Continuous integrated distal capnography in infants ventilated with high frequency ventilation.

Amir Kugelman1, Arieh Riskin, Irit Shoris, Michal Ronen, Iris Shalev Stein, David Bader.   

Abstract

OBJECTIVE: To assess within a feasibility study the correlation, agreement, and trending of continuous integrated distal capnography (dCap) with PaCO(2) in infants on HFV. STUDY
DESIGN: Sixteen premature infants [median (range) gestational age: 26.5 (24.7-34.7) weeks], ventilated with HFV (mean ± SD airway pressure: 8.1 ± 2.1 cmH(2) O, FiO(2) : 0.39 ± 0.21) for RDS, intubated with a double-lumen endotracheal-tube and whose data were recorded on a bedside computer participated in the study. Side-stream dCap was measured via the extra-port of a double-lumen endotracheal-tube by a Microstream capnograph, with a specially designed software for HFV and compared with simultaneous PaCO(2) . Integrated time-window analysis of the data was performed retrospectively on data collected prospectively.
RESULTS: Analysis included 195 measurements. The correlation of dCap with PaCO(2) (r = 0.68, P < 0.0001) and the agreement (bias ± precision: -2.0 ± 10.7 mmHg) were adequate. Area under the ROC curves for dCap to detect high (>60 mmHg) or low (<35 mmHg) PaCO(2) was 0.79 (CI: 0.70-0.89) and 0.87 (CI: 0.73-1.00), respectively; P < 0.0001. Changes in dCap and in PaCO(2) for consecutive measurements within each patient were adequately correlated (r = 0.65, P < 0.0001).
CONCLUSIONS: Continuous integrated dCap is feasible in premature infants ventilated with HFV and can be helpful for trends and alarm for unsafe levels of PaCO(2) .
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22328495     DOI: 10.1002/ppul.22524

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  2 in total

1.  Bench test assessment of mainstream capnography during high frequency oscillatory ventilation.

Authors:  Caroline M Hartdorff; Marc van Heerde; Dick G Markhorst
Journal:  J Clin Monit Comput       Date:  2013-08-23       Impact factor: 2.502

Review 2.  Correlation of End-Tidal Carbon Dioxide with Arterial Carbon Dioxide in Mechanically Ventilated Neonates: A Scoping Review.

Authors:  Ingra Pereira Monti Martins; Adriane Muller Nakato; Paula Karina Hembecker; Sérgio Ossamu Ioshii; Percy Nohama
Journal:  Glob Pediatr Health       Date:  2021-05-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.