Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study.

Kenichi Nakajima1, Koichi Okuda, Shinro Matsuo, Mitsuhiro Yoshita, Junichi Taki, Masahito Yamada, Seigo Kinuya.   

Abstract

PURPOSE: This study was performed to demonstrate that the results obtained with a calibration phantom could be used as a tool for standardizing measurement of heart to mediastinum (H/M) ratio in cardiac metaiodobenzylguanidine (MIBG) imaging.
METHODS: Images of the phantom containing (123)I-MIBG were acquired on the cameras in 10 hospitals (11 camera types) to determine the relationship between H/M ratios using different collimators: low-energy (LE) and medium-energy (ME)/low-medium-energy (LME) collimators. The effect of standardization on the ME-comparable H/M ratio was examined in two settings: a Japanese standard MIBG database (n = 62) and multicentre studies (n = 49). In a multicentre study, probable Alzheimer's disease (AD, n = 18) and probable dementia with Lewy bodies (DLB, n = 31) were studied and standardized by the calibration phantom method.
RESULTS: Linear regression equations between LE and ME collimators were obtained for the phantom study in all institutions. When the H/M ratio with an LE collimator was corrected based upon the calibration phantom, the corrected values were comparable to those obtained using ME collimators. The standard database also exhibited a normal distribution after standardization as determined by skewness and goodness-of-fit test. A mixture of the populations by LE and ME collimators showed significant separation of AD and DLB groups (F ratio = 24.9 for the late H/M), but the corrected values resulted in higher F ratios for both early and late H/M (F ratio = 34.9 for the late H/M).
CONCLUSION: Standardization of H/M ratios by the heart-chest calibration phantom method is feasible among different collimator types. This method could be practically used for multicentre comparison of H/M ratios.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22009380     DOI: 10.1007/s00259-011-1963-1

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  24 in total

Review 1.  Diagnostic performance of myocardial innervation imaging using MIBG scintigraphy in differential diagnosis between dementia with lewy bodies and other dementias: a systematic review and a meta-analysis.

Authors:  Giorgio Treglia; Ernesto Cason
Journal:  J Neuroimaging       Date:  2010-11-17       Impact factor: 2.486

2.  Value of 123I-MIBG radioactivity in the differential diagnosis of DLB from AD.

Authors:  M Yoshita; J Taki; K Yokoyama; M Noguchi-Shinohara; Y Matsumoto; K Nakajima; M Yamada
Journal:  Neurology       Date:  2006-06-27       Impact factor: 9.910

3.  Extent of cardiac sympathetic neuronal damage is determined by the area of ischemia in patients with acute coronary syndromes.

Authors:  I Matsunari; U Schricke; F M Bengel; H U Haase; P Barthel; G Schmidt; S G Nekolla; A Schoemig; M Schwaiger
Journal:  Circulation       Date:  2000-06-06       Impact factor: 29.690

4.  Standardization of the heart-to-mediastinum ratio of 123I-labelled-metaiodobenzylguanidine uptake using the dual energy window method: feasibility of correction with different camera-collimator combinations.

Authors:  Shinro Matsuo; Kenichi Nakajima; Koichi Okuda; Masaya Kawano; Takehiro Ishikawa; Tetsuo Hosoya; Junichi Taki; Seigo Kinuya
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-11-07       Impact factor: 9.236

5.  Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake.

Authors:  Yusuke Inoue; Akira Suzuki; Ichiro Shirouzu; Toru Machida; Yasunori Yoshizawa; Fumihide Akita; Satoshi Ohnishi; Kohki Yoshikawa; Kuni Ohtomo
Journal:  J Nucl Cardiol       Date:  2003 Nov-Dec       Impact factor: 5.952

6.  Correction of iodine-123-labeled meta-iodobenzylguanidine uptake with multi-window methods for standardization of the heart-to-mediastinum ratio.

Authors:  Kenichi Nakajima; Kosuke Matsubara; Takehiro Ishikawa; Nobutoku Motomura; Ryo Maeda; Nasima Akhter; Koichi Okuda; Junichi Taki; Seigo Kinuya
Journal:  J Nucl Cardiol       Date:  2007 Nov-Dec       Impact factor: 5.952

7.  123I-mIBG scintigraphy to predict risk for adverse cardiac outcomes in heart failure patients: design of two prospective multicenter international trials.

Authors:  Arnold F Jacobson; John Lombard; Gopa Banerjee; Paolo G Camici
Journal:  J Nucl Cardiol       Date:  2009-01-20       Impact factor: 5.952

8.  Characterization of Japanese standards for myocardial sympathetic and metabolic imaging in comparison with perfusion imaging.

