Insulin pump therapy with automated insulin suspension in response to hypoglycemia: reduction in nocturnal hypoglycemia in those at greatest risk.

Pratik Choudhary1, John Shin, Yongyin Wang, Mark L Evans, Peter J Hammond, David Kerr, James A M Shaw, John C Pickup, Stephanie A Amiel.   

Abstract

OBJECTIVE: To evaluate a sensor-augmented insulin pump with a low glucose suspend (LGS) feature that automatically suspends basal insulin delivery for up to 2 h in response to sensor-detected hypoglycemia. RESEARCH DESIGN AND METHODS: The LGS feature of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA) was tested for 3 weeks in 31 adults with type 1 diabetes.
RESULTS: There were 166 episodes of LGS: 66% of daytime LGS episodes were terminated within 10 min, and 20 episodes lasted the maximum 2 h. LGS use was associated with reduced nocturnal duration ≤2.2 mmol/L in those in the highest quartile of nocturnal hypoglycemia at baseline (median 46.2 vs. 1.8 min/day, P = 0.02 [LGS-OFF vs. LGS-ON]). Median sensor glucose was 3.9 mmol/L after 2-h LGS and 8.2 mmol/L at 2 h after basal restart.
CONCLUSIONS: Use of an insulin pump with LGS was associated with reduced nocturnal hypoglycemia in those at greatest risk and was well accepted by patients.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21868778      PMCID: PMC3161284          DOI: 10.2337/dc10-2411

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


Continuous glucose monitoring (CGM) can reduce HbA1c in type 1 diabetes (1–3). Despite the use of hypoglycemia alarms, most studies have not demonstrated a significant reduction in hypoglycemia, and prolonged nocturnal hypoglycemia occurs frequently (4). This may be because patients sleep through many of the alarms (5) and insulin delivery continues during hypoglycemia. We report a user evaluation of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA), which can automatically suspend basal insulin delivery for up to 2 h in the event of CGM-detected hypoglycemia, thus reducing the duration of hypoglycemia.

RESEARCH DESIGN AND METHODS

The Veo was evaluated by 31 patients (10 men) with type 1 diabetes (mean age, 41.9 ± 10.6 years) from six U.K. centers. Regional ethics committees approved the study, and patients provided informed consent. The Veo system has alarms for hypoglycemia, predicted hypoglycemia, rate-of-change of glucose, and, uniquely, a low glucose suspend (LGS) feature that is activated when sensor glucose reaches a glucose threshold set by the user. An alarm sounds, and if the user does not respond, basal insulin delivery is suspended for a maximum of 2 h, after which basal insulin delivery is resumed at the programmed rate. The patient may resume basal insulin delivery at any point. During a 2-week run-in, CGM was used with only predictive, rate-of-change, high and low alerts active (LGS-OFF). LGS was then activated for 3 weeks (LGS-ON). We evaluated the response to LGS and compared hypoglycemia exposure and mean blood glucose during LGS-OFF and LGS-ON. Patients were divided into four equal groups (quartiles) by the duration of hypoglycemia during the run-in period, because we wished to test the hypothesis that those with the most hypoglycemia at baseline (without LGS) would have the greatest benefit with LGS. Hypoglycemia was defined as the lower limit of detection of the sensor (2.2 mmol/L) (6). The glucose threshold to trigger LGS was individualized (median 2.4 [range 2.2–3.5] mmol/L). Night was defined as 0000–0800 h. Treatment satisfaction questionnaires were completed at study end. Data were compared using the Student t test, except for skewed data (hypoglycemia duration), which were compared with the Wilcoxon test. Values are mean ± SD or median (range).

RESULTS

Two subjects withdrew during run-in due to difficulties using sensors, and one subject failed to activate the LGS. There were 166 LGS episodes in 25 of 28 (89%) completers (mean 1.9 LGS events/week), of which 76% occurred during daytime, and 55% were terminated within 10 min. LGS continued for the maximum 2 h in 20 episodes (12%), 75% of which were nocturnal. Of 20 completed 2-h suspends, 7 (35%) had no patient response throughout. In the remaining 13 (65%), patients responded to the alarm but elected to continue LGS for 2 h. Mean response time to the LGS alarm was longer at night compared with day (63.2 ± 8.2 vs. 17.4 ± 2.7 min, P < 0.001). LGS use was associated with significant reduction in the duration of nocturnal hypoglycemia (≤2.2 mmol/L) in those in the highest quartile of hypoglycemia duration at baseline: median 46.2 (36.6–191.4) vs. 1.8 (0.0–45) min/day (P = 0.02; LGS-OFF vs. LGS-ON) (Fig. 1) and mean 75.1 ± 54 vs. 10.2 ± 18 min/day (P = 0.02). Mean sensor glucose was not different with LGS-OFF or LGS-ON (6.4 ± 1.3 vs. 6.6 ± 1.1 mmol/L, P = 0.26). After the 20 complete 2-h LGS episodes, median sensor glucose was 3.9 (2.4–14.2) mmol/L at the restart of basal insulin and was 8.2 (3.3–17.3) mmol/L 2 h after restart. Carbohydrate ingestion was not recorded.
Figure 1

Duration of nocturnal hypoglycemia (sensor glucose < 2.2 mmol/L) with and without LGS. The bars show median duration of hypoglycemia at night with LGS-OFF (black bars) and LGS-ON (gray bars) by quartile (q) of nocturnal hypoglycemia exposure at baseline.

