Image quality of low-energy pulmonary CT angiography: comparison with standard CT.

Naama R Bogot1, Alexander Fingerle, Dorith Shaham, Izhak Nissenbaum, Jacob Sosna.   

Abstract

OBJECTIVE: The purpose of this article is to prospectively compare visualization of central and peripheral pulmonary arteries on simultaneously acquired low-energy and standard pulmonary CT angiography. SUBJECTS AND METHODS: Thirty-three consecutive patients (20 women and 13 men; mean age, 55.6 years; range, 21-92 years) with suspected pulmonary embolism (PE) were scanned (140 kVp; 250-300 mA) on a single-source dual-layer dual-energy MDCT scanner. Attenuation and image noise were measured at the main and segmental pulmonary arteries. Signal-to-noise ratios were calculated. Two blinded experienced radiologists assessed segmental and subsegmental artery visibility in consensus, using slab maximum-intensity-projection (MIP) reconstructions. Nonparametric sign test and kappa statistic were used for statistical analysis.
RESULTS: PE was detected in three patients (9.1%); two segmental vessel and subsegmental emboli were seen in the low-energy images only. Higher attenuation was noted in low-energy versus standard images for all arteries evaluated, with a mean (± SD) increase of 66.6 ± 4.4 HU (p < 0.0001). Low-energy images improved visualization of segmental and subsegmental arteries from 97.0% to 99.2% and from 88.0% to 93.9%, respectively. A larger number of subsegmental vessels was seen on low-energy MIP reconstructions in 69.7% (95% CI, 36.5-71.89%) of studies compared with 9.1% on the standard images. Visualization of subsegmental vessels was superior in 55.5% of cases using low-energy imaging. The mean image noise increased by 9.7 ± 0.6 HU (p < 0.0001). The mean signal-to-noise ratio showed no significant difference in the low-energy (8.2) versus standard (8.1) CT images (p = 0.7759).
CONCLUSION: Improved visualization of central and peripheral arteries can be obtained with low-energy pulmonary CT angiography, without a substantial decrease in image quality.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21785053     DOI: 10.2214/AJR.10.5318

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  6 in total

1.  Ultra-low dose contrast CT pulmonary angiography in oncology patients using a high-pitch helical dual-source technology.

Authors:  Prabhakar Rajiah; Leslie Ciancibello; Ronald Novak; Jennifer Sposato; Luis Landeras; Robert Gilkeson
Journal:  Diagn Interv Radiol       Date:  2019-05       Impact factor: 2.630

Review 2.  A new look at the heart-novel imaging techniques.

Authors:  C M Johnston; A J Krafft; M F Russe; E A Rog-Zielinska
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2017-12-14

Review 3.  Imaging of acute pulmonary embolism: an update.

Authors:  Alastair J E Moore; Jason Wachsmann; Murthy R Chamarthy; Lloyd Panjikaran; Yuki Tanabe; Prabhakar Rajiah
Journal:  Cardiovasc Diagn Ther       Date:  2018-06

4.  [Diagnostic accuracy of dual energy CT angiography in patients with diabetes mellitus].

Authors:  C Schabel; M N Bongers; D Ketelsen; R Syha; C Thomas; G Homann; M Notohamiprodjo; K Nikolaou; F Bamberg
Journal:  Radiologe       Date:  2015-04       Impact factor: 0.635

5.  A questionnaire survey reviewing radiologists' and clinical specialist radiographers' knowledge of CT exposure parameters.

Authors:  S J Foley; M G Evanoff; L A Rainford
Journal:  Insights Imaging       Date:  2013-09-05

6.  Impact of iodine concentration and scan parameters on image quality, contrast enhancement and radiation dose in thoracic CT.

Authors:  Marian S Solbak; Mette K Henning; Andrew England; Anne C Martinsen; Trond M Aaløkken; Safora Johansen
Journal:  Eur Radiol Exp       Date:  2020-09-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.