Demonstration of enhanced iodine K-edge imaging using an energy-dispersive X-ray computed tomography system with a 25 mm/s-scan linear cadmium telluride detector and a single comparator.

Eiichi Sato1, Yasuyuki Oda, Abulajiang Abudurexiti, Osahiko Hagiwara, Hiroshi Matsukiyo, Akihiro Osawa, Toshiyuki Enomoto, Manabu Watanabe, Shinya Kusachi, Shigehiro Sato, Akira Ogawa, Jun Onagawa.   

Abstract

An energy-dispersive (ED) X-ray computed tomography (CT) system is useful for carrying out monochromatic imaging. To perform enhanced iodine K-edge CT, we developed an oscillation linear cadmium telluride (CdTe) detector with a scan velocity of 25 mm/s and an energy resolution of 1.2 keV. CT is performed by repeated linear scans and rotations of an object. Penetrating X-ray photons from the object are detected by the CdTe detector, and event signals of X-ray photons are produced using charge-sensitive and shaping amplifiers. The lower photon energy is determined by a comparator device, and the maximum photon energy of 60 keV corresponds to the tube voltage. Rectangular-shaped comparator outputs are counted by a counter card. In the ED-CT, tube voltage and current were 60 kV and 0.30 mA, respectively, and X-ray intensity was 14.8 μGy/s at 1.0m from the source at a tube voltage of 60 kV. Demonstration of enhanced iodine K-edge X-ray CT for cancer diagnosis was carried out by selecting photons with energies ranging from 34 to 60 keV. Copyright Â
© 2012 Elsevier Ltd. All rights reserved.

Entities:  

Year:  2012        PMID: 22364788     DOI: 10.1016/j.apradiso.2012.02.007

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  2 in total

1.  Optimization of K-edge imaging with spectral CT.

Authors:  Peng He; Biao Wei; Wenxiang Cong; Ge Wang
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

2.  Impact of anti-charge sharing on the zero-frequency detective quantum efficiency of CdTe-based photon counting detector system: cascaded systems analysis and experimental validation.

Authors:  Xu Ji; Ran Zhang; Guang-Hong Chen; Ke Li
Journal:  Phys Med Biol       Date:  2018-05-01       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.