The corrosion of CoCrMo, an alloy frequently used in orthopedic implants, was studied with an electrochemical quartz crystal microbalance (QCM) in three physiologically relevant solutions. Mass changes were measured during potentiodynamic tests, showing material deposition in protein solutions at potential levels that caused mass loss when the proteins were not present. X-ray photoelectron spectroscopy (XPS) data indicated that the deposited material was primarily organic and therefore was most likely derived from proteins in the electrolyte. Material deposition consistently occurred at a critical potential and was not dependent on the current density or total charge released into solution. Corrosion studies on pure Co, Cr, and Mo in protein solutions found material deposition only on Mo. We hypothesize that organic deposition results from the interaction of Mo(VI) with proteins in the surrounding solution. The organic layer is reminiscent of tribochemical reaction layers that form on the surface of CoCrMo hip bearings, suggesting that these types of layers can be formed by purely electrochemical means.
The corrosion of CoCrMo, an alloy frequently used in orthopedic implants, was studied with an electrochemical quartz crystal microbalance (QCM) in three physiologically relevant solutions. Mass changes were measured during potentiodynamic tests, showing material deposition in protein solutions at potential levels that caused mass loss when the proteins were not present. X-ray photoelectron spectroscopy (XPS) data indicated that the deposited material was primarily organic and therefore was most likely derived from proteins in the electrolyte. Material deposition consistently occurred at a critical potential and was not dependent on the current density or total charge released into solution. Corrosion studies on pure Co, Cr, and Mo in protein solutions found material deposition only on Mo. We hypothesize that organic deposition results from the interaction of Mo(VI) with proteins in the surrounding solution. The organic layer is reminiscent of tribochemical reaction layers that form on the surface of CoCrMo hip bearings, suggesting that these types of layers can be formed by purely electrochemical means.
Recently, serious concerns
have been reported regarding the mechanical
performance and safety of metal biomedical implants.[1−4] Among the many factors that can affect implant performance, the
corrosive nature of the physiological environment plays a significant
role. Metal-on-metal (MoM) hip replacements, typically made of CoCrMo
alloys, degrade because of the synergistic effects of mechanical wear
from metal–metal contact and electrochemical reactions with
salts, proteins, and other organic species.[5,6] These
degradative processes release metal ions and nanoscopic wear particles
in vivo that often contribute to implant failure.[7,8] However,
studies have shown that a tribofilm, composed of organic and metallic
constituents, can form under these same conditions on the surface
of CoCrMo implants.[8] Implants with such
a tribofilm have demonstrated increased corrosion resistance; the
film may also act as a lubrication layer, enabling better wear resistance.[9] The organic constituents of this tribofilm are
presumably derived from proteins in synovial fluid, but mechanisms
causing the protein decomposition necessary for film formation are
unknown. The current theory suggests that frictional processes contribute
significantly to its development, with the combination of shear stresses
and a rise in local contact temperatures causing protein degradation.[8] Because it has been shown that tribofilms hold
Co, Cr, and Mo ions,[10] it is additionally
hypothesized that local variations in corrosion kinetics involving
proteins and metal ions contribute to the film formation.In
the study described here, we combined standard electrochemical
techniques with a quartz crystal microbalance (QCM) mass sensor to
delineate corrosion mechanisms of CoCrMo and to investigate if tribofilm-like
layers can be generated electrochemically. The QCM enables in situ
quantification of surface changes accompanying corrosion and is very
sensitive to small changes in the total mass of a thin film.[11−15] Here, it was used to measure mass changes associated with the different
sublayers of the alloy surface, including the metal, metal oxide,
and any organic deposit, as depicted in the schematic in Figure 1. The instrument contains a thin piezoelectric quartz
crystal set between two electrodes. When an oscillating voltage is
applied across the electrodes, the quartz vibrates and its resonance
frequency can be measured. If the mass on the electrode surface changes,
so will the resonance frequency, providing detailed information that
quantifies mass changes down to the nanogram level.[16] The QCM is also sensitive to viscoelastic properties that
introduce dissipation and reduce the quality factor of the resonance.[17] Although a variety of equivalent measures can
be used to quantify this dissipation, we use the half-width at half-maximum
of the conductance peak centered at the resonance frequency. The use
of this quantity, which we refer to as Γ, has been used extensively
by others and is discussed in more detail in a review article by Johannsmann.[18]
Figure 1
Schematic representation of a metal surface under oxidizing
conditions
in the presence of macromolecules that are able to form an organic
deposit at the metal/solution interface.
