Use of internal scintillator radioactivity to calibrate DOI function of a PET detector with a dual-ended-scintillator readout.

Chad Bircher1, Yiping Shao.   

Abstract

PURPOSE: Positron emission tomography (PET) detectors that use a dual-ended-scintillator readout to measure depth-of-interaction (DOI) must have an accurate DOI function to provide the relationship between DOI and signal ratios to be used for detector calibration and recalibration. In a previous study, the authors used a novel and simple method to accurately and quickly measure DOI function by irradiating the detector with an external uniform flood source; however, as a practical concern, implementing external uniform flood sources in an assembled PET system is technically challenging and expensive. In the current study, therefore, the authors investigated whether the same method could be used to acquire DOI function from scintillator-generated (i.e., internal) radiation. The authors also developed a method for calibrating the energy scale necessary to select the events within the desired energy window.
METHODS: The authors measured the DOI function of a PET detector with lutetium yttrium orthosilicate (LYSO) scintillators. Radiation events originating from the scintillators' internal Lu-176 beta decay were used to measure DOI functions which were then compared with those measured from both an external uniform flood source and an electronically collimated external point source. The authors conducted these studies with several scintillators of differing geometries (1.5 × 1.5 and 2.0 × 2.0 mm(2) cross-section area and 20, 30, and 40 mm length) and various surface finishes (mirror-finishing, saw-cut rough, and other finishes in between), and in a prototype array.
RESULTS: All measured results using internal and external radiation sources showed excellent agreement in DOI function measurement. The mean difference among DOI values for all scintillators measured from internal and external radiation sources was less than 1.0 mm for different scintillator geometries and various surface finishes.
CONCLUSIONS: The internal radioactivity of LYSO scintillators can be used to accurately measure DOI function in PET detectors, regardless of scintillator geometry or surface finish. Because an external radiation source is not needed, this method of DOI function measurement can be practically applied to individual PET detectors as well as assembled systems.

Entities:  

Mesh:

Year:  2012        PMID: 22320787      PMCID: PMC3272069          DOI: 10.1118/1.3676688

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  For: Is LSO the future of PET?

Authors:  Ronald Nutt
Journal:  Eur J Nucl Med Mol Imaging       Date:  2002-11       Impact factor: 9.236

2.  Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes.

Authors:  Yongfeng Yang; Purushottam A Dokhale; Robert W Silverman; Kanai S Shah; Mickel A McClish; Richard Farrell; Gerald Entine; Simon R Cherry
Journal:  Phys Med Biol       Date:  2006-04-10       Impact factor: 3.609

3.  Depth of interaction calibration for PET detectors with dual-ended readout by PSAPDs.

Authors:  Yongfeng Yang; Jinyi Qi; Yibao Wu; Sara St James; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2008-12-19       Impact factor: 3.609

4.  A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout.

Authors:  Yiping Shao; Rutao Yao; Tianyu Ma
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

5.  Energy and Timing Measurement with Time-Based Detector Readout for PET Applications: Principle and Validation with Discrete Circuit Components.

Authors:  Xishan Sun; Allan K Lan; Chad Bircher; Zhi Deng; Yinong Liu; Yiping Shao
Journal:  Nucl Instrum Methods Phys Res A       Date:  2011-06-11       Impact factor: 1.455

6.  NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes.

Authors:  Brad J Kemp; Chang Kim; John J Williams; Alexander Ganin; Val J Lowe
Journal:  J Nucl Med       Date:  2006-12       Impact factor: 10.057

7.  Lutetium oxyorthosilicate (LSO) intrinsic activity correction and minimal detectable target activity study for SPECT imaging with a LSO-based animal PET scanner.

Authors:  Rutao Yao; Tianyu Ma; Yiping Shao
Journal:  Phys Med Biol       Date:  2008-07-31       Impact factor: 3.609

  7 in total
  6 in total

1.  Field-programable-gate-array-based distributed coincidence processor for high count-rate online positron emission tomography coincidence data acquisition.

Authors:  Xinyi Cheng; Kun Hu; Dongxu Yang; Yiping Shao
Journal:  Phys Med Biol       Date:  2021-02-16       Impact factor: 3.609

2.  A compact and lightweight small animal PET with uniform high-resolution for onboard PET/CT image-guided preclinical radiation oncology research.

Authors:  Xinyi Cheng; Kun Hu; Dongxu Yang; Yiping Shao
Journal:  Phys Med Biol       Date:  2021-10-19       Impact factor: 4.174

3.  Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics.

Authors:  Yiping Shao; Xishan Sun; Kejian A Lan; Chad Bircher; Kai Lou; Zhi Deng
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

4.  Performance evaluation of a depth-of-interaction detector by use of position-sensitive PMT with a super-bialkali photocathode.

Authors:  Yoshiyuki Hirano; Munetaka Nitta; Naoko Inadama; Fumihiko Nishikido; Eiji Yoshida; Hideo Murayama; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2013-08-21

5.  Depth of interaction determination in monolithic scintillator with double side SiPM readout.

Authors:  Matteo Morrocchi; Giovanni Ambrosi; Maria Giuseppina Bisogni; Filippo Bosi; Marco Boretto; Piergiorgio Cerello; Maria Ionica; Ben Liu; Francesco Pennazio; Maria Antonietta Piliero; Giovanni Pirrone; Vasile Postolache; Richard Wheadon; Alberto Del Guerra
Journal:  EJNMMI Phys       Date:  2017-02-16

6.  Analytical study of the effect of the system geometry on photon sensitivity and depth of interaction of positron emission mammography.

Authors:  Pablo Aguiar; Cristina Lois
Journal:  J Oncol       Date:  2012-09-20       Impact factor: 4.375

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.