PURPOSE: To compare intraocular pressure (IOP) measurements after penetrating keratoplasty (PK) using Goldmann applanation tonometry (GAT; Haag-Streit USA), TonoPen XL (Reichert Inc), Pascal Dynamic Contour tonometer (PDCT; Swiss Microtechnology AG), and Ocular Response Analyzer (ORA; Reichert Inc) and to analyze effects and correlation of corneal thickness and curvature on these measurements.
DESIGN: Prospective, cross-sectional study. SETTINGS: Departments of Ophthalmology, University of Auckland and Auckland District Health Board, New Zealand. STUDY POPULATION: Thirty-one eyes of 31 participants with previous PK. OBSERVATIONS: IOP measured using GAT, TonoPen, PDCT, and ORA. Central corneal thickness (CCT) and corneal astigmatism were assessed by Pentacam rotating Scheimpflug tomography. MAIN OUTCOME MEASURE: Degree of agreement in IOP measures and correlation with CCT and corneal astigmatism.
RESULTS: Mean age was 44.5 ± 21.0 years and mean time since PK was 27.7 ± 28.2 months (range 3.0-122.4 months). Mean CCT was 585 ± 149 μm and mean corneal astigmatism 5.5 ± 3.8 diopters. There was no significant difference in IOP measured by GAT and TonoPen; however, both PDCT (2.12 mm Hg, P < .01) and ORA (6.29 mm Hg, P < .01) measured significantly higher IOPs compared to GAT. ORA showed the least agreement with GAT. Significant correlation was identified between each pair of measurement techniques. No instruments correlated significantly with CCT. Only PDCT showed no significant correlation with corneal astigmatism. However, no IOP measures correlated with corneal astigmatism if sutures in situ or less than 1 year post-PK.
CONCLUSIONS: TonoPen or PDCT are the most suitable alternatives for measuring IOP in PK eyes where GAT readings are difficult to obtain. PDCT also offers the advantage of being independent of both CCT and corneal astigmatism. Copyright Â
© 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22000702     DOI: 10.1016/j.ajo.2011.08.020

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  5 in total

1.  Ultrasound biomicroscopic evaluation of anterior segment cysts as a risk factor for ocular hypertension and closure angle glaucoma.

Authors:  Abdurrahim Dusak; Mehmet Baykara; Guven Ozkaya; Cuneyt Erdogan; Hikmet Ozcetin; Ercan Tuncel
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  Changes in intraocular pressure values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology tonometer (CST) in the early phase after small incision lenticule extraction (SMILE).

Authors:  Yang Shen; Xiangjian Su; Xiu Liu; Huamao Miao; Xuejun Fang; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2016-11-18       Impact factor: 2.209

3.  Corneal biomechanics and intraocular pressure assessment after penetrating keratoplasty for non keratoconic patients, long term results.

Authors:  Mohamed Samy Abd Elaziz; Hoda Mohamed Elsobky; Adel Galal Zaky; Eslam Ahmed Maher Hassan; Mahmoud Tawfik KhalafAllah
Journal:  BMC Ophthalmol       Date:  2019-08-07       Impact factor: 2.209

Review 4.  Intraocular Pressure Measurement after Penetrating Keratoplasty.

Authors:  Otilia-Maria Dumitrescu; Sinziana Istrate; Mioara-Laura Macovei; Alina Gabriela Gheorghe
Journal:  Diagnostics (Basel)       Date:  2022-01-19

5.  De novo ocular hypertension after Descemet stripping endothelial keratoplasty: comparative 3-year incidence, risk factors, and outcomes.

Authors:  Errol W Chan; Tina T Wong; Hla M Htoon; Ching L Ho; Donald T Tan; Jodhbir S Mehta
Journal:  Clin Ophthalmol       Date:  2013-09-17
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.