Schematic representation of a metal surface under oxidizing
conditions
in the presence of macromolecules that are able to form an organic
deposit at the metal/solution interface.The QCM has been used to study a variety of corrosion processes
in different materials, including Fe–Cr and Au–Pd–In
alloys.[11−13] Recently, Valero Vidal et al.[19] employed an electrochemical QCM to investigate passive
film growth on CoCrMo films in phosphate-buffered saline (PBS). Mass
loss was observed in all anodic potential ranges investigated, even
during oxide growth, but was accelerated upon transpassive dissolution.
In the present study, the corrosion behavior of CoCrMo thin films
was measured in three different fluids that mimic different aspects
of the physiological environment. Electrochemical QCM measurements
were employed to quantify mass changes associated with the corrosion
process. In protein solutions, it was found that an organic layer
deposits on the surface of the film at elevated potentials. A series
of potentiodynamic and potentiostatic tests indicated critical criteria
for deposition of this organic layer. Additionally, similar corrosion
tests were conducted on thin films of the individual alloying elements
(Co, Cr, and Mo) to determine if a particular ion is responsible for
the organic deposition. As we describe in the following sections,
it was found that a significantly thick organic film formed on the
Mo films, suggesting that Mo ions are essential to promoting organic
film formation on the alloy.
Experimental
Methods
Sample Preparation
Working electrode
specimens were prepared on AT-cut, 5 MHz quartz crystals with chrome
electrodes (Inficon, East Syracuse, NY). CoCrMo, Cr, and Mo films
were fabricated in a magnetron sputter deposition chamber (AJA Orion,
N. Scituate, MA) at a base pressure of 10–7 Torr
with Ar as the working gas. CoCrMo was deposited with a multistep
sputter method modified from Nakagawa et al., which was designed to
control the film’s crystallinity and internal stresses.[20] First, the chamber was heated to 300 C at a
rate of 14 C/min; crystals were then cleaned with a plasma strike.
Subsequently, Cr was sputtered for 200 s at 100 W power and 3 mTorr
pressure to produce an adhesion layer. Using a wrought, low-carbonCoCrMo target (ASTM F1537), a CoCrMo underlayer was deposited by sputtering
for 100 s at 180 W power and 0.25 mTorr. Finally, CoCrMo was sputtered
for 2000 s at 150 W power and 4 mTorr pressure. Cr films were made
by sputtering at 300 C and 3 mTorr pressure for 3000 s at 180 W power
following the plasma cleaning step. Mo films were prepared at room
temperature. Following the plasma strike, a Cr adhesion layer was
deposited by sputtering for 200 s at 100 W power; Mo was then sputtered
for 2000 s at 180 W power. Co films were prepared using an Edwards
Auto 500 electron beam evaporator (BOC Edwards, Wilmington, MA) evacuated
to a base pressure of 10–5 Torr. A Cr adhesion layer
(5 nm thick) was first deposited at a rate of 0.08 nm/s at 40 mA beam
current. Co (International Advanced Materials, Spring Valley, NY)
was deposited at the same rate at 100 mA beam current to a thickness
of between 130 and 300 nm. Low-carbonCoCrMo pins (ASTM F1537) with
a diameter of 7 mm were utilized for bulk specimen tests. Prior to
testing, pins were ground and polished with standard metallographic
methods to a mirror finish (Ra < 0.02 μm).Schematic
of the electrochemical QCM corrosion cell.
Electrochemical QCM
Electrochemical
tests were conducted in a custom three-electrode corrosion cell. The
sputtered metal films on the quartz substrate acted as the working
electrode, with an exposed area of 2.85 cm2. Two electrical
leads connected the QCM holder to a QCM network analyzer and the potentiostat
(BioLogic SP-150, France), as shown schematically in Figure 2. Mammalian Ringer’s salt solution (Electron
Microscopy Services, Hatfield, PA), buffer solution (pH 7.6), or bovinecalf serum (BCS) served as the electrolyte. BCS is commonly used as
a testing fluid for wear tests of orthopedic bearings. Its protein
content is comparable to that of natural joint fluid (synovial fluid)
with 30 g/L protein. Ringer’s solution was used as manufactured
and consisted of NaCl, KCl, MgCl2, CaCl2, NaHCO3, and glucose. Buffer solution was prepared with 0.154 M NaCl,
0.223 M Tris buffer, and 53.7 mM EDTA, adjusted to pH 7.6 with HCl.
BCS was prepared by diluting newborn calf serum (Invitrogen, Grand
Island, NY) in buffer solution.
Figure 2
Schematic
of the electrochemical QCM corrosion cell.
