Comparison of global indices and test duration between two visual field analyzers: Octopus 300 and Topcon SBP-3000.

Jose Javier Garcia-Medina1, Manuel Garcia-Medina, Vicente Zanon-Moreno, Carlos Garcia-Maturana, Francisco Javier Cruz-Espinosa, Maria Dolores Pinazo-Duran.   

Abstract

PURPOSE: To compare the global indices and test duration as measured by Octopus 300 and Topcon SBP-3000 perimeters. To our knowledge, this is the first study performed in this way.
METHODS: Eighty eyes of 40 glaucomatous and ocular hypertensive patients with previous perimetric experience had visual field tests with Octopus 300 (TOP strategy) and TOPCON SBP-3000. All pairs of tests were performed randomly on separate days, but within 1 month of each other. Taking into account reliability factors of both perimetric examinations, 54 eyes of thirty patients were eligible. Only one eye from each patient was considered. Mean sensitivity (MS), mean defect (MD), loss of variance (sLV-Octopus- and LV -Topcon-) and test duration times were considered.
RESULTS: A significant difference was found between the global indices and duration times of the Octopus and the Topcon perimeters (p < 0.05; Wilcoxon test). Moderate degrees of correlation were obtained for MS (Spearman's rho = 0.635; p < 0.001) and MD (Spearman's rho = 0.592; p = 0.001) measurements. There was no correlation between sLV and LV (Spearman's rho = 0.181; p = 0.337). Agreements between pairs of global indices were low as measured by concordance correlation coefficient.
CONCLUSION: Global indices measured by the Octopus and Topcon perimeters are significantly different, so direct comparison of the measured values is not reliable. Because of the poor association and agreement between values obtained by these two perimeters, indirect comparison is also inadvisable.

Entities:  

Mesh:

Year:  2012        PMID: 22270219     DOI: 10.1007/s00417-012-1929-4

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  13 in total

1.  Comparison of two fast strategies, SITA Fast and TOP, for the assessment of visual fields in glaucoma patients.

Authors:  A J W King; A Taguri; A C Wadood; A Azuara-Blanco
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2002-05-15       Impact factor: 3.117

2.  The effect of perimetric experience in normal subjects.

Authors:  A Heijl; G Lindgren; J Olsson
Journal:  Arch Ophthalmol       Date:  1989-01

3.  A comparison of global indices between the Medmont Automated Perimeter and the Humphrey Field Analyzer.

Authors:  John Landers; Alok Sharma; Ivan Goldberg; Stuart Graham
Journal:  Br J Ophthalmol       Date:  2007-03-27       Impact factor: 4.638

4.  Conversion formulas between automated-perimetry indexes as measured by two different types of instrument.

Authors:  A Papp; K Kis; J Németh
Journal:  Ophthalmologica       Date:  2001 Mar-Apr       Impact factor: 3.250

5.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

6.  Glaucomatous visual fields. FASTPAC versus full threshold strategy of the Humphrey Field Analyzer.

Authors:  M Schaumberger; B Schäfer; B J Lachenmayr
Journal:  Invest Ophthalmol Vis Sci       Date:  1995-06       Impact factor: 4.799

7.  Formulas for conversion between Octopus and Humphrey threshold values and indices.

Authors:  T Zeyen; M Roche; L Brigatti; J Caprioli
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1995-10       Impact factor: 3.117

8.  Changes in differential threshold in patients with glaucoma during prolonged perimetry.

Authors:  A Heijl; S M Drance
Journal:  Br J Ophthalmol       Date:  1983-08       Impact factor: 4.638

9.  Evaluation of FASTPAC, a new strategy for threshold estimation with the Humphrey Field Analyzer, in a glaucomatous population.

Authors:  J G Flanagan; J M Wild; G E Trope
Journal:  Ophthalmology       Date:  1993-06       Impact factor: 12.079

10.  Static threshold visual field in glaucoma with the Fastpac algorithm of the Humphrey Field Analyser. Is the gain in examination time offset by any loss of information?

Authors:  J P Nordmann; P Denis; Y Nguer; D Mouton-Chopin; H Saraux
Journal:  Eur J Ophthalmol       Date:  1994 Apr-Jun       Impact factor: 1.922

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.