An empirical model of diagnostic x-ray attenuation under narrow-beam geometry.

Kelsey B Mathieu1, S Cheenu Kappadath, R Allen White, E Neely Atkinson, Dianna D Cody.   

Abstract

PURPOSE: The purpose of this study was to develop and validate a mathematical model to describe narrow-beam attenuation of kilovoltage x-ray beams for the intended applications of half-value layer (HVL) and quarter-value layer (QVL) estimations, patient organ shielding, and computer modeling.
METHODS: An empirical model, which uses the Lambert W function and represents a generalized Lambert-Beer law, was developed. To validate this model, transmission of diagnostic energy x-ray beams was measured over a wide range of attenuator thicknesses [0.49-33.03 mm Al on a computed tomography (CT) scanner, 0.09-1.93 mm Al on two mammography systems, and 0.1-0.45 mm Cu and 0.49-14.87 mm Al using general radiography]. Exposure measurements were acquired under narrow-beam geometry using standard methods, including the appropriate ionization chamber, for each radiographic system. Nonlinear regression was used to find the best-fit curve of the proposed Lambert W model to each measured transmission versus attenuator thickness data set. In addition to validating the Lambert W model, we also assessed the performance of two-point Lambert W interpolation compared to traditional methods for estimating the HVL and QVL [i.e., semi-logarithmic (exponential) and linear interpolation].
RESULTS: The Lambert W model was validated for modeling attenuation versus attenuator thickness with respect to the data collected in this study (R2 > 0.99). Furthermore, Lambert W interpolation was more accurate and less sensitive to the choice of interpolation points used to estimate the HVL and/or QVL than the traditional methods of semilogarithmic and linear interpolation.
CONCLUSIONS: The proposed Lambert W model accurately describes attenuation of both monoenergetic radiation and (kilovoltage) polyenergetic beams (under narrow-beam geometry).

Entities:  

Mesh:

Year:  2011        PMID: 21928626      PMCID: PMC3155576          DOI: 10.1118/1.3592933

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Analytical considerations of beam hardening in medical accelerator photon spectra.

Authors:  C Kleinschmidt
Journal:  Med Phys       Date:  1999-09       Impact factor: 4.071

2.  Measurement of half-value layer in x-ray CT: a comparison of two noninvasive techniques.

Authors:  R L Kruger; C H McCollough; F E Zink
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

3.  Precision of dosimetry-related measurements obtained on current multidetector computed tomography scanners.

Authors:  Kelsey B Mathieu; Michael F McNitt-Gray; Di Zhang; Hyun J Kim; Dianna D Cody
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

4.  On the measurement of half-value layer in film-screen mammography.

Authors:  L K Wagner; B R Archer; F Cerra
Journal:  Med Phys       Date:  1990 Nov-Dec       Impact factor: 4.071

5.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations.

Authors:  Adam C Turner; Di Zhang; Hyun J Kim; John J DeMarco; Chris H Cagnon; Erin Angel; Dianna D Cody; Donna M Stevens; Andrew N Primak; Cynthia H McCollough; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

6.  The AAPM/RSNA physics tutorial for residents. X-ray attenuation.

Authors:  M H McKetty
Journal:  Radiographics       Date:  1998 Jan-Feb       Impact factor: 5.333

7.  Linear accelerator photon beam quality at off-axis points.

Authors:  M K Yu; R S Sloboda; B Murray
Journal:  Med Phys       Date:  1997-02       Impact factor: 4.071

8.  Attenuation in high-energy x-ray beams.

Authors:  B E Bjärngard; H Shackford
Journal:  Med Phys       Date:  1994-07       Impact factor: 4.071

9.  Diagnostic X-ray shielding design based on an empirical model of photon attenuation.

Authors:  B R Archer; J I Thornby; S C Bushong
Journal:  Health Phys       Date:  1983-05       Impact factor: 1.316

  9 in total
  2 in total

1.  The potential impact of ultrathin filter design on dosimetry and relative biological effectiveness in modern image-guided small animal irradiators.

Authors:  Yannick Poirier; Christopher Daniel Johnstone; Charles Kirkby
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

2.  Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case in X-ray computed tomography.

Authors:  Kosuke Matsubara; Katsuhiro Ichikawa; Yuichi Murasaki; Ayaka Hirosawa; Kichiro Koshida
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.