Prior to each experiment, the
open circuit potential was measured for 3 min by allowing the potential
at the working electrode to equilibrate in the electrolyte. Next,
the specimen was held at a cathodic potential of −0.9 V for
10 min to clean the surface and remove any oxide film that may have
formed in air. The open circuit potential was then measured after
the system was allowed to stabilize for 1 h. Subsequently, potentiodynamic
and potentiostatic tests were conducted under varying conditions while
the QCM frequency response was measured at the first, third, and fifth
harmonics (n = 1, 5 MHz; =3, 15 MHz; =5, 25 MHz). In many cases, QCM data
from only the third harmonic are reported for clarity, but the data
showed good agreement among all three harmonics. As discussed in more
detail below, the experiments are generally conducted in the range
where a simple analysis based on only a single harmonic is sufficient.
During potentiodynamic tests, the potential was increased at a rate
of 0.25 mV/s from −0.8 to 1.0 V. All potentials are reported
against a saturated calomel reference electrode (SCE). The corrosion
potential (Ecorr) and current (icorr) were extracted from the polarization curves
by the Tafel method[21] conducted with EC-Lab
Express software. Tests were conducted at room temperature.QCM frequency shifts were related to the mass change per unit area
(ΔM) on the working
electrode surface with the Sauerbrey equation[22]where Zq is the theoretical acoustic
impedance of quartz (8.84 ×
106 kg m–2 s–1), n is the order of the measured harmonic (1, 3, or 5), and f1 is the fundamental frequency (5 MHz). The
baseline frequency for a given harmonic corresponds to the resonance
frequency of the crystal at the outset of a potentiometric scan after
the system equilibrated for 1 h and the open circuit potential was
measured. The Sauerbrey equation was originally applied to quartz
resonators in air or vacuum, but it has been subsequently shown that
the analysis can also be applied to mass deposition from a liquid
medium. The analysis holds in a liquid medium provided that the reference
frequency corresponds to the resonance frequency of the immersed crystal
prior to the deposition process and that the dissipation (Γ)
due to viscoelastic effects is not very high.[23] The thickness of the metal film prior to testing was also extracted
from eq 1 by measuring the resonance frequency
before and after film deposition and by assuming that the sputtered
density is the same as for the bulk material.
X-ray
Photoelectron Spectroscopy and Energy-Dispersive
Spectroscopy
X-ray photoelectron spectroscopy (XPS) was conducted
with a Thermo Scientific ESCALAB 250Xi spectrometer (Thermo Fisher
Scientific) using a monochromatized Al Kα X-ray source at a
90° takeoff angle. The analysis spot size had a diameter of 500
μm. The base pressure in the analysis chamber was ∼10–9 mbar. Survey spectra (0–1100 eV) were obtained
at a 100 eV analyzer pass energy with a 1.0 eV energy step size. High-resolution
spectra of C 1s, O 1s, N 1s, Co 2p, Cr 2p, and Mo 3d were obtained
at a 50 eV analyzer pass energy with a 0.5 eV energy step size. A
flood gun was employed to control the surface potential throughout
spectra acquisition. Peak fitting was conducted with Avantage software
(Thermo Scientific) using a Shirley-type background and Gaussian/Lorentzian
peak shapes. Binding energies were referenced by shifting the lowest-energy
C 1s peak (hydrocarbon peak) to 284.8 eV. Analysis was performed on
three areas for each sample. Five C 1s and O 1s and 10–20 N
1s, Co 2p, Cr 2p, and Mo 3d high-resolution scans were conducted for
each area. Peak binding energies and atomic percentages are reported
as an average ± one standard deviation.Energy-dispersive
spectroscopy (EDS) was conducted with a Hitachi S-3400-NII scanning
electron microscope equipped with an Oxford Incax-act SDD energy dispersive
X-ray spectrometer. Analysis was performed at 15–20 keV with
a 90 s acquisition time.
Results and Discussion
Characterization of the CoCrMo Thin Film
QCM measurements
indicated that sputtered CoCrMo films ranged in
thickness from 200 to 300 nm. EDS confirmed that the chemical composition
matched that of the bulk alloy, with approximately 66% Co, 28% Cr,
and 6% Mo by weight. Crystallinity was confirmed with transmission
electron microscopy (TEM) in another study[24] that showed the film to have longitudinal, nanocrystalline grains
and a hexagonal close-packed (hcp) lattice. Although the microstructure
differs from that of the bulk alloy, it is comparable to the deformed
subsurface of MoM hip joints, which also exhibits a nanocrystalline
hcp structure resulting from a strain-induced phase transformation.[25] To determine how the electrochemical behavior
of the sputtered CoCrMo thin film compared to that of the bulk alloy,
potentiodynamic tests were conducted on both specimens in a BCS electrolyte.
