Literature DB >> 22163763

Subtractive inhibition assay for the detection of E. coli O157:H7 using surface plasmon resonance.

Yixian Wang1, Zunzhong Ye, Chengyan Si, Yibin Ying.   

Abstract

A surface plasmon resonance (SPR) immunosensor was developed for the detection of E. coli O157:H7 by means of a new subtractive inhibition assay. In the subtractive inhibition assay, E. coli O157:H7 cells and goat polyclonal antibodies for E. coli O157:H7 were incubated for a short of time, and then the E. coli O157:H7 cells which bound antibodies were removed by a stepwise centrifugation process. The remaining free unbound antibodies were detected through interaction with rabbit anti-goat IgG polyclonal antibodies immobilized on the sensor chip using a BIAcore 3000 biosensor. The results showed that the signal was inversely correlated with the concentration of E. coli O157:H7 cells in a range from 3.0 × 10(4) to 3.0 × 10(8) cfu/mL with a detection limit of 3.0 × 10(4) cfu/mL. Compared with direct SPR by immobilizing antibodies on the chip surface to capture the bacterial cells and ELISA for E. coli O157:H7 (detection limit: both 3.0 × 10(5) cfu/mL in this paper), the detection limit of subtractive inhibition assay method was reduced by one order of magnitude. The method simplifies bacterial cell detection to protein-protein interaction, which has the potential for providing a practical alternative for the monitoring of E. coli O157:H7 and other pathogens.

Entities:  

Keywords:  E. coli O157:H7; SPR; subtractive inhibition assay

Mesh:

Substances:

Year:  2011        PMID: 22163763      PMCID: PMC3231628          DOI: 10.3390/s110302728

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


Introduction

In recent times, E. coli O157:H7 as one of the major foodborne pathogenic bacteria and as such has attracted considerable attention. According to the U.S. Center for Disease Control and Prevention (CDC), outbreak data and the known ability of the organism to be passed from person to person in nursing homes, day-care centers, and other personal care facilities, indicate that the presence of as few as 10 E. coli O157:H7 could result in disease. It has been reported that there may be about 73,000 infections and 61 deaths occurring due to E. coli O157:H7 each year in the United States [1] and therefore it is of utmost importance to develop rapid and sensitive methods for E. coli O157:H7 detection. By far, the most popular detection methods are culture and colony counting methods, polymerase chain reaction (PCR) and immunology-based methods and biosensors [2-6]. However, they are labor intensive and time consuming or professional operation limited. Biosensors, which incorporate a bioreceptor closely integrated with or connected to a transducer [7,8], have been proved to be a promising method for bacteria detection due to their portability, speed, sensitivity and possibility of on-the-spot detection [6,7,9], Surface plasmon resonance (SPR) biosensors are one kind of biosensor that has been widely used for bacterial detection [10-14]. A large number of direct SPR immunosensors have been used for the detection of bacterial cells by immobilizing antibodies directly on the chip surface to capture the bacterial cells [15-20]. Mazumdar immobilized antibodies on gold surface of each glass prism to capture S. typhimurium using the Plasmonic® SPR assay with a detection limit of 1.25 × 105 cfu/mL [20]. Subramanian reported that the detection limit of direct surface plasmon resonance assay for Escherichia coli O157:H7 detection was 106 cfu/mL [17]. The method based on the surface capture of cells has some limits to reduce the sensitivity of immunosensors [21-24]. Firstly, the effective penetration depth of the evanescent field which arises under conditions of total internal reflection is approximately 300 nm. It means that only refractive index changes occurring within the 300 nm distance from the surface will cause a change in the generated SPR signal. Bacteria such as E. coli O157:H7 with size of about 1 μm probably only interact with the top of the dextran layer that coats the gold surface and therefore only a small portion of the cell which is in close contact with the sensor surface will produce a measurable signal, which decreases the sensitivity of SPR for E. coli O157:H7 detection [21-24]. In addition, due to the large size of bacterial cells, direct cell binding requires that the cell-antibody binding affinity must be high to withstand the effect of shear force created by the laminar flow in the microflow channels [24]. Finally, Biacore instruments average the SPR angle over an area of approximately 0.25 mm2 on the sensor surface. As the sizes of bacterial cells are large, they will not evenly cover the area measured due to steric hindrance, which will decrease the signal response [24]. In this paper, to avoid the defects of SPR detection due to the size of bacteria, a new subtractive inhibition assay using SPR detection of E. coli O157:H7 was developed. In the proposed assay, E. coli O157:H7 cells and antibodies were incubated for a short of time, and the E. coli O157:H7 cells which bound antibodies were removed by a stepwise centrifugation process. Then the remaining free unbound antibodies were quantified through binding with anti-antibody immobilized on the sensor chip using BIAcore 3000 biosensor, which were inversely proportional to the E. coli O157:H7 cell concentration (Figure 1). This method simplifies bacterial cells detection to protein-protein interaction, which increases sensitivity of SPR for E. coli O157:H7 detection.
Figure 1.

