Verification of the linac isocenter for stereotactic radiosurgery using cine-EPID imaging and arc delivery.

Pejman Rowshanfarzad1, Mahsheed Sabet, Daryl J O'Connor, Peter B Greer.   

Abstract

PURPOSE: Verification of the mechanical isocenter position is required as part of comprehensive quality assurance programs for stereotactic radiosurgery/radiotherapy (SRS/SRT) treatments. Several techniques have been proposed for this purpose but each of them has certain drawbacks. In this paper, a new efficient and more comprehensive method using cine-EPID images has been introduced for automatic verification of the isocenter with sufficient accuracy for stereotactic applications.
METHODS: Using a circular collimator fixed to the gantry head to define the field, EPID images of a Winston-Lutz phantom were acquired in cine-imaging mode during 3600 gantry rotations. A robust MATLAB code was developed to analyze the data by finding the center of the field and the center of the ball bearing shadow in each image with sub-pixel accuracy. The distance between these two centers was determined for every image. The method was evaluated by comparison to results of a mechanical pointer and also by detection of a manual shift applied to the phantom position. The repeatability and reproducibility of the method were tested and it was also applied to detect couch and collimator wobble during rotation.
RESULTS: The accuracy of the algorithm was 0.03 +/- 0.02 mm. The repeatability was less than 3 pm and the reproducibility was less than 86 microm. The time elapsed for the analysis of more than 100 cine images of Varian aS1000 and aS500 EPIDs were approximately 65 and 20 s, respectively. Processing of images taken in integrated mode took 0.1 s. The output of the analysis software is printable and shows the isocenter shifts as a function of angle in both in-plane and cross-plane directions. It gives warning messages where the shifts exceed the criteria for SRS/SRT and provides useful data for the necessary adjustments in the system including bearing system and/or room lasers.
CONCLUSIONS: The comprehensive method introduced in this study uses cine-images, is highly accurate, fast, and independent of the observer. It tests all gantry angles and is suitable for pretreatment QA of the isocenter for stereotactic treatments.

Entities:  

Mesh:

Year:  2011        PMID: 21858993     DOI: 10.1118/1.3597836

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy.

Authors:  Weiliang Du; Song Gao; Xiaochun Wang; Rajat J Kudchadker
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

3.  A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

Authors:  P Rowshanfarzad; H L Riis; S J Zimmermann; M A Ebert
Journal:  Br J Radiol       Date:  2015-04-23       Impact factor: 3.039

4.  An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments.

Authors:  Pejman Rowshanfarzad; Conor K McGarry; Michael P Barnes; Mahsheed Sabet; Martin A Ebert
Journal:  Radiat Oncol       Date:  2014-11-26       Impact factor: 3.481

5.  Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter M McCowan; Boyd M C McCurdy; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

6.  On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

Authors:  Weiliang Du; Jennifer L Johnson; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Development and implementation of an EPID-based method for localizing isocenter.

Authors:  Daniel E Hyer; Christopher J Mart; Earl Nixon
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

8.  Improved accuracy for noncoplanar radiotherapy: an EPID-based method for submillimeter alignment of linear accelerator table rotation with MV isocenter.

Authors:  Matthew J Nyflot; Ning Cao; Juergen Meyer; Eric C Ford
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

9.  SafetyNet: Streamlining and Automating QA in radiotherapy.

Authors:  Scott W Hadley; Marc L Kessler; Dale W Litzenberg; Choonik Lee; Jim Irrer; Xiaoping Chen; Eduardo Acosta; Grant Weyburne; Wayne Keranen; Kwok Lam; Elizabeth Covington; Kelly C Younge; Martha M Matuszak; Jean M Moran
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

10.  Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device.

Authors:  Pejman Rowshanfarzad; Peter Häring; Hans L Riis; Sune J Zimmermann; Martin A Ebert
Journal:  Med Devices (Auckl)       Date:  2015-11-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.