As displayed in Figure 3, both materials exhibit
similar electrochemical behavior, which is consistent with the reported
literature data.[19,26,27]Ecorr and icorr of the bulk sample were −0.463 V and 1.85 × 10–7 A/cm2, respectively, and Ecorr and icorr of the thin films ranged from
−0.585 to −0.318 V and 3.79 × 10–8 to 1.82 × 10–7 A/cm2, depending
on the sample. The qualitative features of the potentiodynamic scan
for the bulk material can be used to divide the response into the
following regimes: (1) a cathodic regime below ∼−0.6
V in which the current is determined by the reduction of water or
dissolved oxygen; (2) a passive regime, extending from ∼−0.2
to ∼0.5 V, where the current is nearly independent of the applied
potential, indicating the presence of a protective oxide; and (3)
a transpassive regime above the breakdown potential at ∼0.5
V. The current peak and subsequent stabilization from 0.65 to 0.8
V is typically attributed to the oxidation of Cr(III) to Cr(VI), although
the electrochemical behavior in this potential regime is poorly understood.[26−29]
Figure 3
Comparison
of electrochemical behavior of bulk and thin film CoCrMo.
(Scan rate 0.25 mV/s.)
Comparison
of electrochemical behavior of bulk and thin film CoCrMo.
(Scan rate 0.25 mV/s.)The electrochemical behavior above 1 V could not be studied
because
the thin film was destroyed under this aggressive oxidizing condition.
Overall, the electrochemical behavior is characteristic of behavior
previously reported for CoCrMo alloys.[26] The similarity between the bulk and thin-film responses is taken
as evidence that the thin films are an accurate model for studying
corrosion processes in CoCrMo alloys. The additional information provided
by the QCM experiments with the thin films allows us to better understand
the processes occurring in the different potential regimes introduced
above and is the focus of the remainder of this Article.
Mass Change Behavior During the Corrosion
of CoCrMo Thin Films
Mass changes on CoCrMo films were measured
in situ during potentiodynamic tests in Ringer’s solution,
buffer solution, and BCS; the results are displayed in Figure 4 with the potentiodynamic response. Mass changes
as a function of total charge released, obtained by integrating the
current over time, are displayed in Figure 5. According to Faraday’s law, the amount of mass lost during
a corrosion process should scale linearly with the charge. In our
notation, we write Faraday’s law in the following way[21]where ΔM is the
change in mass per unit area, q is the charge density in equivalents
per unit area, and m is the molar mass per charge ratio. Our sign convention here is
that the current is positive during corrosion processes leading to
decreases in the mass. Note that q is obtained from the integrated charge by dividing by Faraday’s
constant (96 500 C/equiv). In situations where corrosion mechanisms
are changing, the behavior cannot be characterized by a single value
of m. In these cases,
we specify an effective value of m from the slope of a plot of the mass change versus charge:Changes in meff indicate a change in corrosion mechanisms or in the valence state
of the corroding species.
Figure 4
Mass changes and corresponding current density
on CoCrMo thin films
during potentiodynamic scans (scan rate: 0.25 mV/s) in (a) Ringer’s
solution, (b) buffer solution, and (c) BCS.
Figure 5
Change in mass on CoCrMo thin films as a function
of charge released
during potentiodynamic scans in (a) Ringer’s solution, (b)
buffer solution, and (c) BCS.
Mass changes and corresponding current density
on CoCrMo thin films
during potentiodynamic scans (scan rate: 0.25 mV/s) in (a) Ringer’s
solution, (b) buffer solution, and (c) BCS.
Ringer’s Solution
In each
electrolyte, specimens did not exhibit appreciable mass change below
0.4 V, although the current began to increase at 0.3 V. In Ringer’s
solution (Figure 4a), the alloy continuously
lost mass as the potential increased above 0.4 V. Figure 5a illustrates that the alloy corroded with a constant
value of 15 g/equiv for meff. This value is significantly
lower than what would be expected from the ideal mass per charge ratio
of the alloy, which can be determined from the following equation[21]in which f, n, and a are the
mass fraction, valence state, and molar mass of each alloying element,
respectively. If the alloy is considered to have a composition of
Co-29Cr-6Mo with valence states of 2, 3, and 4 for the respective
elements,[27] then a value of approximately
24 g/equiv is expected for m. In contrast, if the valence states of the corroding species
are considered to be 2, 6 and 6, respectively, then the theoretical m ratio is 17, which is significantly
closer to our measured value. The disagreement between the experimentally
determined value for meff and theoretically
expected values for m likely results from the formation of an oxide film in solution,
which grows while the alloy is simultaneously corroding. These results
are consistent with what was reported by Valero Vidal et al.[19] in PBS.Change in mass on CoCrMo thin films as a function
of charge released
during potentiodynamic scans in (a) Ringer’s solution, (b)
buffer solution, and (c) BCS.
Buffer Solution
In buffer solution
(Figures 4b and 5b),
the film displayed a more distinct passive regime between −0.4
and 0.2 V but again began to lose mass at 0.4 V with meff = 16 g/equiv. In contrast to the experiment in Ringer’s
solution, the film that was corroded in buffer solution exhibited
a current peak and subsequent stabilization between 0.65 and 0.8 V.
