed by the paired t test and mixed-effects model. Bonferroni's method was used for multiple comparisons.
RESULTS: The lowest noise of the liver parenchyma was obtained in 6, 15, 31, 29, 7, 1, and 1 patient at 67, 68, 69, 70, 71, 72, and 73 keV, respectively. The mean noise of the liver parenchyma on the 69-keV VMS images in all 90 patients was significantly lower than that on the 67-, 68-, 71-, 72-, and 73-keV VMS images (P < 0.001); however, there was no significant difference in the mean noise of the liver parenchyma between the 69-keV and 70-keV VMS images (P = 0.279). For 95% of the hepatic metastases (288/303 metastases), the highest metastasis-to-liver CNR was obtained in the 66- to 71- keV VMS images. The mean metastasis-to-liver CNR for the 303 metastases was numerically highest at 68 keV; however, there was no significant difference in the mean metastasis-to-liver CNR between the 68-keV and 69-keV images (P = 0.737) or between the 68-keV and 70-keV images (P = 0.103).
CONCLUSIONS: VMS imaging at approximately 70 keV (69-70 keV) yielded the lowest image noise of the liver parenchyma and a high CNR for hypovascular hepatic metastases in the portal-dominant phase.

Entities:  

Mesh:

Year:  2012        PMID: 22472797     DOI: 10.1097/RLI.0b013e318240a874

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  26 in total

Review 1.  Recent developments of dual-energy CT in oncology.

Authors:  David Simons; Marc Kachelriess; Heinz-Peter Schlemmer
Journal:  Eur Radiol       Date:  2014-01-09       Impact factor: 5.315

2.  A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions.

Authors:  Carlo N De Cecco; Damiano Caruso; U Joseph Schoepf; Domenico De Santis; Giuseppe Muscogiuri; Moritz H Albrecht; Felix G Meinel; Julian L Wichmann; Philip F Burchett; Akos Varga-Szemes; Douglas H Sheafor; Andrew D Hardie
Journal:  Eur Radiol       Date:  2018-02-19       Impact factor: 5.315

3.  Automatic spectral imaging protocol and iterative reconstruction for radiation dose reduction in typical hepatic hemangioma computed tomography with reduced iodine load: a preliminary study.

Authors:  Wei Li; Aiyin Li; Bin Wang; Xiuyuan Niu; Xin Cao; Xinyi Wang; Hao Shi
Journal:  Br J Radiol       Date:  2018-05-17       Impact factor: 3.039

4.  Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study.

Authors:  Sebastian Winklhofer; Jack W Lambert; Zhen Jane Wang; Yuxin Sun; Robert G Gould; Ronald J Zagoria; Benjamin M Yeh
Journal:  Abdom Radiol (NY)       Date:  2016-08

5.  Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience.

Authors:  Shun-Yu Gao; Xiao-Peng Zhang; Yong Cui; Ying-Shi Sun; Lei Tang; Xiao-Ting Li; Xiao-Yan Zhang; Jun Shan
Journal:  Chin J Cancer Res       Date:  2014-08       Impact factor: 5.087

6.  Spectral CT evaluation of interstitial brachytherapy in pancreatic carcinoma xenografts: preliminary animal experience.

Authors:  Shudong Hu; Wei Huang; Yerong Chen; Qi Song; Xiaozhu Lin; Zhongmin Wang; Kemin Chen
Journal:  Eur Radiol       Date:  2014-06-06       Impact factor: 5.315

7.  Development of a dual-energy computed tomography quality control program: Characterization of scanner response and definition of relevant parameters for a fast-kVp switching dual-energy computed tomography system.

Authors:  Jessica L Nute; Megan C Jacobsen; Wolfgang Stefan; Wei Wei; Dianna D Cody
Journal:  Med Phys       Date:  2018-03-15       Impact factor: 4.071

8.  Comparison of image quality and radiation exposure between conventional imaging and gemstone spectral imaging in abdominal CT examination.

Authors:  Tianqi Fang; Wei Deng; Martin Wai-Ming Law; Liangping Luo; Liyun Zheng; Ying Guo; Hanwei Chen; Bingsheng Huang
Journal:  Br J Radiol       Date:  2018-06-01       Impact factor: 3.039

Review 9.  Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Joel G Fletcher
Journal:  Radiology       Date:  2015-09       Impact factor: 11.105

10.  Differentiation of low-attenuation intracranial hemorrhage and calcification using dual-energy computed tomography in a phantom system.

Authors:  Jessica L Nute; Lucia Le Roux; Adam G Chandler; Veera Baladandayuthapani; Dawid Schellingerhout; Dianna D Cody
Journal:  Invest Radiol       Date:  2015-01       Impact factor: 6.016

View more