Standardization of fetal ultrasound biometry measurements: improving the quality and consistency of measurements.

I Sarris1, C Ioannou, M Dighe, A Mitidieri, M Oberto, W Qingqing, J Shah, S Sohoni, W Al Zidjali, L Hoch, D G Altman, A T Papageorghiou.   

Abstract

OBJECTIVE: To assess whether a standardization exercise prior to commencing a fetal growth study involving multiple sonographers can reduce interobserver variation.
METHODS: In preparation for an international study assessing fetal growth, nine experienced sonographers from eight countries participated in a standardization exercise consisting of theoretical and practical sessions. Each performed a set of seven standard fetal measurements on pregnant volunteers at 20-37 weeks' gestation, and these were repeated by the lead sonographer; all measurements were taken in a blinded fashion. After this the sonographers had hands-on practice and feedback sessions on other volunteers. This process was repeated three times. Measurement differences between sonographers and the lead sonographer, expressed as a gestational-age-specific Z-score, between the first and third scans were compared using the Wilcoxon signed ranks test, and variance was assessed using Pitman's test. Interobserver agreement was also assessed using the intraclass correlation coefficient (ICC), and all images were scored for quality in a blinded fashion.
RESULTS: At baseline the level of agreement and image scoring were high. A significant reduction in the differences between sonographers and the lead sonographer were seen for fetal biometry overall (head circumference, abdominal circumference and femur length) between the first and third scans (median Z-scores, 0.46 and 0.24; P = 0.005), and a reduction in the variance was also observed (P < 0.001). The ICCs for measurement pairs for every fetal measurement showed a clear trend of increasing ICC (better agreement) with consecutive training scan sessions, although no improvement in image scores was seen.
CONCLUSION: Even for experienced sonographers, a standardization exercise before starting a study of fetal biometry can improve consistency of measurements. This could be of relevance for studies assessing fetal growth in multicenter sites.
Copyright © 2011 ISUOG. Published by John Wiley & Sons, Ltd.

Mesh:

Year:  2011        PMID: 22411446     DOI: 10.1002/uog.8997

Source DB:  PubMed          Journal:  Ultrasound Obstet Gynecol        ISSN: 0960-7692            Impact factor:   7.299


  20 in total

1.  Echocardiographic measurement methods for left ventricular linear dimensions in children result in predictable variations in results.

Authors:  Shelby Kutty; David Russell; Ling Li; Rimsha Hasan; Qinghai Peng; Peter C Frommelt; David A Danford
Journal:  Int J Cardiovasc Imaging       Date:  2013-12-10       Impact factor: 2.357

2.  Intraobserver variability in fibroid size measurements: estimated effects on assessing fibroid growth.

Authors:  Malana Moshesh; Shyamal D Peddada; Tracy Cooper; Donna Baird
Journal:  J Ultrasound Med       Date:  2014-07       Impact factor: 2.153

3.  Automatic image quality assessment and measurement of fetal head in two-dimensional ultrasound image.

Authors:  Lei Zhang; Nicholas J Dudley; Tryphon Lambrou; Nigel Allinson; Xujiong Ye
Journal:  J Med Imaging (Bellingham)       Date:  2017-04-17

4.  Toward point-of-care ultrasound estimation of fetal gestational age from the trans-cerebellar diameter using CNN-based ultrasound image analysis.

Authors:  Mohammad A Maraci; Mohammad Yaqub; Rachel Craik; Sridevi Beriwal; Alice Self; Peter von Dadelszen; Aris Papageorghiou; J Alison Noble
Journal:  J Med Imaging (Bellingham)       Date:  2020-01-13

5.  Are ultrasound foetal circumference measurement methods interchangeable?

Authors:  Nicholas John Dudley
Journal:  Ultrasound       Date:  2019-03-07

6.  Improving precision of second-trimester biometry.

Authors:  Christy Loiacono; Michael Bethune; Michal Schneider; Paul Lombardo
Journal:  Australas J Ultrasound Med       Date:  2018-04-25

7.  A comparison of ultrasound with magnetic resonance imaging in the assessment of fetal biometry and weight in the second trimester of pregnancy: An observer agreement and variability study.

Authors:  Jacqueline Matthew; Christina Malamateniou; Caroline L Knight; Kelly P Baruteau; Tara Fletcher; Alice Davidson; Laura McCabe; Dharmintra Pasupathy; Mary Rutherford
Journal:  Ultrasound       Date:  2018-01-29

8.  Caveats in the monitoring of fetal growth using ultrasound estimated fetal weight.

Authors:  Nicholas John Dudley; Helen Varley
Journal:  Ultrasound       Date:  2020-09-11

9.  Measurement of abdominal circumference in preterm infants.

Authors:  Ilze Meldere; Valdis Urtans; Aigars Petersons; Zane Abola
Journal:  BMC Res Notes       Date:  2015-11-26

10.  WHO multicentre study for the development of growth standards from fetal life to childhood: the fetal component.

Authors:  Mario Merialdi; Mariana Widmer; Ahmet Metin Gülmezoglu; Hany Abdel-Aleem; George Bega; Alexandra Benachi; Guillermo Carroli; Jose Guilherme Cecatti; Anke Diemert; Rogelio Gonzalez; Kurt Hecher; Lisa N Jensen; Synnøve L Johnsen; Torvid Kiserud; Alka Kriplani; Pisake Lumbiganon; Ann Tabor; Sameera A Talegawkar; Antoinette Tshefu; Daniel Wojdyla; Lawrence Platt
Journal:  BMC Pregnancy Childbirth       Date:  2014-05-02       Impact factor: 3.007

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.