Here, the rate of mass loss declined. As demonstrated in Figure 5b, meff dropped as low as
3.6 g/equiv between 0.75 and 0.85 V. Above 0.85 V, the rate of mass
loss began to increase again with the current, approaching 18 g/equiv
by the end of the potentiodynamic scan.
Bovine
Calf Serum
As indicated
in Figure 4c, the corrosion behavior in BCS
was similar to that in buffer solution until the potential reached
∼0.8 V. At this point, a substantial increase in the mass was
observed, despite the increased current that predicts mass loss. Because
the behavior was not seen in Ringer’s or buffer solution and
was not reported by Valero Vidal in PBS,[19] mass gain presumably results from protein deposition. This deposited
film persists to potentials up to about 0.93 V. At higher potentials,
the underlying CoCrMo film corroded rapidly and was removed from the
quartz crystal surface. The plot of mass change with charge (Figure 5c) shows that the alloy initially lost mass with
an meff of 13 g/equiv. The rate of mass gain
was not linear with charge, and the rate of final mass loss was not
consistent among the samples.
Dissipation
Measurements
Dissipation
changes were also measured during potentiodynamic scans. In Ringer’s
and buffer solution, the dissipation remained relatively constant
until the film began to lose mass. Neither the potential nor the degree
of mass loss at which dissipation increased was consistent among the
tests. In BCS, dissipation increased at approximately the same potential
at which mass gain occurred. The ratio of the dissipation shift to
frequency shift (ΔΓ/Δf) from a
potentiodynamic scan in BCS is shown in Figure 6. Surface roughness can increase dissipation and cause an incorrectly
large mass gain reading on the QCM if surface cavities entrap fluid.[30−33] However, frequency shifts resulting from roughness for samples immersed
in a liquid medium scale as n1/2,[34] which was not true of our samples. The presence
of a viscoelastic (rather than rigid) material on the QCM surface
can also enhance dissipation. If viscoelastic effects are small, meaning
ΔΓ/Δf ≪ 1, then eq 1 holds and Δf scales with
the harmonic order, n. If viscoelastic effects are
significant, then ΔM as calculated from eq 1 will depend
on the harmonic and a viscoelastic correction factor must be included
in the analysis.[18,35] In our experiments, as demonstrated
in Figure 7, the value of ΔM obtained from the Sauerbrey equation
was the same for all three harmonics prior to mass deposition and
during the initial stages of the mass deposition process. Above ∼0.85
V, ΔΓ/Δf became significant and
ΔM became dependent
on n. Hence, we attribute the enhanced dissipation
to the viscoelastic character of this adsorbed layer, which supports
the hypothesis that the measured mass increase in BCS results from
protein deposition. Additional information about the viscoelastic
properties of the deposited layer can be obtained from a more detailed
analysis of Δf and ΔΓ in this highly
dissipative regime.[17,18,35] This analysis is the focus of an ongoing investigation.
Figure 6
Ratio of dissipation
to frequency shift on CoCrMo thin films during
potentiodynamic scans in BCS. (Scan rate: 0.25 mV/s.)
Figure 7
Mass changes on CoCrMo thin films during potentiodynamic
scans
in BCS, measured at three harmonics as a function of (a) potential
and (b) released charge. (Scan rate: 0.25 mV/s.)
Ratio of dissipation
to frequency shift on CoCrMo thin films during
potentiodynamic scans in BCS. (Scan rate: 0.25 mV/s.)Mass changes on CoCrMo thin films during potentiodynamic
scans
in BCS, measured at three harmonics as a function of (a) potential
and (b) released charge. (Scan rate: 0.25 mV/s.)
Conditions for Mass Deposition on CoCrMo in
BCS
To determine if there is a critical condition that promotes
mass deposition on CoCrMo in BCS, the potential applied to the film
was increased from −0.8 V and held constant at various levels.
The current density, total charge (current integrated over time),
and potential were quantified at the points at which the initial mass
loss occurred, the rate of mass loss decreased, the mass increase
began, and the final mass loss began. No clear correlation existed
between these processes and the current density or integrated charge.
However, the potentials at each point of mass change were consistent
and are provided in Table 1. Mass deposition
begins at potentials near 0.77 V in all cases, always following the
current peak and stabilization that is characteristic of the transpassive
regime. Thus, it is presumed that mass deposition results from reactions
between proteins and the ionic species that are formed near this potential.