Principle of a subtractive inhibition assay (SIA) using SPR.

Experimental Section

Regents

The reagents were obtained from the following sources: goat polyclonal antibody for E. coli O157:H7 and E. coli O157:H7 positive control were from KPL (Gaithersburg, MD, USA); E. coli DH5α (ATCC PTA-3137) was obtained from College of Food Science at Zhejiang University; Rabbit anti-goat IgG polyclonal antibody was from Boster (Wuhan, China); Sensor chip CM5, HBS running buffer (10 mM HEPES, 150 mM NaCl, 3.8 mM EDTA, 0.05% (v/v) Tween), 10 mM acetate buffer (pH 4.5) and 1 M ethanolamine (pH 8.5) were from GE Healthcare Bio-Sciences AB (Uppsala, Sweden); Phosphate buffered saline (PBS, pH 7.4), N-hydroxysuccinimide (NHS) and N-ethyl-N-(dimethylaminopropyl) carbodiimide hydrochloride (EDC) were from Sigma (St. Louis, MO, USA). All other regents were all chemical analytical grades.

ELISA for E. coli O157:H7 Detection

Microtiter plates were coated with 100 μL of 10 μg/mL of goat polyclonal antibodies for E. coli O157:H7 in phosphate buffered saline (PBS) overnight at 4 °C, and then washed three times with PBS + 0.05% Tween-20 (PBST). 300 μL 3% bovine serum albumin (BSA) was used to block unbound sites each plate for 30 min, followed washing with PBST for three times. One hundred μL of E. coli O157:H7 samples of different concentration from 3.0 × 102–3.0 × 108 cfu/mL were loaded onto each plate and incubated for 1 h at room temperature. The plates were washed three times with PBST and then 100 μL 1,000-fold decreasing dilutions of peroxidase-conjugated affinipure donkey anti-goat IgG (H+L) was added to each well. Following 30 min incubation and a final wash step, the plates were developed with TMB, stopped after 10 min with H2SO4 and read at a wavelength of 450 nm on a spectrophotometer.

Direct SPR by Immobilizing Antibodies on the Chip Surface to Capture the Bacterial Cells

After activation by injecting 70 μL mixture of 0.1 M NHS with 0.4 M EDC at 10 μL/min for 7 min, a 100 μg/mL goat polyclonal antibody for E. coli O157:H7 in 10 mM acetate buffer, pH 4.5, was injected over the chip surface at 10 μL/min for 10 min. Unreacted sites were subsequently deactivated by injecting 1 M ethanolamine, pH 8.5, at 10 μL/min for 7 min. Then, different concentration of diluted E. coli O157:H7 (3.0 × 101–3.0 × 108 cfu/mL) in running buffer were flowed for 5 min at 10 μL/min for direct detection. After the bacterial cells binding, the chip surface was regenerated by injecting 10 μL of 15 mM NaOH every time.

