A motion constrained cross-wire phantom for tracked 2D ultrasound calibration.

Eivind Lyche Melvaer1, Knut Mørken, Eigil Samset.   

Abstract

PURPOSE: Ultrasound-guided 3D interventions require calibration to relate real-time 2D images with the position and orientation of the ultrasound probe. Capturing several images of a single fixed point from different viewpoints is a simple and commonly used approach, but it is cumbersome and tedious. A new phantom for calibration was designed, built and tested to simplify this process.
METHODS: A mechanical phantom that restricts the motion of the ultrasound probe was designed such that the ultrasound image always captures a designated fixed point. Software was implemented which computes calibration parameters. Although the software provides no scientific novelty, it is required to demonstrate the proof of concept and to assess the accuracy and precision of the calibration phantom. The software also illustrates how the phantom enables the fixed point to be located automatically, both in tracker device coordinates and in image pixel coordinates.
RESULTS: The phantom was used to capture several hundred images of a single fixed point in less than 1 min, with different probe positions and orientations around the fixed point and with the single fixed point located in different parts of the ultrasound image. It would not be feasible to capture the same number of images by manual alignment of the probe with the fixed point.
CONCLUSION: Images for single fixed point calibration can be captured easily and quickly with a new calibration phantom. Since a larger number of images can be used to compute the required parameters, the calibration robustness is increased.

Mesh:

Year:  2011        PMID: 22009307     DOI: 10.1007/s11548-011-0661-6

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  12 in total

1.  The effect of acoustic velocity on phantom measurements.

Authors:  A Goldstein
Journal:  Ultrasound Med Biol       Date:  2000-09       Impact factor: 2.998

2.  Respiratory motion compensation for 3-D freehand echocardiography.

Authors:  D Atkinson; M Burcher; J Declerck; J A Noble
Journal:  Ultrasound Med Biol       Date:  2001-12       Impact factor: 2.998

Review 3.  A review of calibration techniques for freehand 3-D ultrasound systems.

Authors:  Laurence Mercier; Thomas Langø; Frank Lindseth; D Louis Collins
Journal:  Ultrasound Med Biol       Date:  2005-04       Impact factor: 2.998

4.  A mechanical instrument for 3D ultrasound probe calibration.

Authors:  Andrew H Gee; Neil E Houghton; Graham M Treece; Richard W Prager
Journal:  Ultrasound Med Biol       Date:  2005-04       Impact factor: 2.998

Review 5.  Freehand 3D ultrasound reconstruction algorithms--a review.

Authors:  Ole Vegard Solberg; Frank Lindseth; Hans Torp; Richard E Blake; Toril A Nagelhus Hernes
Journal:  Ultrasound Med Biol       Date:  2007-05-18       Impact factor: 2.998

6.  Real-time freehand 3D ultrasound calibration.

Authors:  Po-Wei Hsu; Richard W Prager; Andrew H Gee; Graham M Treece
Journal:  Ultrasound Med Biol       Date:  2007-10-23       Impact factor: 2.998

7.  Evaluation of bivariate correlation ratio similarity metric for rigid registration of US/MR images of the liver.

Authors:  Sergiy Milko; Eivind Lyche Melvaer; Eigil Samset; Timor Kadir
Journal:  Int J Comput Assist Radiol Surg       Date:  2009-02-13       Impact factor: 2.924

8.  Rapid calibration for 3-D freehand ultrasound.

Authors:  R W Prager; R N Rohling; A H Gee; L Berman
Journal:  Ultrasound Med Biol       Date:  1998-07       Impact factor: 2.998

9.  A computational approach to edge detection.

Authors:  J Canny
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1986-06       Impact factor: 6.226

10.  3D ultrasonic image feature localization based on magnetic scanhead tracking: in vitro calibration and validation.

Authors:  P R Detmer; G Bashein; T Hodges; K W Beach; E P Filer; D H Burns; D E Strandness
Journal:  Ultrasound Med Biol       Date:  1994       Impact factor: 2.998

View more
  4 in total

1.  Robot-assisted automatic ultrasound calibration.

Authors:  Fereshteh Aalamifar; Alexis Cheng; Younsu Kim; Xiao Hu; Haichong K Zhang; Xiaoyu Guo; Emad M Boctor
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-01-11       Impact factor: 2.924

2.  Phantom with multiple active points for ultrasound calibration.

Authors:  Haichong K Zhang; Alexis Cheng; Younsu Kim; Qianli Ma; Gregory S Chirikjian; Emad M Boctor
Journal:  J Med Imaging (Bellingham)       Date:  2018-11-27

3.  User-friendly freehand ultrasound calibration using Lego bricks and automatic registration.

Authors:  Yiming Xiao; Charles Xiao Bo Yan; Simon Drouin; Dante De Nigris; Anna Kochanowska; D Louis Collins
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-03-16       Impact factor: 2.924

4.  Active phantoms: a paradigm for ultrasound calibration using phantom feedback.

Authors:  Alexis Cheng; Xiaoyu Guo; Haichong K Zhang; Hyun Jae Kang; Ralph Etienne-Cummings; Emad M Boctor
Journal:  J Med Imaging (Bellingham)       Date:  2017-07-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.