Interference of Accel wipes with LifeScan SureStep Flexx glucose meters.

Cindy L Larsen1, Carie Jackson, Martha E Lyon.   

Abstract

OBJECTIVES: To investigate the cause of a persistent test strip error code, error 2b, appearing on LifeScan SureStep Flexx glucose meters (Johnson and Johnson, Rochester, USA). DESIGN AND METHODS: SaniCloth Plus (The Dynarex Corporation, Orangeburg, USA), CaviWipes XL (Metrex, Orange, USA), and Accel (HANSA Med Ltd., Mississauga, Canada) sanitizing agents were tested in an attempt to reproduce the error.
RESULTS: The error 2b code was encountered when the Accel was used as a cleaning agent, but not when SaniCloth Plus or CaviWipes XL were used.
CONCLUSIONS: The hydrogen peroxide contained in Accel (HANSAmed, Ltd., Mississauga, Canada) interferes with the glucose oxidase reaction on the test strip, causing the error 2b code.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16337173     DOI: 10.1016/j.clinbiochem.2005.11.003

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  3 in total

1.  Effect of disinfectants on glucose monitors.

Authors:  John J Mahoney; Christine G Lim
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  Evaluation of different disinfectants on the performance of an on-meter dosed amperometric glucose-oxidase-based glucose meter.

Authors:  Don Sarmaga; Jeffrey A Dubois; Martha E Lyon
Journal:  J Diabetes Sci Technol       Date:  2011-11-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.