Factors and forces influencing stent-graft migration after endovascular aortic aneurysm repair.

Irwin V Mohan1, Peter L Harris, Corine J Van Marrewijk, Robert J Laheij, Thien V How.   

Abstract

PURPOSE: To assess and validate the clinical features predisposing to stent-graft migration and to calculate the distal displacement forces exerted at the proximal fixation site following endovascular aortic aneurysm repair (EVAR).
METHODS: Demographic, anatomical, and graft-related features from 2862 patients were analyzed in a regression model to identify variables associated with stent-graft migration, which was defined as device movement >5 mm or considered significant by the investigator. Using the principles of continuity and momentum, a mathematical model of blood flow was created. The pulse pressure, proximal aortic and distal iliac diameters, and the degree of iliac angulation were varied in the calculations, and the distal displacement force exerted at the proximal fixation site was calculated.
RESULTS: Ninety-nine patients developed stent-graft migration, which was clinically relevant in 85 (3.0%). Hypertension (p=0.015), smoking (p=0.009), maximal aortic diameter (p=0.004), and distal transverse aortic diameter (p=0.03) correlated with migration in the univariate analysis, but iliac angulation did not quite achieve significance (p=0.06). On multivariate analysis, current smoking, hypertension, distal transverse aortic diameter, maximum common iliac diameter, and increasing proximal graft size were significantly associated with stent-graft migration. The mathematical model calculated the distal displacement force exerted on the proximal fixation site of the stent-graft and validated the clinical findings. The ratio of graft-diameter change from proximal aorta to distal iliac influenced the greatest increase in the displacement force.
CONCLUSIONS: The mathematical model validated hypertension, aneurysm morphology, and endograft size as clinical factors significantly associated with stent-graft migration. These findings may have important implications for the choice and design of future stent-grafts.

Entities:  

Mesh:

Year:  2002        PMID: 12546574     DOI: 10.1177/152660280200900606

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  7 in total

1.  Endovascular repositioning of a pipeline embolization device dislocated from the vertebral into the basilar artery using a stent-in-stent technique. Practical and technical considerations.

Authors:  H U Kerl; M Al-Zghloul; C Groden; M A Brockmann
Journal:  Clin Neuroradiol       Date:  2012-03       Impact factor: 3.649

2.  Role of aortic stent graft oversizing and barb characteristics on folding.

Authors:  Kathleen K Lin; Jarin A Kratzberg; Madhavan L Raghavan
Journal:  J Vasc Surg       Date:  2012-02-02       Impact factor: 4.268

3.  A computational study on the biomechanical factors related to stent-graft models in the thoracic aorta.

Authors:  S K Lam; George S K Fung; Stephen W K Cheng; K W Chow
Journal:  Med Biol Eng Comput       Date:  2008-07-11       Impact factor: 2.602

Review 4.  [Simulation of blood flow within the abdominal aorta. Computational fluid dynamics in abdominal aortic aneurysms before and after interventions].

Authors:  T Frauenfelder; E Boutsianis; H Alkadhi; B Marincek; T Schertler
Journal:  Radiologe       Date:  2007-11       Impact factor: 0.635

5.  Complementary roles of open and hybrid approaches to thoracoabdominal aortic aneurysm repair.

Authors:  Ehsan Benrashid; Hanghang Wang; Nicholas D Andersen; Jeffrey E Keenan; Richard L McCann; G Chad Hughes
Journal:  J Vasc Surg       Date:  2016-07-18       Impact factor: 4.268

Review 6.  A perspective review on numerical simulations of hemodynamics in aortic dissection.

Authors:  Wan Naimah Wan Ab Naim; Poo Balan Ganesan; Zhonghua Sun; Kok Han Chee; Shahrul Amry Hashim; Einly Lim
Journal:  ScientificWorldJournal       Date:  2014-02-03

7.  Fast and Accurate Computation of the Displacement Force of Stent Grafts after Endovascular Aneurysm Repair.

Authors:  Ming Qing; Zhan Liu; Tinghui Zheng
Journal:  Bioengineering (Basel)       Date:  2022-09-06