Table 1
Electrode Potential at Critical Points
of Mass Change Behavior of CoCrMo Thin Films Corroded in BCS
potential (V vs SCE)
initial mass loss
0.53 ± 0.04
mass stabilization
0.68 ± 0.02
mass increase
0.77 ± 0.04
final mass loss
0.92 ± 0.05
Time dependence of the potential applied
to CoCrMo thin films (upper)
and of the corresponding mass changes (lower) in experiments to determine
the conditions promoting deposition in BCS solutions. (a) Potential
ramped from −0.8 to 1.2 V. (b) Potential ramped from −0.8
to 0.83 V and held at 0.83 V. (c) Potential ramped from −0.8
to 0.91 V and held at 0.91 V.The data also indicate that there is a maximum potential
for which
the deposited film will remain stable for extended periods of time.
When the potential was increased from −0.8 V at a rate of 0.25
mV/s and then held at 0.83 V, the film experienced comparable deposition
as during the complete potentiodynamic scan but maintained the layer
throughout the test, as displayed in Figure 8a,b. In contrast, when the potential was held near the end of the
mass gain regime at 0.91 V (Figure 8c), the
film corroded in nearly the same manner as when the potential was
continuously increased to 1.0 V, even though the current density was
comparable to that at 0.83 V (data not shown).
Figure 8
Time dependence of the potential applied
to CoCrMo thin films (upper)
and of the corresponding mass changes (lower) in experiments to determine
the conditions promoting deposition in BCS solutions. (a) Potential
ramped from −0.8 to 1.2 V. (b) Potential ramped from −0.8
to 0.83 V and held at 0.83 V. (c) Potential ramped from −0.8
to 0.91 V and held at 0.91 V.
Individual
Alloying Elements
Potentiodynamic
tests were also conducted on the pure alloying elements in BCS to
determine if a particular ion is responsible for mass deposition on
the alloy. Mass changes with potential variation are shown for each
metal in Figure 9. Cobalt specimens (Figure 9a) monotonically lost mass as the potential increased.
Mass loss began near −0.48 V as the current reached a plateau.
The meff (Figure 10a)
averaged 29.3 ± 4.2 g/equiv, approximately corresponding to the
Co(II) oxidation state as expected. Chromium specimens (Figure 9b) gained approximately 3.7 mg/m2 mass
at 0.38 V but continuously lost mass beginning at 0.5 V. The mass
gain at 0.38 V is 1 to 2 orders of magnitude smaller than on the alloy
and corresponds to a stabilization in the current. Hence, it is likely
attributed to an enhancement in the oxide thickness on the chromium
surface. The initial value of meff (Figure 10b) was approximately 9 g/equiv, nearly corresponding
to the dissolution of Cr(VI) ions, but decreased to 6.7 g/equiv above
0.85 V.
Figure 9
Mass changes and corresponding current density of (a) Co, (b) Cr,
and (c) Mo thin films during potentiodynamic scans in BCS. (Scan rate:
0.25 mV/s.)
Figure 10
Change in mass as a function of charge released during
potentiodynamic
scans in BCS on (a) Co, (b) Cr, and (c) Mo thin films.
Mass changes and corresponding current density of (a) Co, (b) Cr,
and (c) Mo thin films during potentiodynamic scans in BCS. (Scan rate:
0.25 mV/s.)Molybdenum samples (Figure 9c) exhibited
significantly different mass change behavior than cobalt or chromium.
The metal film began to lose mass as the potential reached −0.16
± 0.02 V with meff = 15 g/equiv
(Figure 10c), corresponding to the dissolution
of Mo(VI) ions. As the potential neared 0.06 ± 0.02 V or the
current density reached approximately 5 × 10–4 A/cm2, specimens gained at least 450 mg/m2 mass. Some specimens gained nearly 5000 mg/m2 mass. The
QCM signal was lost at ∼0.10 V because of the large frequency
shift; therefore, tests were stopped at this potential. The mass did
not increase on Mo films during potentiodynamic tests in either Ringer’s
or buffer solution. Our hypothesis based on these results is that
Mo(VI) ions are critical to the initiation of mass deposition on both
the Mo and CoCrMo surfaces.Neutral Mo oxidizes to Mo(IV), forming
MoO2 according
to the following reaction with water above the potential E0:[36]This reaction occurs spontaneously at all
pH values relevant to our experiment, thus our assumption is that
a thin layer of MoO2 is always present at the metal surface.