Duration of nocturnal hypoglycemia (sensor glucose < 2.2 mmol/L) with and without LGS. The bars show median duration of hypoglycemia at night with LGS-OFF (black bars) and LGS-ON (gray bars) by quartile (q) of nocturnal hypoglycemia exposure at baseline. Concomitant (within 15 min of LGS) capillary glucose values were available for 43 of 166 episodes (25.9%) of LGS. These were >5 mmol/L in 13 episodes and >10 mmol/L in 4, although we do not know if any carbohydrate was ingested before testing. LGS was terminated within 2 min in all four episodes with capillary glucose >10 mmol/L, with sensor error alerts in two of these. All subjects reported finding LGS “useful,” and 93% reported feeling more secure at night, with reduced anxiety, and wanted to continue using it.

CONCLUSIONS

These data suggest that LGS has the potential to reduce nocturnal hypoglycemia in patients with type 1 diabetes at the highest risk. This is similar to results with insulin pump therapy providing the greatest reduction in hypoglycemia in those with the most hypoglycemia at baseline (7). The risk of ketosis and hyperglycemia after the 2-h suspension of insulin delivery is low (8,9). In our study, median sensor glucose after 2 h of LGS was 3.9 (2.4–14.2) mmol/L. We could not determine if carbohydrate had been consumed during the LGS. There was no evidence of deterioration of overall glucose control with LGS activated. Patients took longer to respond to the LGS alarm at night, and 75% of completed 2-h LGS events occurred overnight. This may relate to the combined effects of sleep and hypoglycemia on alertness/arousability, and reduced counter-regulatory responses during sleep (10). Most daytime LGS episodes were terminated by users within 10 min. The sensor may under-report glucose, particularly during nocturnal hypoglycemia and in view of the lag between interstitial and capillary glucose (11). Although the lowest displayed sensor value is 2.2 mmol/L, the Veo algorithm has improved hypoglycemia detection compared with previous algorithms, with a mean absolute relative difference between 2.2 and 4.4 mmol/L reduced from 24.8 to19.5% (12). We set the LGS threshold low (mean, 2.4 mmol/L), and a higher threshold may have led to greater reduction in hypoglycemia. Our study could not determine rates of false-positive LGS: 4 of 43 LGS had a capillary glucose reading within 15 min >10 mmol/L, and 2 of these were preceded by sensor error alerts. However, these capillary readings may be biased toward episodes when the patient thought the LGS was erroneous. LGS reduces anxiety about nocturnal hypoglycemia. Randomized controlled trials that evaluate hypoglycemia and quality-of-life in type 1 diabetes using LGS pumps compared with insulin pump therapy alone, with or without CGM, are now needed. This is the first system that modulates insulin delivery in response to glucose levels without human intervention and is an important step toward clinically available closed-loop systems.
  12 in total

1.  Do sensor glucose levels accurately predict plasma glucose concentrations during hypoglycemia and hyperinsulinemia?

Authors:  Teresa P Monsod; Daniel E Flanagan; Fran Rife; Rebecca Saenz; Sonia Caprio; Robert S Sherwin; William V Tamborlane
Journal:  Diabetes Care       Date:  2002-05       Impact factor: 19.112

2.  Response to nocturnal alarms using a real-time glucose sensor.

Authors:  Bruce Buckingham; Jen Block; Jonathan Burdick; Andrea Kalajian; Craig Kollman; Michael Choy; Darrell M Wilson; Peter Chase
Journal:  Diabetes Technol Ther       Date:  2005-06       Impact factor: 6.118

3.  Comparison of metabolic deterioration between insulin analog and regular insulin after a 5-hour interruption of a continuous subcutaneous insulin infusion in type 1 diabetic patients.

Authors:  B Guerci; L Meyer; A Sallé; A Charrié; B Dousset; O Ziegler; P Drouin
Journal:  J Clin Endocrinol Metab       Date:  1999-08       Impact factor: 5.958

4.  Safety of continuous subcutaneous insulin infusion: metabolic deterioration and glycaemic autoregulation after deliberate cessation of infusion.