Subtractive Inhibition Assay

Free Antibody Centrifugation Separation

Three hundred μL of a 50 μg/mL anti-E. coli O157:H7 polyclonal antibody solution diluted in PBS, pH 7.4, was mixed with 300 μL of E. coli O157:H7 with series of concentration from 101–108 cfu/mL. The final anti-E. coli O157:H7 polyclonal antibody concentration was 25 μg/mL. Each mixture was incubated in rotating for 1 h at room temperature, followed stepwise centrifugation process for 2 min intervals at 50, 200, 400, 800, 1,200, 1,800 and 3,200 × g to separate the cells bound antibodies from the remaining free unbound antibodies. Get the remaining free unbound antibodies by drawing 500 μL supernatant fluids from the centrifugal tube.

Antibody Immobilization and Assay Setup

The remaining free unbound antibodies were quantified using a BIAcore 3000TM instrument. The CM 5 sensor chip was activated by injecting 70 μL mixture of 0.1 M NHS with 0.4 M EDC at 10 μL/min for 7 min. A 100 μg/mL rabbit anti-goat (Fab portion) polyclonal antibodies in 10 mM acetate buffer, pH 4.5, were injected over the chip surface at 10 μL/min for 10 min. Unreacted sites were subsequently deactivated by injecting 1 M ethanolamine, pH 8.5, at 10 μL/min for 7 min. Then the remaining free unbound antibodies were injected in the chip surface to bind with the secondary antibodies. After the antibodies binding, the chip surface was regenerated by injecting 10 μL of 15 mM NaOH every time.

Specificity Testing

The specificity of the biosensor was confirmed by detecting E. coli DH5α, one of E. coli serotypes. 300 μL 3.0 × 108 cfu/mL E. coli DH5α mixed with 300 μL of a 50 μg/mL anti-E. coli O157:H7 polyclonal antibody solution. And then the mixture was with the same free antibody centrifugation separation and detection method as E. coli O157:H7 detection.

Results and Discussion

ELISA for Validity of Antibody for E. coli O157:H7 Detection

ELISA analysis was used as a conventional immunoassay method for comparison with the biosensor method, and to observe the quality of the E. coli O157:H7 antibodies used in the immunosensor. Eight concentrations of E. coli O157:H7 (3.0 × 101 to 3.0 × 108 cfu/mL) were assayed with the ELISA, and the results are shown in Figure 2. From 3.0 × 105 cfu/mL, increasing the bacteria concentration resulted in an increase of OD value (2.9647, 2.5363, 1.4685 and 0.6323 observed with 3.0 × 108, 3.0 × 107, 3.0 × 106, 3.0 × 105 cfu/mL E. coli O157:H7 cells respectively). And the standard deviation (SD) for each point of the standard curves (3.0 × 108, 3.0 × 107, 3.0 × 106, 3.0 × 105 cfu/mL and control) were 1.5%, 2.0%, 1.2%, 0.9%, 2.2%, respectively. The results indicated that the purchased anti-E. coli O157:H7 antibody allowed the detection of E. coli O157:H7. The detection limit of this ELISA was 3.0 × 105 cfu/mL of E. coli O157:H7.
Figure 2.

Detection of E. coli O157:H7 based on ELISA.

Direct SPR Immobilizing Antibodies on the Chip Surface to Capture the Bacterial Cells

The first step of the direct assay was the immobilization of the capture antibodies on the CM5 chip surface, and a significant increase in the sensorgram signal (RU = 17,000) was observed, indicating a stable binding interaction between the CM5 chip surface and the captured antibody. The incubation time was only 10 min. Through immobilizing antibodies directly on the chip surface to capture the bacterial cells, a detectable change in RU due to its binding to the bacterial cells was obtained only at and above a cell concentration of 3.0 × 105 cfu/mL. 47, 76, 160 and 288 RU observed from binding with 3.0 × 105, 3.0 × 106, 3.0 × 107, 3.0 × 108 cfu/mL E. coli O157:H7 cells, respectively. The standard deviation (SD) for each point of the standard curves (3.0 × 108, 3.0 × 107, 3.0 × 106, 3.0 × 105 cfu/mL and control) were 8.1%, 2.4%, 4.7%, 3.3%, 3.1%, respectively. The lower limit of detection (LLD) is defined as the concentration of cells resulting in a detection signal that is the average value of the detection signal obtained due to control plus 3 times the standard deviation [20]. The detection range was between 3.0 × 105 and 3.0 × 108 cfu/mL with a limit of detection of 3.0 × 105 cfu/mL.

