A new modified fluorescein strip: Its repeatability and usefulness in tear film break-up time analysis.

H Pult1, B H Riede-Pult.   

Abstract

PURPOSE: To (i) analyze the repeatability of fluorescein instillation from a modified fluorescein strip (MFS) compared to a standard fluorescein strip (FS), and to (ii) observe its usefulness in the measurement of the fluorescein break-up time (FBUT) in comparison to the Tearscope (NIBUT).
METHODS: In-vitro: Intra- and inter-observer repeatability in fluorescein instillation from the MFS and FS was evaluated by fluorescence analysis (n=10, each). In-vivo: BUT of the right eye of 20 randomly selected subjects (mean age 43.3±11.5, range=21-60 years, 8 males, 12 females) was measured by use of the Tearscope and MFS. Subjects were grouped by the Ocular Surface Disease Index (OSDI) score into 8 OSDI+ and 12 OSDI- by a cut-off value of 15.
RESULTS: In-vitro: Intra-observer 95% limit of agreement (LoA) of the MFS was similar to the FS LoA in observer 1 (O1), but better than the FS LoA in observer (O2) (MFS: O1: LoA=±1.98 mW; p=0.179; O2: ±2.71; 0.442; FS: O1: ±1.71; 0.246; O2: ±4.11; 0.512). Inter-observer LoA in fluorescence was better in MFS (±1.42; 0.111) than in FS (±3.71; 0.003). In-vivo: MFS-BUT was significantly shorter than the NIBUT (p=0.002), but significantly correlated (r=0.864, p<0.001). NIBUT and MFS-BUT were significant discriminators (p<0.001) of OSDI±(0.948/8s and 0.938/5s [AUC/cut-off value]; NIBUT and MFS-BUT, respectively).
CONCLUSIONS: The MFS was better in the repeatability of fluorescein instillation than the FS. NIBUT and MFS-BUT were good discriminators of dry eye symptoms, but differ in their cut-off values.
Copyright © 2011 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21852182     DOI: 10.1016/j.clae.2011.07.005

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  8 in total

1.  Agreement between Automated and Traditional Measures of Tear Film Breakup.

Authors:  Stephanie M Cox; Kelly K Nichols; Jason J Nichols
Journal:  Optom Vis Sci       Date:  2015-09       Impact factor: 1.973

2.  Meibomian glands visibility assessment through a new quantitative method.

Authors:  José Vicente García-Marqués; Santiago García-Lázaro; Noelia Martínez-Albert; Alejandro Cerviño
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-01-06       Impact factor: 3.117

3.  Repeatability of Non-invasive Keratograph Break-Up Time measurements obtained using Oculus Keratograph 5M.

Authors:  José Vicente García-Marqués; Noelia Martínez-Albert; Cristian Talens-Estarelles; Santiago García-Lázaro; Alejandro Cerviño
Journal:  Int Ophthalmol       Date:  2021-03-16       Impact factor: 2.031

Review 4.  TFOS DEWS II Tear Film Report.

Authors:  Mark D P Willcox; Pablo Argüeso; Georgi A Georgiev; Juha M Holopainen; Gordon W Laurie; Tom J Millar; Eric B Papas; Jannick P Rolland; Tannin A Schmidt; Ulrike Stahl; Tatiana Suarez; Lakshman N Subbaraman; Omür Ö Uçakhan; Lyndon Jones
Journal:  Ocul Surf       Date:  2017-07-20       Impact factor: 5.033

5.  Reliability of a new modified tear breakup time method: dry tear breakup time.

Authors:  Kyung Tae Kim; Jae-Hyung Kim; Young Tae Kong; Ju Byung Chae; Sungmin Hyung
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-06-20       Impact factor: 3.117

6.  Advances in Dry Eye Disease Examination Techniques.

Authors:  Yaying Wu; Chunyang Wang; Xin Wang; Yujie Mou; Kelan Yuan; Xiaodan Huang; Xiuming Jin
Journal:  Front Med (Lausanne)       Date:  2022-01-25

7.  Reliability and efficacy of maximum fluorescein tear break-up time in diagnosing dry eye disease.

Authors:  Yujie Mou; Huan Xiang; Lin Lin; Kelan Yuan; Xin Wang; Yaying Wu; Jinjin Min; Xiuming Jin
Journal:  Sci Rep       Date:  2021-06-01       Impact factor: 4.379

8.  Agreement of noninvasive tear break-up time measurement between Tomey RT-7000 Auto Refractor-Keratometer and Oculus Keratograph 5M.

Authors:  Ryan Lee; Sharon Yeo; Han Tun Aung; Louis Tong
Journal:  Clin Ophthalmol       Date:  2016-09-16
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.