Transtibial prosthetic socket pistoning: static evaluation of Seal-In(®) X5 and Dermo(®) Liner using motion analysis system.

H Gholizadeh1, N A Abu Osman, M Kamyab, A Eshraghi, W A B Wan Abas, M N Azam.   

Abstract

BACKGROUND: The method of attachment of prosthesis to the residual limb (suspension) and socket fitting is a critical issue in the process of providing an amputee with prosthesis. Different suspension methods try to minimize the pistoning movement inside the socket. The Seal-In(®) X5 and Dermo(®) Liner by Ossur are new suspension liners that intend to reduce pistoning between the socket and liner. Since the effects of these new liners on suspension are unclear, the objective of this study was to compare the pistoning effect of Seal-In(®) X5 and Dermo(®) Liner by using Vicon Motion System.
METHODS: Six transtibial amputees, using both the Iceross Seal-In(®) X5 and the Iceross Dermo(®) Liner, participated in the study. The vertical displacement (pistoning) was measured between the liner and socket in single limb support on the prosthetic limb (full-weight bearing), double limb support (semi-weight bearing), and non-weight bearing on the prosthetic limb, and also under three static vertical loading conditions (30 N, 60 N, and 90 N).
FINDINGS: The results demonstrated that the pistoning within the socket when Seal-In(®) X5 was used, decreased by 71% in comparison to the Iceross Dermo(®) Liner. In addition, a significant difference between the two liners under different static conditions was found (p<0.05).
INTERPRETATION: Participants needed to put in extra effort for donning and doffing the prosthesis with Seal-In(®) X5; however, this type of liner provided less pistoning. The new approach that uses the motion analysis system in this study might be an alternative for measuring the pistoning effect in the prosthetic socket.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21794965     DOI: 10.1016/j.clinbiomech.2011.07.004

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  10 in total

1.  Development of Standardized Material Testing Protocols for Prosthetic Liners.

Authors:  John C Cagle; Per G Reinhall; Brian J Hafner; Joan E Sanders
Journal:  J Biomech Eng       Date:  2017-04-01       Impact factor: 2.097

2.  Self-reported socket comfort, mobility, and balance-confidence of individuals with transtibial amputation using pinlock vs suction suspension.

Authors:  Mayank Seth; Emma Haldane Beisheim; Maximilian Tobias Spencer; John Robert Horne; Frank Bernard Sarlo; Jaclyn Megan Sions
Journal:  Prosthet Orthot Int       Date:  2021-06-01       Impact factor: 1.672

3.  Development and evaluation of new coupling system for lower limb prostheses with acoustic alarm system.

Authors:  Arezoo Eshraghi; Noor Azuan Abu Osman; Hossein Gholizadeh; Jalil Ahmadian; Bizhan Rahmati; Wan Abu Bakar Wan Abas
Journal:  Sci Rep       Date:  2013       Impact factor: 4.379

4.  Evaluation of new suspension system for limb prosthetics.

Authors:  Hossein Gholizadeh; Noor Azuan Abu Osman; Arezoo Eshraghi; Sadeeq Ali; Nooranida Arifin; Wan Abu Bakar Wan Abas
Journal:  Biomed Eng Online       Date:  2014-01-10       Impact factor: 2.819

5.  Gait biomechanics of individuals with transtibial amputation: effect of suspension system.

Authors:  Arezoo Eshraghi; Noor Azuan Abu Osman; Mohammad Karimi; Hossein Gholizadeh; Ehsan Soodmand; Wan Abu Bakar Wan Abas
Journal:  PLoS One       Date:  2014-05-27       Impact factor: 3.240

6.  Clinical implication of interface pressure for a new prosthetic suspension system.

Authors:  Hossein Gholizadeh; Noor Azuan Abu Osman; Arezoo Eshraghi; Nasrul Anuar Abd Razak
Journal:  Biomed Eng Online       Date:  2014-06-30       Impact factor: 2.819

7.  Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket.

Authors:  Veronika Noll; Stephan Rinderknecht; Philipp Beckerle
Journal:  Sensors (Basel)       Date:  2018-07-06       Impact factor: 3.576

8.  The effects of suction and pin/lock suspension systems on transtibial amputees' gait performance.

Authors:  Hossein Gholizadeh; Noor Azuan Abu Osman; Arezoo Eshraghi; Sadeeq Ali
Journal:  PLoS One       Date:  2014-05-14       Impact factor: 3.240

9.  Development of an air pneumatic suspension system for transtibial prostheses.

Authors:  Gholamhossein Pirouzi; Noor Azuan Abu Osman; Azim Ataollahi Oshkour; Sadeeq Ali; Hossein Gholizadeh; Wan A B Wan Abas
Journal:  Sensors (Basel)       Date:  2014-09-09       Impact factor: 3.576

10.  Silicone soft socket system for the treatment of geriatric transtibial amputees.

Authors:  Robert Breuer; Klemens Trieb
Journal:  Wien Klin Wochenschr       Date:  2020-04-16       Impact factor: 1.704

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.