At higher potentials, Mo(IV) further oxidizes to Mo(VI) according
to the following reaction:[36]The equilibrium reduction
potential (vs SCE) for this reaction
is[36]No other reaction is predicted
until water is oxidized to evolve O2.[36]Change in mass as a function of charge released during
potentiodynamic
scans in BCS on (a) Co, (b) Cr, and (c) Mo thin films.A useful estimate of the molybdate (MoO4–2) ion concentration at the point at which mass
deposition occurs
can be obtained by assuming that all of the mass loss up to this point
is due to the release of Mo from the surface as MoO4–2. From Figure 9c we see that
mass deposition occurs when ∼400 mg/m2 of Mo has
been released into solution, corresponding to a surface molybdate
concentration, [MoO4–2], of ∼4 × 10–3 mol/m2. These
ions are concentrated within a thin layer of width, ≈ (Dt)1/2, where D is the diffusion
coefficient of the ions
in solution and t is the time that the ions have
had to diffuse. For the molydate ions, we expect D ≈ 3 × 10–10 m2/s, a value
comparable to the diffusion coefficient of similarly sized ions in
water.[37] Mass deposition occurs about 15
min after molybdate ions are first released in solution (Figure 9c). Taking t = 900 s, we obtain ≈ 5 × 10–4 m and [MoO4–2]/ ≈
8 mol/m3, corresponding
to a molar concentration of 8 × 10–3 M.The measured potential at which oxidation to Mo(VI) occurs is consistent
with eq 8 provided that possible changes in
the local pH are accounted for. For example, if we assume [MoO4–2] = 8 mM and a pH equal to the starting
pH of the BCS solution (7.6), then we obtain E0 = −0.6 V from eq 8. At the opposite
extreme, we can assume that the BCS solution has no buffering capacity
so that the eight protons released by the combination of the reactions
shown in eqs 5 and 7 contribute to the decrease in the local pH. In this case, the pH
is reduced to ∼1 and we obtain E0 = 0.182 V from eq 8. Not surprisingly, the
measured potential range over which molybdate ions are released into
solution is between these two extremes. Kinetic effects will also
play a role here because the equilibrium conditions for which eq 8 can be applied quantitatively are not established
under the corrosion conditions relevant to our experiments.It is likely that any change in local pH that occurs upon oxidation
of molydenum will induce changes in the surface charge of the Mo specimen
and proteins in solution, leading to mass deposition from electrostatic
adsorption. However, the magnitude of mass increase measured by the
QCM is significantly greater than what would result from a monolayer
of adsorbed proteins. For example, Ithurbide et al. found that the
adsorption of albumin onto a passivated chromium specimen increased
the mass on the surface by only 4 mg/m2, which is 2 to
3 orders of magnitude smaller than what was measured in this study.[38] Thus, whereas the local protein/substrate interactions
may play an important role in the adhesion of the interfacial layer
to the metallic substrate, the measured mass gain cannot be explained
by adsorption processes that by definition are confined to the immediate
vicinity of the metallic surface. It is more appropriate to think
of the mass gain as resulting from the formation of some sort of interfacial
phase that forms upon the release of Mo(VI) ions (presumably in the
form of molybdates) into a solution that contains appropriate macromolecules.
On the CoCrMo alloy, the mass gain occurs at potentials within the
secondary passive regime where the electrochemical behavior is believed
to be controlled by the oxidation of Cr(III) to Cr(VI). However, because
mass deposition occurred only on the Mo films and not on Co nor Cr,
we conclude that a release of Mo(VI) ions rather than Cr(VI) is necessary
to forming the deposit on the CoCrMo surface.
X-ray Photoelectron Spectroscopy
XPS was used to assess the composition of the deposited layers on
QCM specimens that exhibited material deposition without subsequent
mass loss. The CoCrMo specimen was prepared by conducting a potentiodynamic
scan in BCS from −0.8 to 0.83 V at 0.25 mV/s and then holding
the potential at 0.83 V for 30 min. The Mo sample was prepared by
conducting a potentiodynamic scan in BCS from −0.8 to 0.08
V. Subsequently, specimens were rinsed once in deionized water and
twice in methanol and dried with a compressed air canister to remove
any serum that may have simply adhered to the film. A typical survey
scan of each surface is shown in Figure 11.
The peak binding energy and atomic percentage of each element is included
in Table 2. Data confirm that the deposited
layer on both metals is primarily organic and derived from proteins:
the N 1s peak near 400 eV corresponds to the binding energy of amine
or amide bonds, and the N to total C ratio (0.22 ± 0.02 on CoCrMo,
0.23 ± 0.07 on Mo) is the same as that reported for bovine serum
albumin powder.[38] Additionally, the decomposition
of the C 1s peak (Figure 11b,d) shows evidence
of carbon in an organic binding environment. For CoCrMo, the C 1s
peak was decomposed into two peaks; the peak at 284.8 ± 0.03
eV corresponds to adventitious carbon, and the peak at 287.3 ±
0.34 eV corresponds to carbon in an amine or amide bond.[39] The C 1s peak on the Mo specimen was decomposed
into three peaks. In addition to the adventitious carbon peak, a peak
is present at 286.1 ± 0.06 eV, attributed to C–O and C–N
bonds,[38,40] and at 287.7 ± 0.06 eV, attributed
to peptide bonds.[41] Neither specimen exhibited
an appreciable metallic contribution. High-resolution scans at Cr
2p (570–595 eV) and Mo 3d (220–240 eV) on the CoCrMo
specimen did not detect either element. Our primary conclusion from
these results is that the deposited layer is primarily organic, with
a very small metallic component.