Centrifugation Effect

Haines et al. have reported removal of the cells-bound antibodies by a method such as filtration, however, that method was not suitable for complex and viscous matrices [25]. It has been reported that separate the remaining free antibodies from the bacterial cells-bound antibodies by centrifugation for 1 min intervals at 50, 200, 450, 800, 1,200, 1,800 and 3,200 × g was possible [23]. Therefore, in this paper, a stepwise centrifugation process has been applied to separate the remaining free unbound antibodies from the bacterial cells that bound antibodies. But according to the above method, the signal of 108 cfu/mL was larger than the signal of 3.0 × 107 cfu/mL, possibly because the sizes of E. coli O157:H7 cells (about 1 μm) are smaller than Listeria monocytogenes cells, resulting in the need for greater centrifugal force and longer centrifugation times, so we increased the centrifugation time by centrifuging for 2 min intervals at 50, 200, 400, 800, 1,200, 1,800 and 3,200 × g, confirming the method’s principle, resulting in a decreasing signal with increasing E. coli O157:H7 bacteria cell concentration. Optimization of the centrifugation time according to the size of E. coli O157:H7 cells was the key step for the success of this experiment.

Sensor Chip Preparation

The subtractive inhibition assay was further implemented into a Biacore® 3000 SPR sensor. The scheme for the sensor setup used in this work is shown in Figure 1. The dextran layer was immobilized on the gold surface of CM5 chip beforehand. Then the surface was activated with an NHS-EDC solution, washed with running buffer, and a goat anti-mouse IgG polyclonal antibody was immobilized to a CM5 chip surface (17,556 RU) through an amide bond, as shown in Figure 3. The binding response was compared to that of a reference channel chip surface (same activation using EDC and NHS, blocking using ethanolamine, but without immobilizing secondary antibody), which gave insignificant binding, thereby illustrating the specificity of the binding response (data not shown).
Figure 3.

Plot of sensor chip preparation: (A) activation. (B) rabbit anti-goat IgG polyclonal antibodies immobilization. (C) ethanolamine blocking.

Surface Regeneration

The rabbit anti-goat IgG polyclonal antibodies were immobilized on the CM5 chip and optimal regeneration conditions were investigated. The regeneration solution should remove the remaining free antibodies from the anti-goat IgG polyclonal antibodies without affecting the activity of the secondary antibodies. The regeneration solution of 10 μL 15 mM NaOH at a flow rate of 10 μL/min was sufficient for surface regeneration. The surface activity of the chip decreased 6.0% by repeated 50 binding and regeneration cycles, which proved the excellent long-term surface performance.

Sensor Performance

In the subtractive inhibition assay, different concentrations of E. coli O157:H7 cells and antibodies were incubated and then the E. coli O157:H7 cells which bound antibody were removed by a stepwise centrifugation process. The samples of free unbound antibodies were quantified in duplicate out of order through binding with anti-antibody immobilized on the sensor chip using BIAcore 3000 biosensor. As expected, and as shown in Figure 4, the binding responses were inversely proportional to the concentration of the E. coli O157:H7 cells, which verified the subtractive inhibition assay principle. 166, 285, 373, 389, 419 and 449 RU observed from supernatants from samples incubated with 3.0 × 108, 3.0 × 107, 3.0 × 106, 3.0 × 105, 3.0 × 104 and 0 cfu/mL E. coli O157:H7 cells, respectively.
Figure 4.

