Central corneal thickness measurements: using an ultrasonic instrument and 4 optical instruments.

Ravaughn Williams1, Barbara A Fink, P Ewen King-Smith, G Lynn Mitchell.   

Abstract

PURPOSE: Both between-session and within-session repeatability were determined for measurement of corneal thickness with the following techniques: ultrasound pachymetry (UP), confocal microscopy (CS), Orbscan pachymetry (ORB), spectral oscillation interferometry (SOI), and optical coherence tomography (OCT).
METHODS: The right eyes of 20 normal subjects were tested on 2 different days. For each session, 2 central corneal thickness measurements were determined using the ConfoScan 3 microscope, the Orbscan system, and the Stratus OCT system; 5 measurements were found with the Sonogage ultrasound pachymeter; and between 2 and 10 measurements were found with the SOI system.
RESULTS: This study showed that SOI had the best repeatability, whereas CS had the worst repeatability for both within-session and between-session repeatability. Based on 95% limits of agreement (LoA), the within-session repeatability can be ranked as: (1) SOI; (2) UP; (3) OCT; (4) Orbscan pachymetry; and (5) CS. SOI had a bias of 0.13 and 95% LoA of -1.07 to 1.33, whereas UP had a bias of -0.98 and 95% LoA of -7.35 to 5.40. The between-session repeatability of the instruments can be ranked as: (1) SOI; (2) OCT; (3) UP; (4) Orbscan pachymetry; and (5) CS. Of the optical methods, SOI had the best repeatability with a bias of 1.26 and 95% LoA of -6.14 to 8.66, followed by OCT with a bias of 0.75 and 95% LoA of -16.22 to 17.27.
CONCLUSIONS: Of the 4 instruments that are commercially available, the UP was the most repeatable for within sessions, and the OCT was the most repeatable for between sessions.

Entities:  

Mesh:

Year:  2011        PMID: 21926567     DOI: 10.1097/ICO.0b013e3182152051

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  9 in total

1.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

2.  Decreased central corneal thickness in ankylosing spondylitis.

Authors:  Huseyin Ortak; Ahmet Inanır; Selim Demir; Alper Uysal; Şafak Şahin; Mustafa Sağcan; Yalçın Önder; Sait Alim; Ayşe Kevser Demir
Journal:  Int Ophthalmol       Date:  2013-07-12       Impact factor: 2.031

3.  Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan.

Authors:  Wassia A Khaja; Sandeep Grover; Amy T Kelmenson; Lee R Ferguson; Kumar Sambhav; Kakarla V Chalam
Journal:  Clin Ophthalmol       Date:  2015-06-12

4.  Accuracy of optical coherence tomography (OCT) in pachymetry for glaucoma patients.

Authors:  Marcelo Ayala; Robert Strandås
Journal:  BMC Ophthalmol       Date:  2015-09-29       Impact factor: 2.209

5.  Validity of Tono-pachymetry for Measuring Corrected Intraocular Pressure in Non-surgical and Post-photorefractive Keratectomy Eyes.

Authors:  In Kyun Hahn; Jae Yong Kim; Myoung Joon Kim; Hungwon Tchah; Chan Hee Moon
Journal:  Korean J Ophthalmol       Date:  2017-02-02

6.  Estimating the Osmolarities of Tears During Evaporation Through the "Eyes" of the Corneal Nerves.

Authors:  Harumitsu Hirata; Kamila Mizerska; Valentina Dallacasagrande; Mark I Rosenblatt
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-01-01       Impact factor: 4.799

7.  Comparison of Central Corneal Thickness Measurements by Contact and Non-contact Pachymetry Devices.

Authors:  Huseyin Mayali; Muhammed Altinisik; Ismail Diri; Sami Ilker; Emin Kurt; Ozcan Kayikcioglu
Journal:  J Curr Glaucoma Pract       Date:  2021 Jan-Apr

8.  Central Corneal Thickness Measurements with Different Imaging Devices: Ultrasound Pachymetry, Noncontact Specular Microscopy, and Tono-Pachymetry.

Authors:  Ali Ceylan; Irem Onal; Burak Mergen; Yusuf Yildirim
Journal:  Beyoglu Eye J       Date:  2022-02-18

9.  Effect of Proparacaine 0.375%-Sodium Fluorescein 0.25% Eye Drop Mixture and Fluorescein Strip on Anterior Segment Parameters.

Authors:  Mustafa Doğan; Mehmet Cem Sabaner; Mehmet Akif Erol
Journal:  J Ophthalmol       Date:  2018-09-05       Impact factor: 1.909

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.