Literature DB >> 22010112

Semi-real time electrochemical monitoring for influenza virus RNA by reverse transcription loop-mediated isothermal amplification using a USB powered portable potentiostat.

Naoki Nagatani1, Keiichiro Yamanaka, Masato Saito, Ritsuko Koketsu, Tadahiro Sasaki, Kazuyoshi Ikuta, Toshiro Miyahara, Eiichi Tamiya.   

Abstract

In this paper, the semi-real time electrochemical monitoring method using a screen-printed electrode, which employs reverse transcription loop-mediated isothermal amplification (RT-LAMP) for influenza virus RNA, is presented. The amplified DNA combined with methylene blue (MB), which was used as an electroactive DNA intercalator, and the electrochemical signal was monitored using square wave voltammetry in the presence of RT-LAMP reagent components. MB molecules binding to amplified DNA caused the reduction of the peak current due to the slow diffusion of MB-amplified DNA complex to the electrode surface. We successfully monitored the amplification process of DNA on the basis of RT-LAMP by measuring and analyzing the electrochemical signal of MB with only one screen-printed electrode that connected with a USB powered portable potentiostat. The peak height of the current was related to the extent of amplification of DNA and the amount of input RNA. Since laborious probe immobilization is not required and both the amplification and the monitoring are possible in a single tube, our method does not suffer from potential cross-contamination. Furthermore, our method provides a new rote for the development of electrochemical hand held biosensors.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22010112     DOI: 10.1039/c1an15638a

Source DB:  PubMed          Journal:  Analyst        ISSN: 0003-2654            Impact factor:   4.616


  18 in total

1.  Emerging Loop-Mediated Isothermal Amplification-Based Microchip and Microdevice Technologies for Nucleic Acid Detection.

Authors:  Mohammadali Safavieh; Manoj K Kanakasabapathy; Farhang Tarlan; Minhaz U Ahmed; Mohammed Zourob; Waseem Asghar; Hadi Shafiee
Journal:  ACS Biomater Sci Eng       Date:  2016-01-21

Review 2.  Miniaturized devices for point of care molecular detection of HIV.

Authors:  Michael Mauk; Jinzhao Song; Haim H Bau; Robert Gross; Frederic D Bushman; Ronald G Collman; Changchun Liu
Journal:  Lab Chip       Date:  2017-01-31       Impact factor: 6.799

3.  Rapid, sensitive, and quantitative detection of pathogenic DNA at the point of care through microfluidic electrochemical quantitative loop-mediated isothermal amplification.

Authors:  Kuangwen Hsieh; Adriana S Patterson; B Scott Ferguson; Kevin W Plaxco; H Tom Soh
Journal:  Angew Chem Int Ed Engl       Date:  2012-04-04       Impact factor: 15.336

4.  Molecular Characterization of Increased Amplicon Lengths in SARS-CoV-2 Reverse Transcription Loop-Mediated Isothermal Amplification Assays.

Authors:  Scott W Tighe; Andrew F Hayden; Marcy L Kuentzel; Korin M Eckstrom; Jonathan Foox; Daniel L Vellone; Kristiaan H Finstad; Pheobe K Laaguiby; Jessica J Hoffman; Sridar V Chittur
Journal:  J Biomol Tech       Date:  2021-09

5.  Detection of swine influenza virus in nasal specimens by reverse transcription-loop-mediated isothermal amplification (RT-LAMP).

Authors:  Abhijeet A Bakre; Les P Jones; Hailey K Bennett; Davis E Bobbitt; Ralph A Tripp
Journal:  J Virol Methods       Date:  2020-11-30       Impact factor: 2.014

Review 6.  Updated values for molecular diagnosis for highly pathogenic avian influenza virus.

Authors:  Akira Sakurai; Futoshi Shibasaki
Journal:  Viruses       Date:  2012-08-07       Impact factor: 5.048

Review 7.  Bridging the gap between development of point-of-care nucleic acid testing and patient care for sexually transmitted infections.

Authors:  Kuangwen Hsieh; Johan H Melendez; Charlotte A Gaydos; Tza-Huei Wang
Journal:  Lab Chip       Date:  2022-02-01       Impact factor: 7.517

Review 8.  Future developments in biosensors for field-ready Zika virus diagnostics.

Authors:  Ariana M Nicolini; Katherine E McCracken; Jeong-Yeol Yoon
Journal:  J Biol Eng       Date:  2017-01-23       Impact factor: 4.355

Review 9.  Simple Approaches to Minimally-Instrumented, Microfluidic-Based Point-of-Care Nucleic Acid Amplification Tests.

Authors:  Michael G Mauk; Jinzhao Song; Changchun Liu; Haim H Bau
Journal:  Biosensors (Basel)       Date:  2018-02-26

Review 10.  Diagnostic devices for isothermal nucleic acid amplification.

Authors:  Chia-Chen Chang; Chien-Cheng Chen; Shih-Chung Wei; Hui-Hsin Lu; Yang-Hung Liang; Chii-Wann Lin
Journal:  Sensors (Basel)       Date:  2012-06-14       Impact factor: 3.576

View more