Influence of temperature changes on ambient air NOx chemiluminescence measurements.

Marta Doval Miñarro1, Enrique González Ferradás, Francisco J Marzal Martínez.   

Abstract

Users of automatic air pollution monitors are largely unaware of how certain parameters, like temperature, can affect readings. The present work examines the influence of temperature changes on chemiluminescence NO(x) measurements made with a Thermo Scientific 42i analyzer, a model widely used in air monitoring networks and air pollution studies. These changes are grouped into two categories according to European Standard EN 14211: (1) changes in the air surrounding the analyzers and (2) changes in the sampled air. First, the sensitivity tests described in Standard EN 14211 were performed to determine whether the analyzer performance was adapted to the requirements of the standard. The analyzer met the performance criteria of both tests; however, some differences were detected in readings with temperature changes even though the temperature compensator was on. Sample temperature changes were studied more deeply as they were the most critical (they cannot be controlled and differences of several tens of degrees can be present in a single day). Significant differences in readings were obtained when changing sample temperature; however, maximum deviations were around 3% for temperature ranges of 15°C. If other possible uncertainty contributions are controlled and temperature variations with respect to the calibration temperature are not higher than 15°C, the effect of temperature changes could be acceptable and no data correction should have to be applied.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21964932     DOI: 10.1007/s10661-011-2372-4

Source DB:  PubMed          Journal:  Environ Monit Assess        ISSN: 0167-6369            Impact factor:   2.513


  6 in total

Review 1.  Chemiluminescence in gas analysis and Flame-emission spectrometry.

Authors:  J H Glover
Journal:  Analyst       Date:  1975-07       Impact factor: 4.616

2.  Study of the uncertainty in NO2 chemiluminescence measurements due to the NO-O3 reaction in sampling lines.

Authors:  Marta Doval Miñarro; Enrique González Ferradás; Jonathan Barberá Rico; Fernando Díaz Alonso; Francisco J Marzal Martínez; Cristina Romero Trigueros
Journal:  Environ Sci Pollut Res Int       Date:  2010-09-04       Impact factor: 4.223

3.  A re-examination of ambient air ozone monitor interferences.

Authors:  Chester W Spicer; Darrell W Joseph; Will M Ollison
Journal:  J Air Waste Manag Assoc       Date:  2010-11       Impact factor: 2.235

4.  Potential interference bias in ozone standard compliance monitoring.

Authors:  Alan R Leston; Will M Ollison; Chester W Spicer; Jan Satola
Journal:  J Air Waste Manag Assoc       Date:  2005-10       Impact factor: 2.235

5.  Mechanism and elimination of a water vapor interference in the measurement of ozone by UV absorbance.

Authors:  Kevin L Wilson; John W Birks
Journal:  Environ Sci Technol       Date:  2006-10-15       Impact factor: 9.028

6.  Comparison of ultraviolet absorbance, chemiluminescence, and DOAS instruments for ambient ozone monitoring.

Authors:  E J Williams; F C Fehsenfeld; B T Jobson; W C Kuster; P D Goldan; J Stutz; W A McClenny
Journal:  Environ Sci Technol       Date:  2006-09-15       Impact factor: 9.028