Overlay plot from one assay demonstrating that the binding response is inversely proportional to the E. coli O157:H7 concentration: (A) control sample. (B) 3.0 × 104 cfu/mL. (C) 3.0 × 105 cfu/mL. (D) 3.0 × 106 cfu/mL. (E) 3.0 × 107 cfu/mL. (F) 3.0 × 108 cfu/mL.

The resonance dips of different bacterial cells concentration from 3.0 × 108 to 3.0 × 104 were 283, 164, 76, 60 and 30, respectively. The signal was correlated with the concentration of E. coli O157:H7 cells in a range from 3.0 × 108 to 3.0 × 104 cfu/mL. Normalized data (R/R0) was obtained with the average response for each sample(R) divided by the average of the control (sample without bacterial cells, R0). A calibration curve using normalized data (R/R0) plotted against the E. coli O157:H7 cells concentrations, was constructed (Figure 5). According to Figure 5, the range of detection was found to be approximately 3.0 × 104–3.0 × 108 cfu/mL. The standard deviations for each point of the standard curves (3.0 × 108, 3.0 × 107, 3.0 × 106, 3.0 × 105, 3.0 × 104 cfu/mL) were 7.6%, 1.7%, 3.1%, 5.1%, 1.7%, respectively, which illustrates the good reproducibility of the immunoassay.
Figure 5.

The plot between SPR response and bacterial cells concentration.

In the subtractive inhibition assay, the lower limit of detection (LLD) is defined as the concentration of cells resulting in a detection signal that is the average value of the detection signal obtained due to control minus three times the standard deviation. The relative standard deviation for control samples was 3.5%. Therefore, the detection limit of SPR detection based on subtractive inhibition assay is 3.0 × 104 cfu/mL, which is one order of magnitude less compared with direct SPR detection immobilizing antibodies on the chip surface to capture the bacterial cells and ELISA method for E. coli O157:H7 detection (the detection limit: both 3.0 × 105 cfu/mL). That method is of significance because there have been problems encountered in the use of SPR for bacterial cell detection. It simplifies bacterial cells detection to protein-protein interaction, which avoid the defects of SPR detection due to the size of bacteria, including the cell size exceeding the 300 nm range of the evanescent field wave, high fluid force acting on captured cells, and limited mass transfer. Therefore the subtractive inhibition assay has improved the sensitivity of the sensors for E. coli O157:H7 detection. The traditional culture and colony counting method has been a practical method for the detection and identification of E. coli O157:H7 in food, including microbiological culturing and isolation of the pathogen, followed by confirmation by biochemical and serological tests, which takes up to 5–7 days to get a confirmed result [26]. Although it can provide reliable results, it is time consuming, which is not suitable for rapid assay for E. coli O157:H7 and other pathogens in the food industry. The polymerase chain reaction (PCR) and enzyme-linked immunosorbent assay (ELISA) are a lot less time-consuming than the traditional culture and colony counting methods, and usually take 4–5 hours to produce detection result [6,27]. Compared with above mentioned methods for E. coli O157:H7 detection, the subtractive inhibition assay based on SPR, excluding sample incubation, centrifugation, secondary antibodies immobilization and interaction time, was less than 2 h. It has reduced the testing time effectively and can detect multiple samples in succession with the process of regeneration, which have the potential for providing a practical alternative for the monitoring of E. coli O157:H7 and other pathogens in the food industry.

Specificity of the Biosensor

The specificity of the biosensor was confirmed by detecting E. coli DH5α, one of E. coli serotypes. For E. coli DH5α (the concentration of 3.0 × 108 cfu/mL), the response signal was 412, while the response signal of E. coli O157:H7 cells (3.0 × 108 cfu/mL) was 166. The E. coli O157:H7 gave a binding response of 283 and weak reactivity was found with E. coli DH5α with a signal dip of 37 (Figure 6). This suggests that E. coli DH5α did not have strong binding force with anti-E. coli O157:H7 polyclonal antibodies as E. coli O157:H7, demonstrating specificity of the biosensor in the presence of non-target bacterial cells.
Figure 6.

The cross-reactivity studies plot between E. coli O157:H7 and E. coli DH5α (3.0 × 108).

