Severe corrosion after malpositioning of a metallic head over the Morse taper of a cementless hip arthroplasty. A case report.

E Pansard1, N Fouilleron, G Dereudre, H Migaud, J Girard.   

Abstract

Morse tapers are frequently used in total hip replacement to achieve precise adjustment of lengths and femoral offset. Mechanically, they do not raise any specific problems so long as strict positioning requirements are observed and elements from different manufacturers are not mixed together. We report a case in which the implant induced unexplained pain at 2 years, in relation to a defective fit between the metallic head and the Morse taper. Asymmetric partial fit of the head onto the taper was detected on control X-ray and was implicated as causing metallosis due to excessive release of metal debris from the Morse taper. Revision required femoral stem exchange because of the damage to the Morse taper as well as replacing the cup with new metal-metal bearings. Evolution was favorable at 3 years' follow-up. Most hip replacements include a Morse taper; the present clinical case is a reminder that strict positioning rules are to be respected, without which corrosion and wear may lead to mechanical failure.
Copyright © 2012 Elsevier Masson SAS. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22386704     DOI: 10.1016/j.otsr.2011.05.018

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  9 in total

Review 1.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

Review 2.  The surgical options and clinical evidence for treatment of wear or corrosion occurring with THA or TKA.

Authors:  Charles A Engh; Henry Ho; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

3.  Characterization of Femoral Head Taper Corrosion Features Using a 22-Year Retrieval Database.

Authors:  Jacob Cartner; Patrick Aldinger; Chenxi Li; David Collins
Journal:  HSS J       Date:  2016-08-12

4.  Head-stem trunnion dissociation due to corrosion in total hip arthroplasty.

Authors:  Michael R Whitehouse
Journal:  Ann Transl Med       Date:  2017-09

5.  Direct anterior approach and dual mobility cup: the "head-first" reduction technique: a technical note and early case series.

Authors:  F Randelli; A Fioruzzi; L Scaltrito; M Brioschi; O R Ayeni; P S Randelli
Journal:  Musculoskelet Surg       Date:  2020-01-28

Review 6.  Trunnionosis in total hip arthroplasty: a review.

Authors:  Jaydev B Mistry; Morad Chughtai; Randa K Elmallah; Aloise Diedrich; Sidney Le; Melbin Thomas; Michael A Mont
Journal:  J Orthop Traumatol       Date:  2016-02-11

7.  Bilateral Mechanically-Assisted Crevice Corrosion Resulting in Femoral Stem-Head Dissociation in Metal-on-Polyethylene Total Hip Arthroplasty.

Authors:  Timothy McAleese; Iain Feeley; Andrew Hughes; Eoin Sheehan; Khalid Merghani; Dorothy Niall
Journal:  Arthroplast Today       Date:  2021-02-26

8.  Dominant role of molybdenum in the electrochemical deposition of biological macromolecules on metallic surfaces.

Authors:  Elizabeth J Martin; Robin Pourzal; Mathew T Mathew; Kenneth R Shull
Journal:  Langmuir       Date:  2013-04-03       Impact factor: 3.882

9.  Spontaneous Dissociation of Anatomic Medullary Locking A Plus (AML A Plus) Femoral Component at the Head-Neck Interface.

Authors:  Ketan Pande; Juzaily Fekry Leong; Ngai Nung Lo
Journal:  J Orthop Case Rep       Date:  2015 Jul-Sep
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.