Figure 11
XPS survey scans and
decomposed C 1s spectra of CoCrMo and Mo surfaces
following mass deposition in BCS. (a) CoCrMo survey, (b) CoCrMo C
1s, (c) Mo survey, and (d) Mo C 1s.
Table 2
Binding
Energies and Atomic Percentages
of Elements Present on CoCrMo and Mo Surfaces after Mass Gain in BCS
CoCrMo
Mo
peak BE (eV)
atom %
peak BE (eV)
atom
%
C 1s
285.4 ± 0.1
74.2 ± 5.6
285.8 ± 0.02
60.9 ± 2.1
N 1s
399.8 ± 0.2
16.5 ± 0.03
400.0 ± 0.01
14.1 ± 4.1
O 1s
532.2 ± 0.02
8.8 ± 5.1
531.7 ± 0.1
23.2 ± 1.8
Co 2p
782.8 ± 0.5
0.27 ± 0.27
Cr 2p
Mo 3d
230.0 ± 0.2
4.3 ± 1.4
Summary
An electrochemical quartz crystal microbalance was utilized to
investigate the corrosion of CoCrMo thin films in Ringer’s
solution, buffer solution, and bovinecalf serum. Films were prepared
in a manner such that their microstructure and electrochemical behavior
model those of MoM hip joint surfaces. The key findings emerging from
this work are as follows: (1) If proteins are present in solution,
then the QCM measures a mass increase on the alloy surface during
corrosion at elevated potentials. (2) Typical values for the mass
of the deposited film range from 50 to 200 mg/m2. (3) Material
is deposited at a critical potential, 0.77 V, and can be maintained
for some time at slightly higher potentials, suggesting that deposition
results from redox reactions between oxidized alloying elements and
proteins in solution. (4) Dissipation of the QCM signal increases
with mass gain, providing evidence of a viscoelastic layer. (5) QCM
measurements at multiple harmonics scale with the Sauerbrey equation
prior to large increases in dissipation, indicating that mass gain
measurements are real and not an artifact of roughness effects. (6)
During the corrosion of Co, Cr, and Mo films in protein solutions,
mass deposition was found only on the Mo films under conditions corresponding
to the release of Mo(VI) ions into solution. From this, we infer that
the mass gain results from interactions between Mo(VI) ions and proteins
in solution. (7) XPS analysis of CoCrMo and Mo films after mass gain
reveals a primarily organic surface, further supporting conclusions
that mass gain results from protein deposition.XPS survey scans and
decomposed C 1s spectra of CoCrMo and Mo surfaces
following mass deposition in BCS. (a) CoCrMo survey, (b) CoCrMo C
1s, (c) Mo survey, and (d) Mo C 1s.The organic film is reminiscent of tribochemical reaction
layers
that form on the surface of MoM hip bearings, suggesting the existence
of local electrochemical pathways that promote the protein degradation
essential for their formation. Because of the reproducibility of the
experiments, these purely electrochemical experiments are very well
suited for fundamental investigations of tribochemical processes that
also involve mechanical forces. Further investigations are being undertaken
in order to more completely understand the mechanisms involved in
the formation of these electrochemically induced organic reaction
layers in vivo and their effects on the tribological performance of
CoCrMo alloys.
Authors: Nam-Joon Cho; J Nelson D'Amour; Johan Stalgren; Wolfgang Knoll; Kay Kanazawa; Curtis W Frank Journal: J Colloid Interface Sci Date: 2007-08-13 Impact factor: 8.128
Authors: Kevin J Bozic; Steven M Kurtz; Edmund Lau; Kevin Ong; Thomas P Vail; Daniel J Berry Journal: J Bone Joint Surg Am Date: 2009-01 Impact factor: 5.284
Authors: Garret C DeNolf; Larry Haack; Joe Holubka; Ann Straccia; Kay Blohowiak; Chris Broadbent; Kenneth R Shull Journal: Langmuir Date: 2011-07-18 Impact factor: 3.882
Authors: Heléne Osterlund; Sara Chlot; Mikko Faarinen; Anders Widerlund; Ilia Rodushkin; Johan Ingri; Douglas C Baxter Journal: Anal Chim Acta Date: 2010-10-07 Impact factor: 6.558
Authors: M A Wimmer; M P Laurent; M T Mathew; C Nagelli; Y Liao; L D Marks; J J Jacobs; A Fischer Journal: Wear Date: 2015 May-Jun Impact factor: 3.892