Conclusions

A new subtractive inhibition assay using SPR detection of E. coli O157:H7 has been clearly established. Unlike the direct SPR for bacterial detection by immobilizing the antibody on the chip surface, in the subtractive inhibition assay, the remaining free unbound antibodies obtained by stepwise centrifugation process after E. coli O157:H7 cells and relevant antibodies incubation, were detected through interaction with secondary antibody immobilized on the sensor chip using a BIAcore 3000 biosensor. The results showed that the signal was inversely correlated with the concentration of E. coli O157:H7 cells in a range from 3.0 × 104 to 3.0 × 108 cfu/mL. The detection limit was 3.0 × 104 cfu/mL, compared with direct SPR by immobilizing antibodies on the chip surface to capture the bacterial cells and ELISA methods with the same bacterial cells and relevant antibodies (detection limit: 3.0 × 105 cfu/mL in this paper), the detection limit of subtractive inhibition assay method was reduced by one order of magnitude. The subtractive inhibition assay method simplified bacterial cell detection to protein-protein interaction, and can increase the sensitivity of SPR biosensors for E. coli O157:H7 detection. The assay time required for sample detection was less than 2 h and the sample requirement for each analysis was only 5 μL, which reduced the testing time and sample volume effectively compared with the traditional methods for bacterial detection. The present assay can be used in automated mode with the ability to rapidly analyze a large number of samples and has specificity in the presence of non-target bacterial cells. The subtractive inhibition assay method based on SPR can also be applied to the detection of other organisms suffering from oversize effects, such as virus, bacteria, fungal, cells and so on. The simplicity of this SPR-based subtractive inhibition assay for detection of E. coli O157:H7 clearly demonstrates the potential of the SPR technique to be used as a rapid detection tool for food microbiological safety.
  19 in total

Review 1.  Methodology for detection and typing of foodborne microorganisms.

Authors:  E de Boer; R R Beumer
Journal:  Int J Food Microbiol       Date:  1999-09-15       Impact factor: 5.277

Review 2.  Biosensor technologies for detecting microbiological foodborne hazards.

Authors:  Robert H Hall
Journal:  Microbes Infect       Date:  2002-04       Impact factor: 2.700

Review 3.  A review of conventional detection and enumeration methods for pathogenic bacteria in food.

Authors:  Kiev S Gracias; John L McKillip
Journal:  Can J Microbiol       Date:  2004-11       Impact factor: 2.419

4.  Quantitative and simultaneous detection of four foodborne bacterial pathogens with a multi-channel SPR sensor.

Authors:  Allen D Taylor; Jon Ladd; Qiuming Yu; Shengfu Chen; Jirí Homola; Shaoyi Jiang
Journal:  Biosens Bioelectron       Date:  2006-04-25       Impact factor: 10.618

Review 5.  Pathogen detection: a perspective of traditional methods and biosensors.

Authors:  Olivier Lazcka; F Javier Del Campo; F Xavier Muñoz
Journal:  Biosens Bioelectron       Date:  2006-08-28       Impact factor: 10.618

Review 6.  Rapid detection of food-borne pathogenic bacteria.

Authors:  B Swaminathan; P Feng
Journal:  Annu Rev Microbiol       Date:  1994       Impact factor: 15.500

7.  Rapid determination of Phytophthora infestans sporangia using a surface plasmon resonance immunosensor.

Authors:  Peter Skottrup; Mogens Nicolaisen; Annemarie F Justesen
Journal:  J Microbiol Methods       Date:  2006-12-08       Impact factor: 2.363

8.  A mixed self-assembled monolayer-based surface plasmon immunosensor for detection of E. coli O157:H7.

Authors:  Anand Subramanian; Joseph Irudayaraj; Thomas Ryan
Journal:  Biosens Bioelectron       Date:  2006-01-15       Impact factor: 10.618

9.  A generic approach for the detection of whole Listeria monocytogenes cells in contaminated samples using surface plasmon resonance.

