Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma.

Sushmita Kaushik1, Surinder Singh Pandav, Anupam Banger, Kanika Aggarwal, Amod Gupta.   

Abstract

PURPOSE: To evaluate corneal biomechanical properties across the glaucoma spectrum and study the relationship between these measurements and intraocular pressure measured by Goldmann applanation tonometry (GAT-IOP) and central corneal thickness (CCT).
DESIGN: Prospective cross-sectional study.
SETTING: Tertiary-care teaching institute. STUDY POPULATION: A total of 323 eyes of 323 participants (71 normal, 101 glaucoma suspect [GS], 38 ocular hypertension [OHT], 59 primary angle-closure disease [PACD], 36 primary open-angle glaucoma [POAG], and 18 normal-tension glaucoma [NTG]) who had received no ophthalmic treatment. OBSERVATION PROCEDURES: Corneal hysteresis (CH), corneal resistance factor (CRF), corneal-compensated IOP (IOPcc), and Goldmann-correlated IOP (IOPg) measured by the Ocular Response Analyzer (ORA). GAT-IOP and CCT recorded in all subjects. MAIN OUTCOME MEASURES: Regression analysis used to determine the relationship between GAT-IOP, CCT, age, CRF, and CH. Bland-Altman plots used to assess agreement between IOP measured by GAT and the ORA (IOPg).
RESULTS: CH measurements were significantly less in POAG and NTG compared to normal subjects (P = .034 and P = .030 respectively), regardless of the IOP. The CRF was significantly less in NTG and maximum in POAG and OHT. Regression analysis with CH as dependant variable showed significant association with GAT-IOP and CRF (P < .001) but not CCT, persisting on multivariate analysis (adjusted R(2) = 0.483). GAT-IOP correlated strongly with Goldmann-correlated IOP on the ORA (IOPg) (r = 0.82; P < .001), but limits of agreement between the measurements were poor.
CONCLUSIONS: CH and CRF may constitute a pressure-independent risk factor for glaucoma. CRF appears to influence GAT-IOP measurements more than simple geometric thickness measured by CCT. However, IOP measurements from the ORA are not interchangeable with, and are unlikely to replace, Goldmann applanation tonometry at the present time.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22310080     DOI: 10.1016/j.ajo.2011.10.032

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  42 in total

1.  Genome-wide analysis of central corneal thickness in primary open-angle glaucoma cases in the NEIGHBOR and GLAUGEN consortia.

Authors:  Megan Ulmer; Jun Li; Brian L Yaspan; Ayse Bilge Ozel; Julia E Richards; Sayoko E Moroi; Felicia Hawthorne; Donald L Budenz; David S Friedman; Douglas Gaasterland; Jonathan Haines; Jae H Kang; Richard Lee; Paul Lichter; Yutao Liu; Louis R Pasquale; Margaret Pericak-Vance; Anthony Realini; Joel S Schuman; Kuldev Singh; Douglas Vollrath; Robert Weinreb; Gadi Wollstein; Donald J Zack; Kang Zhang; Terri Young; R Rand Allingham; Janey L Wiggs; Allison Ashley-Koch; Michael A Hauser
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-07-03       Impact factor: 4.799

2.  Corneal biomechanical changes after trabeculectomy with mitomycin C in primary open-angle glaucoma and pseudoexfoliation glaucoma.

Authors:  Rana Sorkhabi; Farhad Najafzadeh; Ali Sadeghi; Mohamadhosein Ahoor; Ali Mahdavifard
Journal:  Int Ophthalmol       Date:  2019-05-27       Impact factor: 2.031

3.  A Prospective Longitudinal Study to Investigate Corneal Hysteresis as a Risk Factor for Predicting Development of Glaucoma.

Authors:  Carolina N Susanna; Alberto Diniz-Filho; Fábio B Daga; Bianca N Susanna; Feilin Zhu; Nara G Ogata; Felipe A Medeiros
Journal:  Am J Ophthalmol       Date:  2018-01-02       Impact factor: 5.258

4.  Corneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas.

Authors:  Gabor Nemeth; Eszter Szalai; Ziad Hassan; Agnes Lipecz; Zsuzsa Flasko; Laszlo Modis
Journal:  Int J Ophthalmol       Date:  2017-02-18       Impact factor: 1.779

5.  IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Yang Zhang; Lin Zheng; Ailing Bian; Qi Zhou
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

6.  Short-term effects of overnight orthokeratology on corneal epithelial permeability and biomechanical properties.

Authors:  Thao N Yeh; Harry M Green; Yixiu Zhou; Julie Pitts; Britney Kitamata-Wong; Sophia Lee; Shiyin L Wang; Meng C Lin
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-06-06       Impact factor: 4.799

7.  Corneal biomechanics in different age groups.

Authors:  Ahmed Abdel Karim El Massry; Amr Ahmed Said; Ihab Mohamed Osman; Amr Saad Bessa; Mohammed Ahmed Elmasry; Eman Nabil Elsayed; Nader Hussein Lotfy Bayoumi
Journal:  Int Ophthalmol       Date:  2020-01-08       Impact factor: 2.031

8.  Corneal properties in children with congenital isolated growth hormone deficiency.

Authors:  Pinar Nalcacioglu-Yuksekkaya; Emine Sen; Ufuk Elgin; Mumin Hocaoglu; Faruk Ozturk; Sebahat Agladıoglu Yilmaz; Havva Nur Kendirci; Semra Cetinkaya; Zehra Aycan
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

9.  Synergic effect of corneal hysteresis and central corneal thickness in the risk of early-stage primary open-angle glaucoma progression.

Authors:  Maria A Jiménez-Santos; Federico Saénz-Francés; Rubén Sánchez-Jean; José María Martinez-de-la Casa; Julian García-Feijoo; Luis Jañez-Escalada
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-05-07       Impact factor: 3.117

10.  Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children.

Authors:  Ramin Salouti; Ali Agha Alishiri; Reza Gharebaghi; Mostafa Naderi; Khosrow Jadidi; Ahmad Shojaei-Baghini; Mohammadreza Talebnejad; Zahra Nasiri; Seyedmorteza Hosseini; Fatemeh Heidary
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.