Keratometry evaluations with the Pentacam high resolution in comparison with the automated keratometry and conventional corneal topography.

László Módis1, Eszter Szalai, Bence Kolozsvári, Gábor Németh, Attila Vajas, András Berta.   

Abstract

PURPOSE: To determine the reliability and repeatability of keratometry (K) measurements obtained with the Pentacam high resolution (HR), automated keratometry, and corneal topography systems.
METHODS: The right eyes of 46 healthy subjects were examined prospectively. Keratometry measurements in the flat (Kf) and steep (Ks) meridians were taken by 2 independent investigators with the Pentacam HR (Oculus, Wetzlar, Germany) followed by automated keratorefractometry (KR-8100; Topcon, Tokyo, Japan), and corneal topography (TMS-4; Tomey, Erlangen, Germany).
RESULTS: The mean K readings of the Pentacam HR, automated keratometry, and corneal topography were 43.40/43.34 diopter (D), 43.99/43.98 D, and 43.80/43.83 D, respectively. The difference between the values was statistically significant (P < 0.0001, repeated measures analysis of variance). Strong significant correlation was observed between the Pentacam HR and keratometry (Kf: r = 0.952/0.954; Ks: r = 0.845, Spearman rank test), and Pentacam HR and corneal topography (Kf: r = 0.933/0.930; Ks: r = 0.838/0.829) (P < 0.0001). No significant difference was presented between the 2 investigators for any of the instruments (P = 0.215-0.983). Moreover, high correlation was found between the K readings of the observers (interoperator intraclass correlation coefficients ranged from 0.95 to 0.99).
CONCLUSIONS: The Pentacam HR provided reliable K measurements in clinical practice in comparison with an automated keratometer and a corneal topographer. Based on the results, for patient follow-up, one keratometry device is recommended.

Entities:  

Mesh:

Year:  2012        PMID: 22081146     DOI: 10.1097/ICO.0b013e318204c666

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  8 in total

1.  Longitudinal change of refraction over at least 5 years in 15,000 patients.

Authors:  David Goldblum; Annette Brugger; Andreas Haselhoff; Stefanie Schmickler
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-11-28       Impact factor: 3.117

2.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

3.  Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOL Master, and Javal Keratometer.

Authors:  Zahra Dehnavi; Mehdi Khabazkhoob; Ali Mirzajani; Mahmood Jabbarvand; Abbasali Yekta; Ebrahim Jafarzadehpur
Journal:  Middle East Afr J Ophthalmol       Date:  2015 Apr-Jun

4.  Comparison of corneal curvature and anterior chamber depth measurements using the manual keratometer, Lenstar LS 900 and the Pentacam.

Authors:  Omür Ö Uçakhan; Veysel Akbel; Zeynep Bıyıklı; Ayfer Kanpolat
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

5.  Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography.

Authors:  Onur Polat; Zeki Baysal; Serkan Özcan; Sibel İnan; Ümit Übeyt İnan
Journal:  Turk J Ophthalmol       Date:  2016-12-01

Review 6.  Pentacam® Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part I.

Authors:  Mahsaw N Motlagh; Majid Moshirfar; Michael S Murri; David F Skanchy; Hamed Momeni-Moghaddam; Yasmyne C Ronquillo; Phillip C Hoopes
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019

7.  Agreement between Pentacam and handheld Auto-Refractor/Keratometer for keratometry measurement.

Authors:  Hassan Hashemi; Samira Heydarian; Abbas Ali Yekta; Mohamadreza Aghamirsalim; Mahin Ahmadi-Pishkuhi; Mehrnaz Valadkhan; Hadi Ostadimoghaddam; Ahmad Ahmadzadeh Amiri; Mehdi Khabazkhoob
Journal:  J Optom       Date:  2019-07-10

8.  Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.

Authors:  Jing Dong; Maolong Tang; Yaqin Zhang; Yading Jia; Haining Zhang; Zhijie Jia; Xiaogang Wang
Journal:  PLoS One       Date:  2015-11-17       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.