An unusual surgical success with a defective penile prosthesis.

Chen Zhao1, Bo Ram Choi, Young Beom Jeong, Jong Kwan Park.   

Abstract

A defective penile prosthesis is disconcerting for the surgeon performing an implantation in a patient with erectile dysfunction. We became aware of the defective Titan penile prosthesis just after insertion of the left cylinder, which had an abnormal rotation (180 degrees) due to abnormal rotated tubing, which occurred at the manufacturing plant. We successfully performed the implantation by cutting and re-connecting the tubing between the cylinder and pump with a spare connector. When the surgeon encounters a defective penile prosthesis with abnormal rotation of the cylinder due to abnormal rotated tubing in the operative field, disconnection and re-connection of the tubing using an extra connector are alternative cost-effective maneuvers.

Entities:  

Year:  2011        PMID: 22154174      PMCID: PMC3235186          DOI: 10.5489/cuaj.10162

Source DB:  PubMed          Journal:  Can Urol Assoc J        ISSN: 1911-6470            Impact factor:   1.862


  6 in total

1.  Prevention of urethral stricture in insertion of an inflatable penile prosthesis.

Authors:  Sung Won Lee; Bong Hee Park; Joo Hyun Lim; Wan Shou Cui; Myung Ki Kim; Jong Kwan Park
Journal:  Int J Urol       Date:  2008-02       Impact factor: 3.369

2.  An unusual success with simultaneous urethral repair and reimplantation of penile prosthesis in a patient with urethral stricture induced by rotated tubing.

Authors:  Wan Shou Cui; Sang Deuk Kim; Kyung Soo Choi; Chen Zhao; Jong Kwan Park
Journal:  J Sex Med       Date:  2009-03-30       Impact factor: 3.802

3.  Outcome of penile prosthesis implantation for treating erectile dysfunction: experience with 504 procedures.

Authors:  Andrea Minervini; David J Ralph; John P Pryor
Journal:  BJU Int       Date:  2006-01       Impact factor: 5.588

Review 4.  Diagnosis, treatment and prevention of penile prosthesis infection.

Authors:  C C Carson
Journal:  Int J Impot Res       Date:  2003-10       Impact factor: 2.896

5.  Risk factors for penile prosthetic infection.

Authors:  J P Jarow
Journal:  J Urol       Date:  1996-08       Impact factor: 7.450

6.  Long-term survival of inflatable penile prostheses: single surgical group experience with 2,384 first-time implants spanning two decades.

Authors:  Steven K Wilson; John R Delk; Emad A Salem; Mario A Cleves
Journal:  J Sex Med       Date:  2007-07       Impact factor: 3.802

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.