Authors:  J C Pickup; G C Viberti; R W Bilous; H Keen; K G Alberti; P D Home; C Binder
Journal:  Diabetologia       Date:  1982-03       Impact factor: 10.122

5.  Glycaemic impact of patient-led use of sensor-guided pump therapy in type 1 diabetes: a randomised controlled trial.

Authors:  M A O'Connell; S Donath; D N O'Neal; P G Colman; G R Ambler; T W Jones; E A Davis; F J Cameron
Journal:  Diabetologia       Date:  2009-04-25       Impact factor: 10.122

6.  Decreased epinephrine responses to hypoglycemia during sleep.

Authors:  T W Jones; P Porter; R S Sherwin; E A Davis; P O'Leary; F Frazer; G Byrne; S Stick; W V Tamborlane
Journal:  N Engl J Med       Date:  1998-06-04       Impact factor: 91.245

7.  Continuous glucose monitoring and intensive treatment of type 1 diabetes.

Authors:  William V Tamborlane; Roy W Beck; Bruce W Bode; Bruce Buckingham; H Peter Chase; Robert Clemons; Rosanna Fiallo-Scharer; Larry A Fox; Lisa K Gilliam; Irl B Hirsch; Elbert S Huang; Craig Kollman; Aaron J Kowalski; Lori Laffel; Jean M Lawrence; Joyce Lee; Nelly Mauras; Michael O'Grady; Katrina J Ruedy; Michael Tansey; Eva Tsalikian; Stuart Weinzimer; Darrell M Wilson; Howard Wolpert; Tim Wysocki; Dongyuan Xing
Journal:  N Engl J Med       Date:  2008-09-08       Impact factor: 91.245

Review 8.  Severe hypoglycaemia and glycaemic control in Type 1 diabetes: meta-analysis of multiple daily insulin injections compared with continuous subcutaneous insulin infusion.

Authors:  J C Pickup; A J Sutton
Journal:  Diabet Med       Date:  2008-07       Impact factor: 4.359

9.  Risk of hypoglycaemia in types 1 and 2 diabetes: effects of treatment modalities and their duration.

Authors: 
Journal:  Diabetologia       Date:  2007-04-06       Impact factor: 10.122

10.  Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study.

Authors:  Denis Raccah; Véronique Sulmont; Yves Reznik; Bruno Guerci; Eric Renard; Hélène Hanaire; Nathalie Jeandidier; Marc Nicolino
Journal:  Diabetes Care       Date:  2009-09-18       Impact factor: 19.112

View more
  52 in total

Review 1.  Sensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes.

Authors:  Isabelle Steineck; Ajenthen Ranjan; Kirsten Nørgaard; Signe Schmidt
Journal:  J Diabetes Sci Technol       Date:  2016-10-06

2.  Nocturnal continuous glucose and sleep stage data in adults with type 1 diabetes in real-world conditions.

Authors:  Stephanie Feudjio Feupe; Patrick F Frias; Sara C Mednick; Elizabeth A McDevitt; Nathaniel D Heintzman
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

Review 3.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

Review 4.  Management of diabetes mellitus: is the pump mightier than the pen?

Authors:  John C Pickup
Journal:  Nat Rev Endocrinol       Date:  2012-02-28       Impact factor: 43.330

Review 5.  Outpatient Management of Pediatric Type 1 Diabetes.

Authors:  Joni K Beck; Fran R Cogen
Journal:  J Pediatr Pharmacol Ther       Date:  2015 Sep-Oct

6.  7(th) Annual Symposium on Self-Monitoring of Blood Glucose (SMBG), May 8-10, 2014, Helsinki, Finland.

Authors:  Christopher G Parkin; Anita Mlinac; Rolf Hinzmann
Journal:  Diabetes Technol Ther       Date:  2014-09-11       Impact factor: 6.118

Review 7.  Advances in management of type 1 diabetes mellitus.

Authors:  Ravindranath Aathira; Vandana Jain
Journal:  World J Diabetes       Date:  2014-10-15

Review 8.  Technology to optimize pediatric diabetes management and outcomes.

Authors:  Jessica T Markowitz; Kara R Harrington; Lori M B Laffel
Journal:  Curr Diab Rep       Date:  2013-12       Impact factor: 4.810

9.  Inpatient studies of a Kalman-filter-based predictive pump shutoff algorithm.

Authors:  Fraser Cameron; Darrell M Wilson; Bruce A Buckingham; Hasmik Arzumanyan; Paula Clinton; H Peter Chase; John Lum; David M Maahs; Peter M Calhoun; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

10.  Multisite Study of an Implanted Continuous Glucose Sensor Over 90 Days in Patients With Diabetes Mellitus.

Authors:  Andrew Dehennis; Mark A Mortellaro; Sorin Ioacara
Journal:  J Diabetes Sci Technol       Date:  2015-07-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.