Conversion formulas between automated-perimetry indexes as measured by two different types of instrument.

A Papp1, K Kis, J Németh.   

Abstract

PURPOSE: Comparison between the results of measurements made using different types of automated-perimetry equipment is difficult, because of the lack of standardisation. The purpose of the study was to compare the two main indexes, mean deviation or mean defect (MD) and pattern standard deviation (PSD), as measured by two widely used instruments (Interzeag Octopus 500EZ and Humphrey Field Analyser 740) in the same group of patients. PATIENTS AND METHODS: Thirty eyes of 17 patients with different stages of primary open-angle glaucoma were tested, and for 25 eyes the indexes measured using program G1 of the Octopus perimetry instrument were compared with those from program C30-2 of the Humphrey Field Analyser.
RESULTS: Using the one-tailed Student t test, a significant difference was found between the MD and the PSD values of the Humphrey and of the Octopus perimetry equipments. However, subsequent use of the conversion formulas to adjust the measurements then indicated a good correlation between the two instruments (MD: r = 0.91; PSD: r = 0.79).
CONCLUSION: MD and PSD values measured by the Octopus and Humphrey perimetry instruments are based on measurements made under significantly different conditions, so the direct comparison of the indexes is not reliable. Nonetheless, empirically valid conversion formulas exist and can be used to compare the perimetric results for a patient tested with the two types of instrument. Copyright 2001 S. Karger AG, Basel

Entities:  

Mesh:

Year:  2001        PMID: 11244336     DOI: 10.1159/000050835

Source DB:  PubMed          Journal:  Ophthalmologica        ISSN: 0030-3755            Impact factor:   3.250


  7 in total

1.  [Functional diagnostic options for advanced and end stage glaucoma].

Authors:  A F Scheuerle; U Schiefer; K Rohrschneider
Journal:  Ophthalmologe       Date:  2012-04       Impact factor: 1.059

2.  Transcranial direct current stimulation affects visual perception measured by threshold perimetry.

Authors:  Antje Kraft; Jasper Roehmel; Manuel C Olma; Sein Schmidt; Kerstin Irlbacher; Stephan A Brandt
Journal:  Exp Brain Res       Date:  2010-11-03       Impact factor: 1.972

3.  Comparison of global indices and test duration between two visual field analyzers: Octopus 300 and Topcon SBP-3000.

Authors:  Jose Javier Garcia-Medina; Manuel Garcia-Medina; Vicente Zanon-Moreno; Carlos Garcia-Maturana; Francisco Javier Cruz-Espinosa; Maria Dolores Pinazo-Duran
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-01-25       Impact factor: 3.117

4.  Comparison of macular ganglion cell complex thickness to total retinal thickness ratio between Hungarian and Japanese eyes.

Authors:  Yoshiyuki Kita; Farzaneh Naghizadeh; Ritsuko Kita; Goji Tomita; Gábor Holló
Journal:  Jpn J Ophthalmol       Date:  2013-08-27       Impact factor: 2.447

5.  Contrasting effects of transcranial direct current stimulation on central and peripheral visual fields.

Authors:  Thiago L Costa; Mirella Gualtieri; Mirella T S Barboni; Rafael K Katayama; Paulo S Boggio; Dora F Ventura
Journal:  Exp Brain Res       Date:  2015-02-04       Impact factor: 1.972

6.  Diagnostic ability of Humphrey perimetry, Octopus perimetry, and optical coherence tomography for glaucomatous optic neuropathy.

Authors:  B Monsalve; A Ferreras; P Calvo; J A Urcola; M Figus; J Monsalve; P Frezzotti
Journal:  Eye (Lond)       Date:  2016-11-11       Impact factor: 3.775

7.  Axial Myopia Is Associated with Visual Field Prognosis of Primary Open-Angle Glaucoma.

Authors:  Chen Qiu; Shaohong Qian; Xinghuai Sun; Chuandi Zhou; Fanrong Meng
Journal:  PLoS One       Date:  2015-07-27       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.