Literature DB >> 22401326

Effect of disinfectants on glucose monitors.

John J Mahoney1, Christine G Lim.   

Abstract

BACKGROUND: Monitoring blood glucose levels is an integral part of routine diabetes management. To minimize the risk of transmission of bloodborne pathogens during monitoring, the Centers for Disease Control and Prevention (CDC) recommends that glucose meters be disinfected after each use whenever they are used to test multiple patients. The objective of this study is to assess the compatibility of some common disinfectants with certain blood glucose meter systems.
METHODS: We tested six disinfectants for adverse impact on meter performance or the exterior meter surfaces. The disinfectants tested were 0.525% sodium hypochlorite, 20% 2-propanol and 10% ethanol, 17.2% isopropanol, 55% isopropanol, 70% isopropanol, and hydrogen peroxide. To assess meter performance, we tested OneTouch® Ultra® blood glucose monitoring systems with control solution before and after application of either water or disinfectant. To assess the effect on exterior meter surfaces, we performed a soaking test to simulate long-term exposure to disinfectant.
RESULTS: Paired t-test results showed that the control solution data associated with disinfectant and with water application were not significantly different for each meter type. However, most of the meter types were adversely affected by hydrogen peroxide and/or by the higher concentrations of alcohol-based disinfectants.
CONCLUSIONS: Although none of the six disinfectants affected meter performance, hydrogen peroxide and isopropanol >20% adversely affected the exterior surfaces of the tested meters. When complying with CDC instructions for meter disinfection, users should use caution and choose disinfectants that have been validated by the meter manufacturer.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22401326      PMCID: PMC3320825          DOI: 10.1177/193229681200600111

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  12 in total

1.  Evaluation of the potential for bloodborne pathogen transmission associated with diabetes care practices in nursing homes and assisted living facilities, Pinellas County.

Authors:  Nicola D Thompson; Vaughn Barry; Karen Alelis; Dongming Cui; Joseph F Perz
Journal:  J Am Geriatr Soc       Date:  2010-03-30       Impact factor: 5.562

2.  Nosocomial transmission of hepatitis B virus associated with a spring-loaded fingerstick device--California.

Authors: 
Journal:  MMWR Morb Mortal Wkly Rep       Date:  1990-09-07       Impact factor: 17.586

3.  Assisted monitoring of blood glucose: special safety needs for a new paradigm in testing glucose.

Authors:  David C Klonoff; Joseph F Perz
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

4.  Interference of Accel wipes with LifeScan SureStep Flexx glucose meters.

Authors:  Cindy L Larsen; Carie Jackson; Martha E Lyon
Journal:  Clin Biochem       Date:  2005-12-06       Impact factor: 3.281

5.  Survival of hepatitis B virus after drying and storage for one week.

Authors:  W W Bond; M S Favero; N J Petersen; C R Gravelle; J W Ebert; J E Maynard
Journal:  Lancet       Date:  1981-03-07       Impact factor: 79.321

6.  Disinfectant wipes containing hydrogen peroxide induce overestimation of glucose results obtained with Lifescan SureStep Flexx® glucose meter.

Authors:  Philippe Desmeules; Johanne Ethier; Pierre Allard
Journal:  Clin Biochem       Date:  2010-09-29       Impact factor: 3.281

7.  Nosocomial hepatitis B virus infection associated with reusable fingerstick blood sampling devices--Ohio and New York City, 1996.

Authors: 
Journal:  MMWR Morb Mortal Wkly Rep       Date:  1997-03-14       Impact factor: 17.586

8.  Deja vu: nosocomial hepatitis B virus transmission and fingerstick monitoring.

Authors:  J M Quale; D Landman; B Wallace; E Atwood; V Ditore; G Fruchter
Journal:  Am J Med       Date:  1998-10       Impact factor: 4.965

9.  Nosocomial transmission of hepatitis B virus associated with the use of a spring-loaded finger-stick device.

Authors:  L B Polish; C N Shapiro; F Bauer; P Klotz; P Ginier; R R Roberto; H S Margolis; M J Alter
Journal:  N Engl J Med       Date:  1992-03-12       Impact factor: 91.245

10.  Nosocomial transmission of hepatitis B virus infection among residents with diabetes in a skilled nursing facility.

Authors:  Amy J Khan; Suzanne M Cotter; Beth Schulz; Xiaolei Hu; Jon Rosenberg; Betty H Robertson; Anthony E Fiore; Beth P Bell
Journal:  Infect Control Hosp Epidemiol       Date:  2002-06       Impact factor: 3.254

View more
  1 in total

1.  Demonstration of disinfection procedure for the development of accurate blood glucose meters in accordance with ISO 15197:2013.

Authors:  Shu-Ping Lin; Wen-Ye Lin; Jung-Tzu Chang; Chun-Feng Chu
Journal:  PLoS One       Date:  2017-07-06       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.