Literature DB >> 22298873

Virtual touch tissue quantification using acoustic radiation force impulse technology: initial clinical experience with solid breast masses.

Min Bai1, Lianfang Du, Jiying Gu, Fan Li, Xiao Jia.   

Abstract

OBJECTIVES: The purpose of this study was to investigate the clinical usage of Virtual Touch tissue quantification (VTQ; Siemens Medical Solutions, Mountain View, CA) implementing sonographic acoustic radiation force impulse technology for differentiation between benign and malignant solid breast masses.
METHODS: A total of 143 solid breast masses were examined with VTQ, and their shear wave velocities (SWVs) were measured. From all of the masses, 30 were examined by two independent operators to evaluate the reproducibility of the results of VTQ measurement. All masses were later surgically resected, and the histologic results were correlated with the SWV results. A receiver operating characteristic curve was calculated to assess the diagnostic performance of VTQ.
RESULTS: A total of 102 benign lesions and 41 carcinomas were diagnosed on the basis of histologic examination. The VTQ measurements performed by the two independent operators yielded a correlation coefficient of 0.885. Applying a cutoff point of 3.065 m/s, a significant difference (P < .001) was found between the SWVs of the benign (mean ± SD, 2.25 ± 0.59 m/s) and malignant (5.96 ± 2.96 m/s) masses. The sensitivity, specificity, and area under the receiver operating characteristic curve for the differentiation were 75.6%, 95.1%, and 85.6%, respectively. When the repeated non-numeric result X.XX of the SWV measurements was designated as an indicator of malignancy, the sensitivity, specificity, and accuracy were 63.4%, 100%, and 89.5%.
CONCLUSIONS: Virtual Touch tissue quantification can yield reproducible and quantitative diagnostic information on solid breast masses and serve as an effective diagnostic tool for differentiation between benign and malignant solid masses.

Entities:  

Mesh:

Year:  2012        PMID: 22298873     DOI: 10.7863/jum.2012.31.2.289

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  46 in total

1.  Combination of two-dimensional shear wave elastography with ultrasound breast imaging reporting and data system in the diagnosis of breast lesions: a new method to increase the diagnostic performance.

Authors:  Dan-Dan Li; Hui-Xiong Xu; Le-Hang Guo; Xiao-Wan Bo; Xiao-Long Li; Rong Wu; Jun-Mei Xu; Yi-Feng Zhang; Kun Zhang
Journal:  Eur Radiol       Date:  2015-12-29       Impact factor: 5.315

2.  Breast elastography: A literature review.

Authors:  A Goddi; M Bonardi; S Alessi
Journal:  J Ultrasound       Date:  2012-06-30

Review 3.  Acoustic radiation force elasticity imaging in diagnostic ultrasound.

Authors:  Joshua R Doherty; Gregg E Trahey; Kathryn R Nightingale; Mark L Palmeri
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2013-04       Impact factor: 2.725

4.  Elasticity of healthy Achilles tendon decreases with the increase of age as determined by acoustic radiation force impulse imaging.

Authors:  Zhengmin Ruan; Bin Zhao; Hengtao Qi; Yuan Zhang; Feixue Zhang; Mei Wu; Guangrui Shao
Journal:  Int J Clin Exp Med       Date:  2015-01-15

5.  The value of virtual touch tissue imaging quantification in the differential diagnosis between benign and malignant breast lesions.

Authors:  Wen-Tao Kong; Wei-Jun Zhou; Yin Wang; Xiao-Min Zhuang; Min Wu
Journal:  J Med Ultrason (2001)       Date:  2019-05-17       Impact factor: 1.314

Review 6.  Ultrasound Imaging Technologies for Breast Cancer Detection and Management: A Review.

Authors:  Rongrong Guo; Guolan Lu; Binjie Qin; Baowei Fei
Journal:  Ultrasound Med Biol       Date:  2017-10-26       Impact factor: 2.998

7.  Acoustic Radiation Force Impulse Imaging in Benign and Malignant Breast Lesions.

Authors:  Jayapriya Jayaraman; Venkatraman Indiran; Kanakaraj Kannan; Prabakaran Maduraimuthu
Journal:  Cureus       Date:  2017-06-01

8.  Acoustic radiation force impulse (ARFI) in the evaluation of the renal parenchymal stiffness in paediatric patients with vesicoureteral reflux: preliminary results.

Authors:  Costanza Bruno; Giuliana Caliari; Marco Zaffanello; Milena Brugnara; Michele Zuffante; Mariangela Cecchetto; Salvatore Minniti; Alice Pedot; Giorgio Talamini; Roberto Pozzi-Mucelli
Journal:  Eur Radiol       Date:  2013-07-24       Impact factor: 5.315

Review 9.  Acoustic radiation force impulse elastography for differentiation of malignant and benign breast lesions: a meta-analysis.

Authors:  Dan-Dan Li; Le-Hang Guo; Hui-Xiong Xu; Chang Liu; Jun-Mei Xu; Li-Ping Sun; Jian Wu; Bo-Ji Liu; Lin-Na Liu; Xiao-Hong Xu
Journal:  Int J Clin Exp Med       Date:  2015-04-15

10.  Update on Breast Cancer Detection Using Comb-Push Ultrasound Shear Elastography.

Authors:  Max Denis; Mahdi Bayat; Mohammad Mehrmohammadi; Adriana Gregory; Pengfei Song; Dana H Whaley; Sandhya Pruthi; Shigao Chen; Mostafa Fatemi; Azra Alizad
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2015-09       Impact factor: 2.725

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.