Literature DB >> 21852567

Abdominal dynamic CT in patients with renal dysfunction: contrast agent dose reduction with low tube voltage and high tube current-time product settings at 256-detector row CT.

Takeshi Nakaura1, Kazuo Awai, Natsuki Maruyama, Noboru Takata, Ichiro Yoshinaka, Kazunori Harada, Shouzaburou Uemura, Yasuyuki Yamashita.   

Abstract

PURPOSE: To evaluate the feasibility of a low-contrast agent dose protocol at abdominal dynamic computed tomography (CT) with a low tube voltage high tube current-time product technique and a 256-detector row CT unit.
MATERIALS AND METHODS: This prospective study received institutional review board approval; written informed consent to participate was obtained from all patients. The study included 151 patients; 117 had an estimated glomerular filtration rate (eGFR) greater than or equal to 60 mL/min/1.73 m(2). These patients were examined with the conventional 120-kVp protocol. The other 34 patients underwent scanning with an 80-kVp tube voltage, a high tube current-time product, and a 40% reduction in contrast agent dose. Effective dose (ED), image noise, attenuation, contrast-to-noise ratio (CNR), and figure of merit of the aorta in the arterial phase and of the portal vein and hepatic parenchyma in the portal venous phase in the two groups were compared with the Student t test.
RESULTS: Estimated ED was about 20% lower with the 80-kVp protocol than with the 120-kVp protocol. There were no significant differences in CNR in any region of interest between the 80-kVp protocol and the 120-kVp protocol (abdominal aorta: 36.9 ± 9.7 [standard deviation] vs 36.1 ± 8.1, P = .63; portal vein: 13.4 ± 3.2 vs 13.1 ± 3.2, P = .65; hepatic parenchyma: 6.4 ± 2.6 vs 6.7 ± 2.3, P = .51).
CONCLUSION: Contrast dose at hepatic dynamic 256-detector row CT in patients with renal dysfunction can be decreased by 40% with this protocol by using the 80-kVp setting and a high tube current-time product. RSNA, 2011

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21852567     DOI: 10.1148/radiol.11110021

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  25 in total

1.  Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients.

Authors:  Fan Zhang; Li Yang; Xiang Song; Ying-Na Li; Yan Jiang; Xing-Hua Zhang; Hai-Yue Ju; Jian Wu; Rui-Ping Chang
Journal:  Br J Radiol       Date:  2015-11-26       Impact factor: 3.039

2.  Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT.

Authors:  Julian L Wichmann; Andrew D Hardie; U Joseph Schoepf; Lloyd M Felmly; Jonathan D Perry; Akos Varga-Szemes; Stefanie Mangold; Damiano Caruso; Christian Canstein; Thomas J Vogl; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2016-05-10       Impact factor: 5.315

3.  Guideline on the use of iodinated contrast media in patients with kidney disease 2018.

Authors:  Yoshitaka Isaka; Hiromitsu Hayashi; Kazutaka Aonuma; Masaru Horio; Yoshio Terada; Kent Doi; Yoshihide Fujigaki; Hideo Yasuda; Taichi Sato; Tomoyuki Fujikura; Ryohei Kuwatsuru; Hiroshi Toei; Ryusuke Murakami; Yoshihiko Saito; Atsushi Hirayama; Toyoaki Murohara; Akira Sato; Hideki Ishii; Tadateru Takayama; Makoto Watanabe; Kazuo Awai; Seitaro Oda; Takamichi Murakami; Yukinobu Yagyu; Nobuhiko Joki; Yasuhiro Komatsu; Takamasa Miyauchi; Yugo Ito; Ryo Miyazawa; Yoshihiko Kanno; Tomonari Ogawa; Hiroki Hayashi; Eri Koshi; Tomoki Kosugi; Yoshinari Yasuda
Journal:  Clin Exp Nephrol       Date:  2020-01       Impact factor: 2.801

4.  Using 80 kVp on a 320-row scanner for hepatic multiphasic CT reduces the contrast dose by 50 % in patients at risk for contrast-induced nephropathy.

Authors:  Narumi Taguchi; Seitaro Oda; Daisuke Utsunomiya; Yoshinori Funama; Takeshi Nakaura; Masanori Imuta; Sadahiro Yamamura; Hideaki Yuki; Masafumi Kidoh; Kenichiro Hirata; Tomohiro Namimoto; Masahiro Hatemura; Noriyuki Kai; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2016-05-30       Impact factor: 5.315

5.  Determination of optimal intravenous contrast agent iodine dose for the detection of liver metastasis at 80-kVp CT.

Authors:  Satoshi Goshima; Masayuki Kanematsu; Yoshifumi Noda; Hiroshi Kondo; Haruo Watanabe; Hiroshi Kawada; Nobuyuki Kawai; Yukichi Tanahashi; Kyongtae T Bae
Journal:  Eur Radiol       Date:  2014-05-28       Impact factor: 5.315

Review 6.  Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose.

Authors:  Sock Keow Tan; Chai Hong Yeong; Raja Rizal Azman Raja Aman; Kwan Hoong Ng; Yang Faridah Abdul Aziz; Kok Han Chee; Zhonghua Sun
Journal:  Br J Radiol       Date:  2018-03-29       Impact factor: 3.039

7.  Guideline on the use of iodinated contrast media in patients with kidney disease 2018.

Authors:  Yoshitaka Isaka; Hiromitsu Hayashi; Kazutaka Aonuma; Masaru Horio; Yoshio Terada; Kent Doi; Yoshihide Fujigaki; Hideo Yasuda; Taichi Sato; Tomoyuki Fujikura; Ryohei Kuwatsuru; Hiroshi Toei; Ryusuke Murakami; Yoshihiko Saito; Atsushi Hirayama; Toyoaki Murohara; Akira Sato; Hideki Ishii; Tadateru Takayama; Makoto Watanabe; Kazuo Awai; Seitaro Oda; Takamichi Murakami; Yukinobu Yagyu; Nobuhiko Joki; Yasuhiro Komatsu; Takamasa Miyauchi; Yugo Ito; Ryo Miyazawa; Yoshihiko Kanno; Tomonari Ogawa; Hiroki Hayashi; Eri Koshi; Tomoki Kosugi; Yoshinari Yasuda
Journal:  Jpn J Radiol       Date:  2020-01       Impact factor: 2.374

8.  Low contrast dose protocol involving a 100 kVp tube voltage for hypervascular hepatocellular carcinoma in patients with renal dysfunction.

Authors:  Takeshi Nakaura; Yasunori Nagayama; Masafumi Kidoh; Shinichi Nakamura; Tomohiro Namimoto; Kazuo Awai; Kazunori Harada; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2015-07-11       Impact factor: 2.374

9.  Can sufficient preoperative information of intracranial aneurysms be obtained by using 320-row detector CT angiography alone?

Authors:  Eri Hayashida; Akira Sasao; Toshinori Hirai; Kiyotoshi Hamasaki; Toru Nishi; Daisuke Utsunomiya; Seitaro Oda; Yasuhiko Iryo; Joji Urata; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2013-06-16       Impact factor: 2.374

10.  Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol.

Authors:  Jianwen Li; Zhifeng Mai; Zhihong Zhang; Jiamang Cui; Mingjie Yang; Xia Ma; Yan Wang
Journal:  Jpn J Radiol       Date:  2020-02-11       Impact factor: 2.374