Authors:  Paul Leonard; Stephen Hearty; John Quinn; Richard O'Kennedy
Journal:  Biosens Bioelectron       Date:  2004-05-15       Impact factor: 10.618

Review 10.  Food-related illness and death in the United States.

Authors:  P S Mead; L Slutsker; V Dietz; L F McCaig; J S Bresee; C Shapiro; P M Griffin; R V Tauxe
Journal:  Emerg Infect Dis       Date:  1999 Sep-Oct       Impact factor: 6.883

View more
  16 in total

1.  Generalized figure of merit for plasmonic dip measurement-based surface plasmon resonance sensors.

Authors:  Treesukon Treebupachatsakul; Apivitch Boosamalee; Kamejira Chaithatwanitch; Suejit Pechprasarn
Journal:  Biomed Opt Express       Date:  2022-03-02       Impact factor: 3.562

2.  Foodborne Pathogen Screening Using Magneto-fluorescent Nanosensor: Rapid Detection of E. Coli O157:H7.

Authors:  Tyler Shelby; Shoukath Sulthana; James McAfee; Tuhina Banerjee; Santimukul Santra
Journal:  J Vis Exp       Date:  2017-09-17       Impact factor: 1.355

3.  An aptamer-based biosensor for colorimetric detection of Escherichia coli O157:H7.

Authors:  Wenhe Wu; Jie Zhang; Meiqin Zheng; Yuhong Zhong; Jie Yang; Yuhong Zhao; Wenping Wu; Wei Ye; Jie Wen; Qi Wang; Jianxin Lu
Journal:  PLoS One       Date:  2012-11-07       Impact factor: 3.240

Review 4.  New trends in impedimetric biosensors for the detection of foodborne pathogenic bacteria.

Authors:  Yixian Wang; Zunzhong Ye; Yibin Ying
Journal:  Sensors (Basel)       Date:  2012-03-12       Impact factor: 3.576

Review 5.  Rapid methods for the detection of foodborne bacterial pathogens: principles, applications, advantages and limitations.

Authors:  Jodi Woan-Fei Law; Nurul-Syakima Ab Mutalib; Kok-Gan Chan; Learn-Han Lee
Journal:  Front Microbiol       Date:  2015-01-12       Impact factor: 5.640

6.  Fiber-Optic Fluoroimmunoassay System with a Flow-Through Cell for Rapid On-Site Determination of Escherichia coli O157:H7 by Monitoring Fluorescence Dynamics.

Authors:  Kumiko Miyajima; Tomoyuki Koshida; Takahiro Arakawa; Hiroyuki Kudo; Hirokazu Saito; Kazuyoshi Yano; Kohji Mitsubayashi
Journal:  Biosensors (Basel)       Date:  2013-03-08

7.  A Label-Free Impedance Immunosensor Using Screen-Printed Interdigitated Electrodes and Magnetic Nanobeads for the Detection of E. coli O157:H7.

Authors:  Ronghui Wang; Jacob Lum; Zach Callaway; Jianhan Lin; Walter Bottje; Yanbin Li
Journal:  Biosensors (Basel)       Date:  2015-12-15

Review 8.  Detection Methodologies for Pathogen and Toxins: A Review.

Authors:  Md Eshrat E Alahi; Subhas Chandra Mukhopadhyay
Journal:  Sensors (Basel)       Date:  2017-08-16       Impact factor: 3.576

Review 9.  A review on detection methods used for foodborne pathogens.

Authors:  B Priyanka; Rajashekhar K Patil; Sulatha Dwarakanath
Journal:  Indian J Med Res       Date:  2016-09       Impact factor: 2.375

10.  Aptasensors for rapid detection of Escherichia coli O157:H7 and Salmonella typhimurium.

Authors:  Wen-He Wu; Min Li; Yue Wang; Hou-Xian Ouyang; Lin Wang; Ci-Xiu Li; Yu-Chen Cao; Qing-He Meng; Jian-Xin Lu
Journal:  Nanoscale Res Lett       Date:  2012-11-29       Impact factor: 4.703

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.