Authors:  Shinro Matsuo; Kenichi Nakajima; Shohei Yamashina; Kazuyuki Sakata; Mitsuru Momose; Jun Hashimoto; Shinichiro Kumita; Masaya Kawano; Koichi Okuda
Journal:  Ann Nucl Med       Date:  2009-06-12       Impact factor: 2.668

9.  Consensus guidelines for the clinical and pathologic diagnosis of dementia with Lewy bodies (DLB): report of the Consortium on DLB International Workshop.

Authors:  Ian G McKeith
Journal:  J Alzheimers Dis       Date:  2006       Impact factor: 4.472

10.  Variations in 123I-metaiodobenzylguanidine (MIBG) late heart mediastinal ratios in chronic heart failure: a need for standardisation and validation.

Authors:  Hein J Verberne; Jan B A Habraken; Berthe L F van Eck-Smit; Denis Agostini; Arnold F Jacobson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-06       Impact factor: 9.236

View more
  28 in total

1.  The time has come to standardize (123)I-MIBG heart-to-mediastinum ratios including planar and SPECT methods.

Authors:  Kenichi Nakajima; Koichi Okuda; Shinro Matsuo; Denis Agostini
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-11-23       Impact factor: 9.236

2.  123I-mIBG and the phantom tollbooth.

Authors:  Albert Flotats; Ignasi Carrió
Journal:  J Nucl Cardiol       Date:  2017-03-13       Impact factor: 5.952

3.  Development and validation of a direct-comparison method for cardiac (123)I-metaiodobenzylguanidine washout rates derived from late 3-hour and 4-hour imaging.

Authors:  Koichi Okuda; Kenichi Nakajima; Shuichi Sugino; Yumiko Kirihara; Shinro Matsuo; Junichi Taki; Mitsumasa Hashimoto; Seigo Kinuya
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-08-23       Impact factor: 9.236

4.  Quantification of I-123-meta-iodobenzylguanidine heart-to-mediastinum ratios: not so simple after all.

Authors:  Piotr J Slomka; Puja K Mehta; Guido Germano; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2014-07-09       Impact factor: 5.952

5.  ASNC imaging guidelines for SPECT nuclear cardiology procedures: Stress, protocols, and tracers.

Authors:  Milena J Henzlova; W Lane Duvall; Andrew J Einstein; Mark I Travin; Hein J Verberne
Journal:  J Nucl Cardiol       Date:  2016-06       Impact factor: 5.952

6.  Observer reproducibility of results from a low-dose 123I-metaiodobenzylguanidine cardiac imaging protocol in patients with heart failure.

Authors:  Teresa Pellegrino; Mario Petretta; Serena De Luca; Stefania Paolillo; Antonio Boemio; Raffaella Carotenuto; Maria Piera Petretta; Carmine di Nuzzo; Pasquale Perrone-Filardi; Alberto Cuocolo
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-06-06       Impact factor: 9.236

Review 7.  Radionuclide imaging of cardiac sympathetic innervation in heart failure: unlocking untapped potential.

Authors:  Shuchita Gupta; Aman Amanullah
Journal:  Heart Fail Rev       Date:  2015-03       Impact factor: 4.214

8.  First determination of the heart-to-mediastinum ratio using cardiac dual isotope (¹²³I-MIBG/⁹⁹mTc-tetrofosmin) CZT imaging in patients with heart failure: the ADRECARD study.

Authors:  Dimitri Bellevre; Alain Manrique; Damien Legallois; Samy Bross; Rafael Baavour; Nathaniel Roth; Tanguy Blaire; Cédric Desmonts; Alban Bailliez; Denis Agostini
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-07-31       Impact factor: 9.236

Review 9.  (123)I-Meta-iodobenzylguanidine Sympathetic Imaging: Standardization and Application to Neurological Diseases.

Authors:  Kenichi Nakajima; Masahito Yamada
Journal:  Chonnam Med J       Date:  2016-09-23

10.  A prediction model for 5-year cardiac mortality in patients with chronic heart failure using ¹²³I-metaiodobenzylguanidine imaging.

Authors:  Kenichi Nakajima; Tomoaki Nakata; Takahisa Yamada; Shohei Yamashina; Mitsuru Momose; Shu Kasama; Toshiki Matsui; Shinro Matsuo; Mark I Travin; Arnold F Jacobson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-03-25       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.