Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors.

G Cranmer-Sargison1, S Weston, J A Evans, N P Sidhu, D I Thwaites.   

Abstract

PURPOSE: The goal of this work was to implement a recently proposed small field dosimetry formalism [Alfonso et al., Med. Phys. 35(12), 5179-5186 (2008)] for a comprehensive set of diode detectors and provide the required Monte Carlo generated factors to correct measurement.
METHODS: Jaw collimated square small field sizes of side 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, and 3.0 cm normalized to a reference field of 5.0 cm × 5.0 cm were used throughout this study. Initial linac modeling was performed with electron source parameters at 6.0, 6.1, and 6.2 MeV with the Gaussian FWHM decreased in steps of 0.010 cm from 0.150 to 0.100 cm. DOSRZnrc was used to develop models of the IBA stereotactic field diode (SFD) as well as the PTW T60008, T60012, T60016, and T60017 field diodes. Simulations were run and isocentric, detector specific, output ratios (OR(det)) calculated at depths of 1.5, 5.0, and 10.0 cm. This was performed using the following source parameter subset: 6.1 and 6.2 MeV with a FWHM = 0.100, 0.110, and 0.120 cm. The source parameters were finalized by comparing experimental detector specific output ratios with simulation. Simulations were then run with the active volume and surrounding materials set to water and the replacement correction factors calculated according to the newly proposed formalism.
RESULTS: In all cases, the experimental field size widths (at the 50% level) were found to be smaller than the nominal, and therefore, the simulated field sizes were adjusted accordingly. At a FWHM = 0.150 cm simulation produced penumbral widths that were too broad. The fit improved as the FWHM was decreased, yet for all but the smallest field size worsened again at a FWHM = 0.100 cm. The simulated OR(det) were found to be greater than, equivalent to and less than experiment for spot size FWHM = 0.100, 0.110, and 0.120 cm, respectively. This is due to the change in source occlusion as a function of FWHM and field size. The corrections required for the 0.5 cm field size were 0.95 (± 1.0%) for the SFD, T60012 and T60017 diodes and 0.90 (± 1.0%) for the T60008 and T60016 diodes-indicating measured output ratios to be 5% and 10% high, respectively. Our results also revealed the correction factors to be the same within statistical variation at all depths considered.
CONCLUSIONS: A number of general conclusions are evident: (1) small field OR(det) are very sensitive to the simulated source parameters, and therefore, rigorous Monte Carlo linac model commissioning, with respect to measurement, must be pursued prior to use, (2) backscattered dose to the monitor chamber should be included in simulated OR(det) calculations, (3) the corrections required for diode detectors are design dependent and therefore detailed detector modeling is required, and (4) the reported detector specific correction factors may be applied to experimental small field OR(det) consistent with those presented here.

Mesh:

Year:  2011        PMID: 22149841     DOI: 10.1118/1.3658572

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam.

Authors:  Kyohei Fukata; Satoru Sugimoto; Chie Kurokawa; Akito Saito; Tatsuya Inoue; Keisuke Sasai
Journal:  Radiol Phys Technol       Date:  2018-04-04

2.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

3.  Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

Authors:  Yujiao Qin; Hualiang Zhong; Ning Wen; Karen Snyder; Yimei Huang; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

4.  Dosimetric characterization of Elekta stereotactic cones.

Authors:  Egor Borzov; Alexander Nevelsky; Raquel Bar-Deroma; Itzhak Orion
Journal:  J Appl Clin Med Phys       Date:  2017-12-20       Impact factor: 2.102

5.  Survey of 5 mm small-field output factor measurements in Australia.

Authors:  Christopher P Oliver; Duncan J Butler; Viliami Takau; Ivan Williams
Journal:  J Appl Clin Med Phys       Date:  2018-01-25       Impact factor: 2.102

6.  Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system.

Authors:  Jason Y Cheng; Holly Ning; Barbara C Arora; Ying Zhuge; Robert W Miller
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

7.  A Novel GPU-based Fast Monte Carlo Photon Dose Calculating Method for Accurate Radiotherapy Treatment Planning.

Authors:  Karbalaee M; Shahbazi-Gahrouei D; Tavakoli M B
Journal:  J Biomed Phys Eng       Date:  2020-06-01

8.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

9.  Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator.

Authors:  Paula Monasor Denia; María Del Carmen Castellet García; Carla Manjón García; Juan David Quirós Higueras; Noelia de Marco Blancas; Jorge Bonaque Alandí; Xavier Jordi Juan Senabre; Agustín Santos Serra; Juan López-Tarjuelo
Journal:  PLoS One       Date:  2019-03-11       Impact factor: 3.240

10.  Characteristics of a novel treatment system for linear accelerator-based stereotactic radiosurgery.

Authors:  Ning Wen; Haisen Li; Kwang Song; Karen Chin-Snyder; Yujiao Qin; Jinkoo Kim; Maria Bellon; Misbah Gulam; Stephen Gardner; Anthony Doemer; Suneetha Devpura; James Gordon; Indrin Chetty; Farzan Siddiqui; Munther Ajlouni; Robert Pompa; Zane Hammoud; Michael Simoff; Steven Kalkanis; Benjamin Movsas; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.