Radiation dose savings for adult pulmonary embolus 64-MDCT using bismuth breast shields, lower peak kilovoltage, and automatic tube current modulation.

Lynne M Hurwitz1, Terry T Yoshizumi, Philip C Goodman, Rendon C Nelson, Greta Toncheva, Giao B Nguyen, Carolyn Lowry, Colin Anderson-Evans.   

Abstract

OBJECTIVE: The purpose of this study was to assess whether radiation dose savings using a lower peak kilovoltage (kVp) setting, bismuth breast shields, and automatic tube current modulation could be achieved while preserving the image quality of MDCT scans obtained to assess for pulmonary embolus (PE).
MATERIALS AND METHODS: CT angiography (CTA) examinations were performed to assess for the presence or absence of pulmonary artery emboli using a 64-MDCT scanner with automatic tube current modulation (noise level=10 HU), two kVp settings (120 and 140 kVp), and bismuth breast shields. Absorbed organ doses were measured using anthropomorphic phantoms and metal oxide semiconductor field effect transistor (MOSFET) detectors. Image quality was assessed quantitatively as well as qualitatively in various anatomic sites of the thorax.
RESULTS: Using a lower kVp (120 vs 140 kVp) and automatic tube current modulation resulted in a dose savings of 27% to the breast and 47% to the lungs. The use of a lower kVp (120 kVp), automatic tube current modulation, and bismuth shields placed directly on the anterior chest wall reduced absorbed breast and lung doses by 55% and 45%, respectively. Qualitative assessment of the images showed no change in image quality of the lungs and mediastinum when using a lower kVp, bismuth shields, or both.
CONCLUSION: The use of bismuth breast shields together with a lower kVp and automatic tube current modulation will reduce the absorbed radiation dose to the breast and lungs without degradation of image quality to the organs of the thorax for CTA detection of PE.

Entities:  

Mesh:

Year:  2009        PMID: 19098206     DOI: 10.2214/AJR.08.1066

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  23 in total

1.  Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Authors:  Andrew J Einstein; Carl D Elliston; Daniel W Groves; Bin Cheng; Steven D Wolff; Gregory D N Pearson; M Robert Peters; Lynne L Johnson; Sabahat Bokhari; Gary W Johnson; Ketan Bhatia; Theodore Pozniakoff; David J Brenner
Journal:  J Nucl Cardiol       Date:  2011-11-09       Impact factor: 5.952

2.  Ventilation/perfusion SPECT--an essential but underrated method for diagnosis of pulmonary embolism and other diseases.

Authors:  Marika Bajc; Björn Jonson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03-31       Impact factor: 9.236

3.  EANM guidelines for ventilation/perfusion scintigraphy : Part 2. Algorithms and clinical considerations for diagnosis of pulmonary emboli with V/P(SPECT) and MDCT.

Authors:  M Bajc; J B Neilly; M Miniati; C Schuemichen; M Meignan; B Jonson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-09       Impact factor: 9.236

Review 4.  SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT.

Authors:  Sandra S Halliburton; Suhny Abbara; Marcus Y Chen; Ralph Gentry; Mahadevappa Mahesh; Gilbert L Raff; Leslee J Shaw; Jörg Hausleiter
Journal:  J Cardiovasc Comput Tomogr       Date:  2011 Jul-Aug

5.  Radiological patterns of primary graft dysfunction after lung transplantation evaluated by 64-multi-slice computed tomography: a descriptive study.

Authors:  Esther Okeke Belmaati; Ida Steffensen; Claus Jensen; Klaus F Kofoed; Jann Mortensen; Michael B Nielsen; Martin Iversen
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-02-28

6.  Differences in behavior of tube current modulation techniques for thoracic CT examinations between male and female anthropomorphic phantoms.

Authors:  Kosuke Matsubara; Pei-Jan Paul Lin; Atsushi Fukuda; Kichiro Koshida
Journal:  Radiol Phys Technol       Date:  2014-05-27

7.  CT pulmonary angiography using organ dose modulation with an iterative reconstruction algorithm and 3D Smart mA in different body mass indices: image quality and radiation dose.

Authors:  Yongxia Zhao; Ziwei Zuo; Shujie Cheng; Yanmin Wu
Journal:  Radiol Med       Date:  2018-05-03       Impact factor: 3.469

8.  Patient-specific radiation dose and cancer risk estimation in CT: part II. Application to patients.

Authors:  Xiang Li; Ehsan Samei; W Paul Segars; Gregory M Sturgeon; James G Colsher; Greta Toncheva; Terry T Yoshizumi; Donald P Frush
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

9.  Dose to radiosensitive organs during routine chest CT: effects of tube current modulation.

Authors:  Erin Angel; Nazanin Yaghmai; Cecilia Matilda Jude; John J DeMarco; Christopher H Cagnon; Jonathan G Goldin; Cynthia H McCollough; Andrew N Primak; Dianna D Cody; Donna M Stevens; Michael F McNitt-Gray
Journal:  AJR Am J Roentgenol       Date:  2009-11       Impact factor: 3.959

Review 10.  Modelling the physics in the iterative reconstruction for transmission computed tomography.

Authors:  Johan Nuyts; Bruno De Man; Jeffrey A Fessler; Wojciech Zbijewski; Freek J Beekman
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.