Magnet resonance imaging safety of the Vibrant Soundbridge system: a review.

Jan Hendrik Wagner1, Arne Ernst, Ingo Todt.   

Abstract

OBJECTIVE: To review the current knowledge on magnetic resonance imaging (MRI) safety of the Vibrant Soundbridge (VSB) system, an implantable, active middle ear implant. DATA SOURCES: National Library of Medicine's online database and clinical reports. STUDY SELECTION: All available articles from 1995 to 2010 on MRI compatibility of the VSB and other active middle ear implants.
RESULTS: Minor demagnetization was found for alignments of the magnets antiparallel to the magnetic field of the scanner, which played no clinically important role. In some cases, the torque forces upon MRI scanning can lead to pain and/or dislocation of the implant or dislocation of the floating mass transducer (FMT) depending on its type of positioning and attachment in the middle ear. Consecutive need of revision surgery is possible. Voltage induction may cause loud audible sounds; however, there were no reports for cochlear hearing loss. No significant heating effects upon MRI scanning were reported. Image artifacts occur especially because of the large magnet. In none of the patient or temporal bone studies, the ossicles, round or oval window, or other middle ear structures were injured, and there was no functional loss of the implant performance (i.e., demagnetization of the FMT).
CONCLUSION: MRI examinations of up to 1.5T may be of crucial diagnostic importance to the patients implanted with a VSB, but there seems no serious risk of harm to the patient or damage to the VSB. A dislocation of the FMT can be possible during MRI, and this does depend on transducer position and the security of the transducer to the vibratory structure and the coupling mode used. The next generation of VSB systems should consider those possible changes in its design.

Entities:  

Mesh:

Year:  2011        PMID: 21844783     DOI: 10.1097/MAO.0b013e31822a9816

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  6 in total

Review 1.  [Magnetic resonance imaging in patients with magnetic hearing implants: overview and procedural management].

Authors:  S Nospes; W Mann; A Keilmann
Journal:  Radiologe       Date:  2013-11       Impact factor: 0.635

2.  Indications and outcome of subtotal petrosectomy for active middle ear implants.

Authors:  Nicolas Verhaert; Hamidreza Mojallal; Burkard Schwab
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-07-25       Impact factor: 2.503

3.  Functional results after Bonebridge implantation in adults and children with conductive and mixed hearing loss.

Authors:  Torsten Rahne; Ingmar Seiwerth; Gerrit Götze; Cornelia Heider; Florian Radetzki; Michael Herzog; Stefan K Plontke
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-11-26       Impact factor: 2.503

4.  Vibrant Soundbridge® in preschool children with unilateral aural atresia: acceptance and benefit.

Authors:  M Leinung; E Zaretsky; B P Lange; V Hoffmann; T Stöver; C Hey
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-08-25       Impact factor: 2.503

5.  Response to: Comment on "Original Solution for Middle Ear Implant and Anesthetic/Surgical Management in a Child with Severe Craniofacial Dysmorphism".

Authors:  Giovanni Bianchin; Lorenzo Tribi; Aronne Reverzani; Patrizia Formigoni; Valeria Polizzi
Journal:  Case Rep Otolaryngol       Date:  2016-12-05

Review 6.  New Insights into MR Safety for Implantable Medical Devices.

Authors:  Kagayaki Kuroda; Satoshi Yatsushiro
Journal:  Magn Reson Med Sci       Date:  2022-02-25       Impact factor: 2.760

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.