Development of a dual-energy computed tomography quality control program: Characterization of scanner response and definition of relevant parameters for a fast-kVp switching dual-energy computed tomography system.

Jessica L Nute1, Megan C Jacobsen1,2, Wolfgang Stefan1, Wei Wei3, Dianna D Cody1.   

Abstract

PURPOSE: A prototype QC phantom system and analysis process were developed to characterize the spectral capabilities of a fast kV-switching dual-energy computed tomography (DECT) scanner. This work addresses the current lack of quantitative oversight for this technology, with the goal of identifying relevant scan parameters and test metrics instrumental to the development of a dual-energy quality control (DEQC).
METHODS: A prototype elliptical phantom (effective diameter: 35 cm) was designed with multiple material inserts for DECT imaging. Inserts included tissue equivalent and material rods (including iodine and calcium at varying concentrations). The phantom was scanned on a fast kV-switching DECT system using 16 dual-energy acquisitions (CTDIvol range: 10.3-62 mGy) with varying pitch, rotation time, and tube current. The circular head phantom (22 cm diameter) was scanned using a similar protocol (12 acquisitions; CTDIvol range: 36.7-132.6 mGy). All acquisitions were reconstructed at 50, 70, 110, and 140 keV and using a water-iodine material basis pair. The images were evaluated for iodine quantification accuracy, stability of monoenergetic reconstruction CT number, noise, and positional constancy. Variance component analysis was used to identify technique parameters that drove deviations in test metrics. Variances were compared to thresholds derived from manufacturer tolerances to determine technique parameters that had a nominally significant effect on test metrics.
RESULTS: Iodine quantification error was largely unaffected by any of the technique parameters investigated. Monoenergetic HU stability was found to be affected by mAs, with a threshold under which spectral separation was unsuccessful, diminishing the utility of DECT imaging. Noise was found to be affected by CTDIvol in the DEQC body phantom, and CTDIvol and mA in the DEQC head phantom. Positional constancy was found to be affected by mAs in the DEQC body phantom and mA in the DEQC head phantom.
CONCLUSION: A streamlined scan protocol was developed to further investigate the effects of CTDIvol and rotation time while limiting data collection to the DEQC body phantom. Further data collection will be pursued to determine baseline values and statistically based failure thresholds for the validation of long-term DECT scanner performance.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  computed tomography; dual-energy CT; quality control; spectral imaging

Mesh:

Year:  2018        PMID: 29446082      PMCID: PMC5903965          DOI: 10.1002/mp.12812

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  34 in total

1.  Centiles for adult head circumference.

Authors:  K M Bushby; T Cole; J N Matthews; J A Goodship
Journal:  Arch Dis Child       Date:  1992-10       Impact factor: 3.791

Review 2.  Dual-energy CT of the urinary tract.

Authors:  Ravi K Kaza; Joel F Platt; Alec J Megibow
Journal:  Abdom Imaging       Date:  2013-02

3.  Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode.

Authors:  Da Zhang; Xinhua Li; Bob Liu
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

4.  Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy.

Authors:  Sebastian Feuerlein; Tobias J Heye; Mustafa R Bashir; Daniel T Boll
Journal:  Invest Radiol       Date:  2012-11       Impact factor: 6.016

5.  Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter: Assessment of Image Quality, Radiation Dose, and Accuracy of Dual-Energy Applications in an In Vitro and In Vivo Study.

Authors:  André Euler; Anushri Parakh; Anna L Falkowski; Sebastian Manneck; David Dashti; Bernhard Krauss; Zsolt Szucs-Farkas; Sebastian T Schindera
Journal:  Invest Radiol       Date:  2016-08       Impact factor: 6.016

6.  Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT.

Authors:  Kazuhiro Matsumoto; Masahiro Jinzaki; Yutaka Tanami; Akihisa Ueno; Minoru Yamada; Sachio Kuribayashi
Journal:  Radiology       Date:  2011-02-17       Impact factor: 11.105

7.  Metal artifact reduction by dual-energy computed tomography using energetic extrapolation: a systematically optimized protocol.

Authors:  Felix G Meinel; Bernhard Bischoff; Qiaowei Zhang; Fabian Bamberg; Maximilian F Reiser; Thorsten R C Johnson
Journal:  Invest Radiol       Date:  2012-07       Impact factor: 6.016

8.  Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.

Authors:  Fabian Bamberg; Alexander Dierks; Konstantin Nikolaou; Maximilian F Reiser; Christoph R Becker; Thorsten R C Johnson
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

9.  Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures.

Authors:  Changsheng Zhou; Yan E Zhao; Song Luo; Hongyuan Shi; Lin Li; Ling Zheng; Long Jiang Zhang; Guangming Lu
Journal:  Acad Radiol       Date:  2011-10       Impact factor: 3.173

10.  Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT.

Authors:  Gert Jan Pelgrim; Robbert W van Hamersvelt; Martin J Willemink; Bernhard T Schmidt; Thomas Flohr; Arnold Schilham; Julien Milles; Matthijs Oudkerk; Tim Leiner; Rozemarijn Vliegenthart
Journal:  Eur Radiol       Date:  2017-02-06       Impact factor: 5.315

View more
  6 in total

1.  Deep-learning-based direct inversion for material decomposition.

Authors:  Hao Gong; Shengzhen Tao; Kishore Rajendran; Wei Zhou; Cynthia H McCollough; Shuai Leng
Journal:  Med Phys       Date:  2020-10-30       Impact factor: 4.071

2.  Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study.

Authors:  Xiaomei Lu; Zaiming Lu; Jiandong Yin; Yuying Gao; Xingbiao Chen; Qiyong Guo
Journal:  Quant Imaging Med Surg       Date:  2019-02

3.  Quantification and homogenization of image noise between two CT scanner models.

Authors:  Samuel A Einstein; Xiujiang John Rong; Corey T Jensen; Xinming Liu
Journal:  J Appl Clin Med Phys       Date:  2019-12-20       Impact factor: 2.102

4.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

5.  Characterization of size-specific effects during dual-energy CT material decomposition of non-iodine materials.

Authors:  Jessica Miller; Lianna DiMaso; Jessie Huang-Vredevoogd; Jainil Shah; Michael Lawless
Journal:  J Appl Clin Med Phys       Date:  2021-11-16       Impact factor: 2.102

6.  A suggested method for setting up GSI profiles on the GE Revolution CT scanner.

Authors:  David M Gauntt
Journal:  J Appl Clin Med Phys       Date:  2019-11-05       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.