Role of aortic stent graft oversizing and barb characteristics on folding.

Kathleen K Lin1, Jarin A Kratzberg, Madhavan L Raghavan.   

Abstract

OBJECTIVE: To evaluate folding in infrarenal stent grafts in relation to oversizing, barb angle, and barb length using computed tomography images of stent grafts deployed in explanted bovine aortas.
METHODS: Computed tomography data from an in vitro investigation on the effect of oversizing of 4% to 45% (n = 19), barb length of 2 to 7 mm (n = 11), and barb angle of 10° to 90° (n = 7) on device fixation were examined for instances of folding. Folding was classified as circumferential or longitudinal and quantified on an ordinal scale based on codified criteria. Cumulative fold ranking from 0 (no fold) to 6 (two severe folds) for each deployment was used as the measure of folding observed.
RESULTS: Of the 37 cases, cumulative mean ± standard deviation fold ranking for stent grafts oversized >30% (n = 5) was significantly greater than the rest (3.4 ± 1.7 vs 0.5 ± 1.2, respectively; Mann-Whitney U test; P < .005). When barb length was varied from 2 to 7 mm (oversizing held at 10%-20%), folding was noted in one of 11 cases. Similarly, when barb angle was varied from 0° (vertical) to 90° (horizontal), folding was not noted in any of the seven cases. The pullout force was not significantly different between stent grafts with and without folding (5.4 ± 1.95 vs 5.12 ± 1.89 N, respectively; P > .5). At least one instance of folding was noted in the seven of seven (100%) stent grafts with oversizing >23.5% and in only five of 30 (14%) stent grafts with oversizing <23.5%.
CONCLUSIONS: Stent graft folding was prevalent when oversized >30%. Large variations in barb length and angle did not aggravate folding risk when oversized within the recommended range of 10% to 20%.
Copyright © 2012 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22305271      PMCID: PMC3465615          DOI: 10.1016/j.jvs.2011.10.080

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  15 in total

1.  Eccentric stent graft compression: an indicator of insecure proximal fixation of aortic stent graft.

Authors:  Y G Wolf; B B Hill; W A Lee; C M Corcoran; T J Fogarty; C K Zarins
Journal:  J Vasc Surg       Date:  2001-03       Impact factor: 4.268

2.  The impact of stent design on proximal stent-graft fixation in the abdominal aorta: an experimental study.

Authors:  T Resch; M Malina; B Lindblad; J Malina; J Brunkwall; K Ivancev
Journal:  Eur J Vasc Endovasc Surg       Date:  2000-08       Impact factor: 7.069

3.  Influence of endograft oversizing on device migration, endoleak, aneurysm shrinkage, and aortic neck dilation: results from the Zenith Multicenter Trial.

Authors:  W Charles Sternbergh; Samuel R Money; Roy K Greenberg; Timothy A M Chuter
Journal:  J Vasc Surg       Date:  2004-01       Impact factor: 4.268

4.  Stent graft migration after endovascular aneurysm repair: importance of proximal fixation.

Authors:  Christopher K Zarins; Daniel A Bloch; Tami Crabtree; Alan H Matsumoto; Rodney A White; Thomas J Fogarty
Journal:  J Vasc Surg       Date:  2003-12       Impact factor: 4.268

5.  A framework for geometric analysis of vascular structures: application to cerebral aneurysms.

Authors:  Marina Piccinelli; Alessandro Veneziani; David A Steinman; Andrea Remuzzi; Luca Antiga
Journal:  IEEE Trans Med Imaging       Date:  2009-05-12       Impact factor: 10.048

Review 6.  Endograft collapse after thoracic endovascular aortic repair.

Authors:  Frederik H W Jonker; Felix J V Schlosser; Arnar Geirsson; Bauer E Sumpio; Frans L Moll; Bart E Muhs
Journal:  J Endovasc Ther       Date:  2010-12       Impact factor: 3.487

7.  Computational geometry for patient-specific reconstruction and meshing of blood vessels from MR and CT angiography.

Authors:  Luca Antiga; Bogdan Ene-Iordache; Andrea Remuzzi
Journal:  IEEE Trans Med Imaging       Date:  2003-05       Impact factor: 10.048

Review 8.  Oversizing of aortic stent grafts for abdominal aneurysm repair: a systematic review of the benefits and risks.

Authors:  J van Prehn; F J V Schlösser; B E Muhs; H J M Verhagen; F L Moll; J A van Herwaarden
Journal:  Eur J Vasc Endovasc Surg       Date:  2009-05-09       Impact factor: 7.069

9.  Factors and forces influencing stent-graft migration after endovascular aortic aneurysm repair.

Authors:  Irwin V Mohan; Peter L Harris; Corine J Van Marrewijk; Robert J Laheij; Thien V How
Journal:  J Endovasc Ther       Date:  2002-12       Impact factor: 3.487

10.  Endograft migration one to four years after endovascular abdominal aortic aneurysm repair with the AneuRx device: a cautionary note.

Authors:  Michael S Conners; W Charles Sternbergh; Glen Carter; Britt H Tonnessen; Moises Yoselevitz; Samuel R Money
Journal:  J Vasc Surg       Date:  2002-09       Impact factor: 4.268

View more
  3 in total

Review 1.  Contemporary Role of Computational Analysis in Endovascular Treatment for Thoracic Aortic Disease.

Authors:  Guido H W van Bogerijen; Jip L Tolenaar; Michele Conti; Ferdinando Auricchio; Francesco Secchi; Francesco Sardanelli; Frans L Moll; Joost A van Herwaarden; Vincenzo Rampoldi; Santi Trimarchi
Journal:  Aorta (Stamford)       Date:  2013-08-01

2.  Endovascular Aneurysm Repair in Patients with Conical Neck Anatomy.

Authors:  Jae Hoon Lee; Ki Hyuk Park
Journal:  Vasc Specialist Int       Date:  2017-06-30

3.  Virtual stenting with simplex mesh and mechanical contact analysis for real-time planning of thoracic endovascular aortic repair.

Authors:  Duanduan Chen; Jianyong Wei; Yiming Deng; Huanming Xu; Zhenfeng Li; Haoye Meng; Xiaofeng Han; Yonghao Wang; Jia Wan; Tianyi Yan; Jiang Xiong; Xiaoying Tang
Journal:  Theranostics       Date:  2018-11-10       Impact factor: 11.556

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.