Comparison of central corneal thickness measurements by Galilei Dual-Scheimpflug analyzer® and ultrasound pachymeter in myopic eyes.

Volkan Yeter1, Baris Sönmez, Umit Beden.   

Abstract

BACKGROUND AND
OBJECTIVE: To compare the central corneal thickness (CCT) measurements by Galilei Dual-Scheimpflug analyzer (GSA) (Ziemer Group, Port, Switzerland) and ultrasound pachymeter (UP) in myopic eyes. PATIENTS AND METHODS: In this prospective study, CCT of 161 myopic eyes of 81 refractive surgery candidates (35 female, 46 male; mean age: 23.18 ± 4.08 years) were measured by GSA and UP consecutively. The data were analyzed statistically by paired t test, Bland-Altman plot, and Pearson correlation test to assess the agreement of the measurements.
RESULTS: The mean CCTs obtained by GSA and UP were 559.85 ± 30.87 and 560.41 ± 34.45 μm, respectively. No significant difference was found between the measurements (P = .684) and regression analysis showed a high correlation between the measurements obtained with both devices (r = 0.86; P < .001). A strong agreement between the two methods was also found by Bland-Altman plot.
CONCLUSION: In myopic eyes, the CCT measurements of GSA and UP are similar and highly correlated. Because of high agreement between these devices, the GSA is a non-contact method that may be an alternative to UP for measurement of CCT. Copyright 2012, SLACK Incorporated.

Entities:  

Mesh:

Year:  2012        PMID: 22320416     DOI: 10.3928/15428877-20120102-08

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  4 in total

1.  Effect of obstructive sleep apnea syndrome on corneal thickness.

Authors:  Handan Inonu Koseoglu; Asiye Kanbay; Huseyin Ortak; Remzi Karadağ; Osman Demir; Selim Demir; Alper Gunes; Sibel Doruk
Journal:  Int Ophthalmol       Date:  2015-08-21       Impact factor: 2.031

Review 2.  The thick and thin of the central corneal thickness in glaucoma.

Authors:  Graham W Belovay; Ivan Goldberg
Journal:  Eye (Lond)       Date:  2018-02-15       Impact factor: 3.775

3.  Measurement of central corneal thickness and pre-corneal tear film thickness of rabbits using the Scheimpflug system.

Authors:  Jing Dong; Qiang Wu; Xiao-Gang Wang
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

4.  Decreased central corneal thickness in ankylosing spondylitis.

Authors:  Huseyin Ortak; Ahmet Inanır; Selim Demir; Alper Uysal; Şafak Şahin; Mustafa Sağcan; Yalçın Önder; Sait Alim; Ayşe Kevser Demir
Journal:  Int Ophthalmol       Date:  2013-07-12       Impact factor: 2.031

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.