INTRODUCTION: This study was conducted in vivo to investigate the accuracy of Root ZX II in locating the apical foramen in teeth with apical periodontitis (AP).
METHODS: Twenty-seven single-root teeth scheduled for extractions were selected; in AP group, there were 12 teeth with pulp necrosis and radiographic evidence of apical lesions, and VT group consisted of 15 vital teeth. After endodontic access, the coronal portion of the canal was flared, and electronic measurement was performed by using a 15 K-file until the device level reading "APEX" was reached. The file was fixed in place, and the tooth was extracted. The apical third of the root was shaved until exposure of the file. The distance from the file tip to the most coronal border of the apical foramen was obtained.
RESULTS: The mean distance in AP group was +0.117 (standard deviation, 0.373) and in VT group was -0.105 (standard deviation, 0.218). The unpaired t test showed no difference between the groups when comparing the mean distances (P > .05). The apical foramen was accurately located within ±0.5 mm in 83% of the teeth in AP group and in 100% of VT group. Statistical analysis showed no difference between the groups (P = .1092).
CONCLUSIONS: The Root ZX II device was accurate in locating the apical foramen regardless of the presence of AP.
Copyright © 2011 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21846536     DOI: 10.1016/j.joen.2011.06.006

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

2.  Resolution of endodontic issues linked to complex anatomy.

Authors:  Guido Migliau; Erlind Pepla; Laith Kostantinos Besharat; Livio Gallottini
Journal:  Ann Stomatol (Roma)       Date:  2014-03-31

3.  Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.

Authors:  Firdevs Çınar; Yakup Üstün
Journal:  Eur Endod J       Date:  2020-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.